﻿<?xml version="1.0" encoding="utf-8"?><Regulation regulation-type="SOR" xml:lang="en" startdate="20060601"><Identification Code="id=&quot;&quot;" hasPreviousVersion="true"><LimsAuthority><Alpha>R-1</Alpha><AuthorityTitle>Radiation Emitting Devices Act</AuthorityTitle></LimsAuthority><InstrumentNumber>C.R.C., c. 1370</InstrumentNumber><LastAmendedDate><Date><YYYY>2005</YYYY><MM>2</MM><DD>23</DD></Date></LastAmendedDate><ConsolidationDate><Date><YYYY>2013</YYYY><MM>01</MM><DD>28</DD></Date></ConsolidationDate><LastModifiedDate><Date><YYYY>2006</YYYY><MM>7</MM><DD>21</DD></Date></LastModifiedDate><EnablingAuthority Code="id=&quot;&quot;,ea=&quot;&quot;"><XRefExternal reference-type="act">RADIATION EMITTING DEVICES ACT</XRefExternal></EnablingAuthority><ShortTitle Code="id=&quot;&quot;,st=&quot;&quot;">Radiation Emitting Devices Regulations</ShortTitle><LongTitle Code="id=&quot;&quot;,lt=&quot;&quot;">Regulations Respecting Radiation Emitting Devices</LongTitle></Identification><Body><Heading Code="ga=&quot;s_1&quot;,h1=&quot;&quot;" level="1"><TitleText Code="ga=&quot;s_1&quot;,h1=&quot;&quot;,t1=&quot;&quot;,nb=&quot;1&quot;">SHORT TITLE</TitleText></Heading><Section Code="se=&quot;1&quot;"><Label>1.</Label><Text>These Regulations may be cited as the <XRefExternal reference-type="regulation" link="C.R.C.,_c._1370">Radiation Emitting Devices Regulations</XRefExternal>.</Text></Section><Heading Code="ga=&quot;s_2&quot;,h1=&quot;&quot;" level="1"><TitleText Code="ga=&quot;s_2&quot;,h1=&quot;&quot;,t1=&quot;&quot;,nb=&quot;2&quot;">INTERPRETATION</TitleText></Heading><Section Code="se=&quot;2&quot;"><Label>2.</Label><Text>In these Regulations,</Text><Definition generate-in-text="no" Code="se=&quot;2&quot;,df=&quot;{Act}{Loi}&quot;"><Text><DefinedTermEn>Act</DefinedTermEn> means the <XRefExternal reference-type="act" link="R-1">Radiation Emitting Devices Act</XRefExternal>; (<DefinedTermFr>Loi</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no" Code="se=&quot;2&quot;,df=&quot;{device}{dispositif}&quot;"><Text><DefinedTermEn>device</DefinedTermEn> means a radiation emitting device that is within a class of radiation emitting devices prescribed in these Regulations; (<DefinedTermFr>dispositif</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no" Code="se=&quot;2&quot;,df=&quot;{Minister}{ministre}&quot;"><Text><DefinedTermEn>Minister</DefinedTermEn> means the Minister of National Health and Welfare. (<DefinedTermFr>ministre</DefinedTermFr>)</Text></Definition></Section><Heading Code="ga=&quot;s_3&quot;,h1=&quot;&quot;" level="1"><TitleText Code="ga=&quot;s_3&quot;,h1=&quot;&quot;,t1=&quot;&quot;,nb=&quot;3&quot;">PRESCRIPTION OF CLASSES OF RADIATION EMITTING DEVICES AND STANDARDS THEREFOR</TitleText></Heading><Section Code="se=&quot;3&quot;"><Label>3.</Label><Subsection Code="se=&quot;3&quot;,ss=&quot;1&quot;"><Label>(1)</Label><Text>The radiation emitting devices described in Schedule I are prescribed as classes of radiation emitting devices for the purposes of the Act.</Text></Subsection><Subsection Code="se=&quot;3&quot;,ss=&quot;2&quot;"><Label>(2)</Label><Text>The standards set out in Schedule II for prescribed classes of radiation emitting devices are prescribed as standards regulating the design, construction or functioning of those prescribed classes of radiation emitting devices and the components thereof.</Text></Subsection></Section><Heading Code="ga=&quot;s_4&quot;,h1=&quot;&quot;" level="1"><TitleText Code="ga=&quot;s_4&quot;,h1=&quot;&quot;,t1=&quot;&quot;,nb=&quot;4&quot;">DETENTION OF SEIZED DEVICES</TitleText></Heading><Section Code="se=&quot;4&quot;"><Label>4.</Label><Subsection Code="se=&quot;4&quot;,ss=&quot;1&quot;"><Label>(1)</Label><Text>An inspector who seizes a device or a component of a device pursuant to subsection 10(1) of the Act may</Text><Paragraph Code="se=&quot;4&quot;,ss=&quot;1&quot;,p1=&quot;a&quot;"><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>detain the device or component for such time as is necessary for him to ascertain whether the device or component complies with the standards prescribed therefor;</Text></Paragraph><Paragraph Code="se=&quot;4&quot;,ss=&quot;1&quot;,p1=&quot;b&quot;"><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>keep or store the device or component in the building or place where it was seized or remove it to any other place for testing to ascertain whether or not the component or device complies with the standards prescribed therefor; or</Text></Paragraph><Paragraph Code="se=&quot;4&quot;,ss=&quot;1&quot;,p1=&quot;c&quot;"><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>release the device or component unconditionally or on the condition that</Text><Subparagraph Code="se=&quot;4&quot;,ss=&quot;1&quot;,p1=&quot;c&quot;,p2=&quot;i&quot;"><Label>(i)</Label><Text>it be destroyed forthwith or retained by the owner until it is destroyed, or</Text></Subparagraph><Subparagraph Code="se=&quot;4&quot;,ss=&quot;1&quot;,p1=&quot;c&quot;,p2=&quot;ii&quot;"><Label>(ii)</Label><Text>within a period fixed by the inspector, it be returned</Text><Clause Code="se=&quot;4&quot;,ss=&quot;1&quot;,p1=&quot;c&quot;,p2=&quot;ii&quot;,c1=&quot;A&quot;"><Label>(A)</Label><Text>to the country from which it was exported, or</Text></Clause><Clause Code="se=&quot;4&quot;,ss=&quot;1&quot;,p1=&quot;c&quot;,p2=&quot;ii&quot;,c1=&quot;B&quot;"><Label>(B)</Label><Text>to the manufacturer, distributor or importer thereof and be made to comply with the standards prescribed therefor.</Text></Clause></Subparagraph></Paragraph></Subsection><Subsection Code="se=&quot;4&quot;,ss=&quot;2&quot;"><Label>(2)</Label><Text>An inspector who seizes a device or a component of a device pursuant to subsection 10(1) of the Act shall keep a written record of any such seizure, signed by him and setting out</Text><Paragraph Code="se=&quot;4&quot;,ss=&quot;2&quot;,p1=&quot;a&quot;"><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the place, date and hour of the seizure;</Text></Paragraph><Paragraph Code="se=&quot;4&quot;,ss=&quot;2&quot;,p1=&quot;b&quot;"><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the name of the person who owns the device or component or on whose premises the device or component was seized;</Text></Paragraph><Paragraph Code="se=&quot;4&quot;,ss=&quot;2&quot;,p1=&quot;c&quot;"><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the name of the manufacturer, the model number and the date and place of manufacture of the device or component;</Text></Paragraph><Paragraph Code="se=&quot;4&quot;,ss=&quot;2&quot;,p1=&quot;d&quot;"><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the place, if any, to which the device or component has been removed; and</Text></Paragraph><Paragraph Code="se=&quot;4&quot;,ss=&quot;2&quot;,p1=&quot;e&quot;"><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>particulars of the manner in which the device or component has been dealt with after seizure.</Text></Paragraph></Subsection><Subsection Code="se=&quot;4&quot;,ss=&quot;3&quot;"><Label>(3)</Label><Text>The person who owns a device or a component of a device seized pursuant to subsection 10(1) of the Act or on whose premises the device or component was seized or an employee or an agent of such person shall sign the record kept pursuant to subsection (2) and shall receive a copy thereof.</Text></Subsection><Subsection Code="se=&quot;4&quot;,ss=&quot;4&quot;"><Label>(4)</Label><Text>Where a device or a component of a device has been released pursuant to paragraph (1)(<Emphasis style="italic">c</Emphasis>) on a condition and the condition has not been complied with, the inspector may again seize and detain the device or component.</Text></Subsection></Section><Heading Code="ga=&quot;s_5&quot;,h1=&quot;&quot;" level="1"><TitleText Code="ga=&quot;s_5&quot;,h1=&quot;&quot;,t1=&quot;&quot;,nb=&quot;5&quot;">DISPOSITION OF FORFEITED DEVICES</TitleText></Heading><Section Code="se=&quot;5&quot;"><Label>5.</Label><Text>Where a court has, pursuant to subsection 16(1) of the Act, directed that a device or a component of a device be forfeited to Her Majesty, and where no application under section 75 of the <XRefExternal reference-type="act" link="F-14">Fisheries Act</XRefExternal> has been made or a declaration as provided in subsection 75(4) of that Act is not included in the final order made pursuant to that section, the device or component shall be disposed of as the Minister directs.</Text><HistoricalNote><ul><li>SOR/91-408, s. 1.</li></ul></HistoricalNote></Section></Body><Schedule spanlanguages="no" bilingual="no" Code="sc=&quot;&quot;,nb=&quot;1&quot;"><ScheduleFormHeading><Label>SCHEDULE I</Label><OriginatingRef>(s. 3)</OriginatingRef></ScheduleFormHeading><RegulationPiece><Section><Label>1.</Label><Text>Television receivers, including video monitors and video display systems, being electronic appliances designed to display a picture or alphanumeric information, or both, after receiving signals through electromagnetic waves, cable or other means of transmission and including the cabinet or case of such appliances.</Text></Section><Section><Label>2.</Label><Text>Dental X-ray equipment with an extra-oral source, being X-ray generating equipment that is designed primarily for the examination of dental structures in humans and that has an X-ray generating tube designed to be used outside the mouth.</Text></Section><Section><Label>3.</Label><Text>Microwave ovens, being appliances or sets of components that are designed to supply microwave energy to material within a cavity.</Text></Section><Section><Label>4.</Label><Text>A baggage inspection X-ray device, being an X-ray generating appliance designed primarily for the examination of carry-on baggage, or the examination of parcels, mail or similar items, including the X-ray generator, the X-ray detector and display and control systems.</Text></Section><Section><Label>5.</Label><Text>A demonstration-type gas discharge device, being a device that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>contains an electronic device in which glow discharges or X-rays or both may be produced by the acceleration of electrons and ions; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is designed to demonstrate the production, properties or effects of glow discharges or X-rays, or the flow of electrons or ions.</Text></Paragraph></Section><Section><Label>6.</Label><Text>Photofluorographic X-ray equipment being X-ray generating appliances designed primarily for the examination of the human chest and the recording photographically in reduced size of the image produced on a fluorescent screen.</Text></Section><Section><Label>7.</Label><Text>A laser scanner, being a device that uses scanned laser radiation within the wavelength range of 400 to 1400 nanometres to decipher or generate codes represented by drawn or printed geometrical patterns.</Text></Section><Section><Label>8.</Label><Text>A demonstration laser, being a device that consists of or incorporates a laser and that is primarily intended to be used for demonstrating the principles of optics in educational institutions.</Text></Section><Section><Label>9.</Label><Text>Low Energy Electron Microscopes being electron-optical devices with an operating energy of 500 kilo-electron volts (keV) or less in which a beam of electrons, focused by means of electron lenses, is used to produce an enlarged image of a minute object on a fluorescent screen, photographic plate or any other detector-display system, including both the transmission and scanning types of devices.</Text></Section><Section><Label>10.</Label><Text>High intensity mercury vapour discharge lamps, being lamps incorporating a high-pressure arc discharge tube with a fill consisting primarily of mercury, whether such lamps are described as mercury vapour lamps, metal halide lamps, self-ballasted lamps or otherwise when sold, but not including tungsten filament self-ballasted lamps.</Text></Section><Section><Label>11.</Label><Text>Tanning equipment as defined in section 1 of Part XI of Schedule II.</Text></Section><Section><Label>12.</Label><Text>Diagnostic X-ray equipment, being X-ray devices that are used for the examination of the human body, not including dental X-ray equipment with an extra-oral source that is subject to Part II of these Regulations, photofluorographic X-ray equipment, radiation therapy simulators and computer-assisted tomographic equipment.</Text></Section><Section><Label>13.</Label><Text>Ultrasound therapy devices, being devices designed to generate and emit ultrasonic power at acoustic frequencies above 20 kHz for use in physical therapy.</Text></Section><Section><Label>14.</Label><Text>Analytical X-ray equipment, being X-ray generating devices that contain an X-ray tube and that use X-radiation to determine the elemental composition, or examine the microstructure, of material.</Text></Section><Section><Label>15.</Label><Text>Cabinet X-ray equipment, being X-ray generating devices, not including analytical X-ray equipment or baggage inspection X-ray devices, that have the X-ray tube permanently installed in a cabinet and are designed primarily for the examination of material, part or all of which is placed within the cabinet.</Text></Section><Section><Label>16.</Label><Text><Repealed>[Revoked, SOR/88-471, s. 1]</Repealed></Text></Section></RegulationPiece><HistoricalNote><ul><li>SOR/78-407, s. 1;</li><li> SOR/79-229, s. 1;</li><li> SOR/80-381, s. 1;</li><li> SOR/80-464, s. 1;</li><li> SOR/81-23, s. 1;</li><li> SOR/81-286, s. 1;</li><li> SOR/81-545, s. 1;</li><li> SOR/82-981, s. 1;</li><li> SOR/83-495, s. 1;</li><li> SOR/88-471, s. 1;</li><li> SOR/93-201, s. 1;</li><li> SOR/97-511, s. 1;</li><li> SOR/2001-252, s. 1;</li><li> SOR/2005-33, s. 1.</li></ul></HistoricalNote></Schedule><Schedule spanlanguages="no" bilingual="no" Code="sc=&quot;&quot;,nb=&quot;2&quot;"><ScheduleFormHeading><Label>SCHEDULE II</Label><OriginatingRef>(s. 3)</OriginatingRef></ScheduleFormHeading><RegulationPiece><Heading level="1"><Label>PART I</Label><TitleText>TELEVISION RECEIVERS</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>maximum test voltage</DefinedTermEn> means</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>127 volts, if the television receiver is designed to operate from a 110-120 volt power source, or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>110 per cent of the maximum voltage from which the television receiver is designed to operate, if the television receiver is designed to operate otherwise than from a 110-120 volt power source; (<DefinedTermFr>tension maximale d’essai</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>service control</DefinedTermEn> means a control that is installed on a television receiver by the manufacturer thereof for the purpose of adjustment and that, under normal usage, is not accessible to the user of the receiver; (<DefinedTermFr>commande interne</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>user control</DefinedTermEn> means a control that is provided on or external to a television receiver by the manufacturer thereof for the purpose of adjustment and that for a fully assembled television receiver under normal usage, is accessible to the user. (<DefinedTermFr>commande externe</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Subsection><Label>(1)</Label><Text>Every television receiver shall be designed and constructed in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>under normal conditions of use, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>on failure or malfunction of any one component or subsequent failure or malfunction of other components caused by that failure or malfunction,</Text></Paragraph><ContinuedSectionSubsection><Text>it functions in accordance with the standards of functioning described in section 3 for as long as the receiver has its original components or has replacement components recommended by the manufacturer.</Text></ContinuedSectionSubsection></Subsection><Subsection><Label>(2)</Label><Text>Without limiting the generality of subsection (1), the components of a receiver that emit X-rays shall have sufficient shielding to enable the receiver to comply with the standards of functioning set out in section 3.</Text></Subsection><Subsection><Label>(3)</Label><Text>Where the shielding required by subsection (2) necessitates the use of individual shields, such shields shall either</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>be non-removable, or interlocked; or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>bear a radiation warning sign that is permanently affixed and clearly visible under the conditions of servicing, and cautions against operation of the receiver when the shield is removed.</Text></Paragraph></Subsection><Subsection><Label>(4)</Label><Text>Subject to subsection (6), every television receiver shall bear on the rear external surface of its cabinet or case, a permanent label that sets out, with respect to the receiver</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the name and address of the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the model number;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the city and country of manufacture or a code by which that city and country are identified and the key to which is supplied by the manufacturer to the Minister before the sale of the receiver;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the month and year of manufacture, without abbreviation, with the year shown as a four-digit number;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>the brand name of the receiver;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>the chassis family designation; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>the serial number.</Text></Paragraph></Subsection><Subsection><Label>(5)</Label><Text>The permanent label required by subsection (4) shall be clearly visible at any time.</Text></Subsection><Subsection><Label>(6)</Label><Text>Where a television receiver is manufactured for sale under the trade name of a person who is not the manufacturer of the receiver, the trade name and the address of that person may, for the purposes of paragraph (4)(<Emphasis style="italic">a</Emphasis>), be substituted for the name and address of the manufacturer if the name and address of the manufacturer are permanently affixed to the inside of the cabinet or case of the receiver and are clearly visible under conditions of servicing.</Text></Subsection><Subsection><Label>(7) and (8)</Label><Text><Repealed>[Revoked, SOR/94-40, s. 1]</Repealed></Text></Subsection></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>3.</Label><Text>Every television receiver shall function in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>when the receiver</Text><Subparagraph><Label>(i)</Label><Text>is fully assembled,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is used with any supply voltage up to the maximum test voltage,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>is used with any settings of the user controls and service controls, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>displays a synchronous raster covering at least 60 per cent of the viewable screen area,</Text></Subparagraph><ContinuedParagraph><Text>the emission of ionizing radiation therefrom is such that the exposure rate of X-rays, when averaged over a period of five minutes, to an object having a 10 square centimetre cross section and centred at five centimetres from any accessible external surface of the television receiver does not exceed 0.5 milliroentgen per hour; and</Text></ContinuedParagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>when the receiver, without its cabinet or case, or any part of the chassis, cabinet or case,</Text><Subparagraph><Label>(i)</Label><Text>is used with any supply voltage up to the maximum test voltage,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is used with any settings of the user controls and service controls, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>displays a synchronous raster covering at least 60 per cent of the viewable screen area,</Text></Subparagraph><ContinuedParagraph><Text>the emission of ionizing radiation therefrom is such that the exposure rate of X-rays, when averaged over a period of five minutes, to an object having a 10 square centimetre cross section and centred at 10 centimetres from the surface of any component of the receiver does not exceed 2.5 milliroentgen per hour.</Text></ContinuedParagraph></Paragraph></Section><Heading level="1"><Label>PART II</Label><TitleText>DENTAL X-RAY EQUIPMENT WITH AN EXTRA-ORAL SOURCE</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Subsection><Label>(1)</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>coefficient of variation</DefinedTermEn> means the ratio of the standard deviation to the mean value of a series of measurements, calculated by using the following equation:</Text><FormulaGroup><Formula><ImageGroup position="center"><AlternateText>The coefficient of variation, calculated by the equation “C equals one divided by x bar multiplied by, open square bracket, sum from i equals one to n, open parenthesis, x sub i minus x bar, close parenthesis, squared, divided by n minus one, close square bracket, to the power of one half”</AlternateText><Image source="CRC1370img\CRC1370_e014.png" /></ImageGroup></Formula><FormulaConnector>where</FormulaConnector><FormulaDefinition><FormulaTerm>C</FormulaTerm><Text>is the coefficient of variation,</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>X<Sub>i</Sub></FormulaTerm><Text>is the value of the i<Sup>th</Sup> measurement,</Text></FormulaDefinition><FormulaDefinition><FormulaTerm><Emphasis style="overbar">X</Emphasis></FormulaTerm><Text>is the mean value of the measurements, and</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>n</FormulaTerm><Text>is the number of measurements; (<DefinedTermFr>coefficient de variation</DefinedTermFr>)</Text></FormulaDefinition></FormulaGroup></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>radiation exposure to the X-ray image receptor</DefinedTermEn> means the amount of radiation registered by one or more detectors located in proximity to the X-ray image receptor. (<DefinedTermFr>dose d’irradiation au récepteur d’image radiologique</DefinedTermFr>)</Text></Definition></Subsection><Subsection><Label>(2)</Label><Text>Unless otherwise provided, the other words and expressions used in this Part have the same meaning as in the International Electrotechnical Commission Standard entitled <XRefExternal reference-type="other">Medical radiology — Terminology</XRefExternal>, Publication 788, First edition, 1984.</Text></Subsection></Section><Heading level="2"><TitleText>Design Standards</TitleText></Heading><Section><Label>2.</Label><Subsection><Label>(1)</Label><Text>Dental X-ray equipment with an extra-oral source shall be designed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>electrical meters or other indicators that</Text><Subparagraph><Label>(i)</Label><Text>are clearly visible to the operator,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>are securely affixed to the X-ray equipment control panel,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>show a set of loading factors, by control of which the operator is able to obtain radiograms of diagnostic quality, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>if the X-ray equipment operates by automatic exposure control when the X-ray tube is energized, provide a readily discernible visual or aural warning signal whenever the equipment cannot provide a radiogram of diagnostic quality;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>separate warning indicators that</Text><Subparagraph><Label>(i)</Label><Text>where aural, are clearly audible to the operator,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>where visual, are</Text><Clause><Label>(A)</Label><Text>clearly visible to the operator, and</Text></Clause><Clause><Label>(B)</Label><Text>affixed to the X-ray equipment control panel, and</Text></Clause></Subparagraph><Subparagraph><Label>(iii)</Label><Text>are readily discernible and clearly marked to indicate</Text><Clause><Label>(A)</Label><Text>visually when</Text><Subclause><Label>(I)</Label><Text>the filament of the X-ray tube is carrying current,</Text></Subclause><Subclause><Label>(II)</Label><Text>the control panel is energized,</Text></Subclause><Subclause><Label>(III)</Label><Text>the automatic exposure control has been selected, and</Text></Subclause><Subclause><Label>(IV)</Label><Text>the loading factors controlled by the automatic exposure control have reached the limits specified in subparagraph 4(1)(<Emphasis style="italic">e</Emphasis>)(ii), and</Text></Subclause></Clause><Clause><Label>(B)</Label><Text>visually and aurally when X-rays are being produced;</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>an irradiation switch that</Text><Subparagraph><Label>(i)</Label><Text>requires continuous pressure by the operator until the completion of an irradiation, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is installed so as to allow the operator to stand at least 3 m from the X-ray source when the X-ray tube is energized;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>a controlling timer that</Text><Subparagraph><Label>(i)</Label><Text>when the equipment is not operating in panoramic mode, automatically resets itself to its original setting or to zero on the termination of an irradiation,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>prevents an irradiation from being initiated when it is set at zero or in the off position,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>causes the production of X-rays to be automatically terminated on the attainment of a preset</Text><Clause><Label>(A)</Label><Text>irradiation time,</Text></Clause><Clause><Label>(B)</Label><Text>current time product, or</Text></Clause><Clause><Label>(C)</Label><Text>radiation exposure to the X-ray image receptor, and</Text></Clause></Subparagraph><Subparagraph><Label>(iv)</Label><Text>when the equipment is operating in automatic exposure mode, ensures that the maximum irradiation time or the maximum current time product does not exceed the limits specified in clause 4(1)(<Emphasis style="italic">e</Emphasis>)(ii)(C) or subparagraph 4(1)<Emphasis style="italic">(e</Emphasis>)(iv), whichever is applicable;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>a localizing cone or other device that limits the focal spot to skin distance to not less than</Text><Subparagraph><Label>(i)</Label><Text>15 cm, for equipment designed for panoramic examinations, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>18 cm, for all other equipment;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>beam limiting devices that</Text><Subparagraph><Label>(i)</Label><Text>provide a degree of radiation protection from stray radiation such that stray radiation does not exceed the limit for leakage radiation from the X-ray tube housing set out in paragraph 4(1)(<Emphasis style="italic">g</Emphasis>), and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>limit the size of the X-ray beam</Text><Clause><Label>(A)</Label><Text>at the X-ray image receptor of equipment designed for panoramic examinations, to a size that does not exceed any dimension of the scanning slit by more than one-half of that dimension or more than 2 per cent of the focal spot to image receptor distance, whichever is the lesser,</Text></Clause><Clause><Label>(B)</Label><Text>where the equipment is designed for and operated in cephalometric mode, to a circle not more than 30 cm in diameter or a rectangle not more than 800 cm<Sup>2</Sup> in area, fully intercepted by the X-ray image receptor, at a distance of 1.5 m or at the maximum focal spot to image receptor distance, whichever is the lesser, and</Text></Clause><Clause><Label>(C)</Label><Text>where the equipment is operated with an intra-oral X-ray image receptor, to a circle not more than 7 cm in diameter or a rectangle not more than 38.5 cm<Sup>2</Sup> in area;</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>radiation-absorbing filters that</Text><Subparagraph><Label>(i)</Label><Text>are securely installed in the path of the X-ray beam, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>provide a degree of attenuation of the X-ray beam such that the first half-value layer of aluminum is not less than the value shown in column II of the table to this subparagraph that corresponds to the X-ray tube voltage shown in column I of the table, or is not less than the value obtained by linear interpolation from that table;</Text><TableGroup pointsize="9" topmarginspacing="4"><table frame="topbot"><tgroup cols="3"><colspec colname="1" colwidth="84.0*" htmlwidth="18%" /><colspec colname="2" colwidth="192.00*" htmlwidth="41%" /><colspec colname="3" colwidth="192.00*" htmlwidth="41%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t853h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t853h1">Column I</entry><entry colsep="0" rowsep="0" valign="bottom" id="t853h2">Column II</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t853h3" headers="t853h1">X-ray Tube Voltage (Kilovolts (Peak Value))</entry><entry colsep="0" rowsep="1" valign="bottom" id="t853h4" headers="t853h2">First Half-value Layer of Aluminum (mm)</entry></row></thead><tbody><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t853h0">1.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t853h1 t853h3">50</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t853h2 t853h4">1.5</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t853h0">2.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t853h1 t853h3">60</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t853h2 t853h4">1.5</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t853h0">3.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t853h1 t853h3">70</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t853h2 t853h4">1.5</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t853h0">4.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t853h1 t853h3">71</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t853h2 t853h4">2.1</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t853h0">5.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t853h1 t853h3">80</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t853h2 t853h4">2.3</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t853h0">6.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t853h1 t853h3">90</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t853h2 t853h4">2.5</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t853h0">7.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t853h1 t853h3">100</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t853h2 t853h4">2.7</entry></row></tbody></tgroup></table></TableGroup></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>on the external surface of the X-ray tube housing or on a suitable structure rigidly and permanently affixed to the X-ray tube housing, a clearly visible mark or marks indicating, to within 4 mm, the location along the X-ray beam axis of the focal spot on the target; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>where the equipment is equipped with an automatic exposure control, an interlock that, when an automatically-timed irradiation has terminated because the limits specified in subparagraph 4(1)(<Emphasis style="italic">e</Emphasis>)(ii) have been reached, requires the operator to manually reset the equipment to its original setting before another irradiation can be made.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Dental X-ray equipment with an extra-oral source shall, where more than one X-ray tube is controlled by one control panel, be designed to include, in addition to the safety features required by subsection (1),</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>an interlock that prevents the energizing of more than one X-ray tube at the same time;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>on or near each X-ray tube housing, so as to be clearly visible to the operator, a visual indicator that indicates when the X-ray tube is connected and ready to be energized; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>on the control panel, so as to be clearly visible to the operator, a visual indicator that indicates which X-ray tubes are connected and ready to be energized.</Text></Paragraph></Subsection></Section><Heading level="2"><TitleText>Construction Standards</TitleText></Heading><Section><Label>3.</Label><Text>Dental X-ray equipment with an extra-oral source shall be constructed of such materials and in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the X-ray tube is securely fixed and correctly aligned within the X-ray tube housing;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the X-ray source assembly maintains its position or its intended motion without tipping, excessive drift or vibration during irradiation;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>where the equipment has its original components or replacement components recommended by the manufacturer, the equipment functions, under normal conditions of use, in accordance with the functioning standards set out in subsection 4(1); and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the exposure of ionizing radiation or kerma emitted by the X-ray source assembly when the irradiation control circuit has not been activated, or by any other component at any time, does not exceed 645 nanocoulombs per kilogram (2.5 milliroentgens) or 22 micrograys, in any one-hour period, when averaged over a detection area of 10 cm<Sup>2</Sup> and measured at a distance of 5 cm from any accessible surface of the equipment.</Text></Paragraph></Section><Heading level="2"><TitleText>Functioning Standards</TitleText></Heading><Section><Label>4.</Label><Subsection><Label>(1)</Label><Text>Dental X-ray equipment with an extra-oral source shall, when fully assembled for use and tested under the test conditions referred to in subsection (2), function in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the preset X-ray tube voltage cannot be below 50 kilovolts (peak value);</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>where a series of 10 consecutive radiation measurements is taken at the same distance from the target in the X-ray beam within a period of one hour, and where all variable controls for loading factors are adjusted to other settings and reset to the test setting before each measurement, the coefficient of variation of the measurements is not greater than 0.05;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the actual operating X-ray tube voltage</Text><Subparagraph><Label>(i)</Label><Text>is not less than 50 kilovolts (peak value), and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>does not deviate from the indicated value by more than the maximum allowable deviation specified by the manufacturer in accordance with paragraph 5(2)(<Emphasis style="italic">b</Emphasis>);</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>where the design of the equipment allows the X-ray tube voltage to fall below 50 kilovolts (peak value) during an irradiation, a warning indicator gives a clearly visible or audible signal when conditions that result in an X-ray tube voltage lower than 50 kilovolts (peak value) occur;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>the controlling timer referred to in paragraph 2(1)(<Emphasis style="italic">d</Emphasis>)</Text><Subparagraph><Label>(i)</Label><Text>at each setting meets the accuracy limits specified by the manufacturer in accordance with subparagraph 5(2)(<Emphasis style="italic">b</Emphasis>)(ii),</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>where the equipment is designed for conventional dental examinations,</Text><Clause><Label>(A)</Label><Text>is such that the minimum value at which it can be set is equal to or less than the longest of the minimum irradiation times set out in columns II to IV of the table to this clause for the minimum X-ray tube voltage shown in column I of the table,</Text><TableGroup pointsize="9" spanlanguages="yes" topmarginspacing="4"><table frame="topbot"><tgroup cols="5"><colspec colname="1" colwidth="46.29*" htmlwidth="10%" /><colspec colname="2" colwidth="142.79*" htmlwidth="31%" /><colspec colname="3" colwidth="105.02*" htmlwidth="22%" /><colspec colname="4" colwidth="80.38*" htmlwidth="17%" /><colspec colname="5" colwidth="93.52*" htmlwidth="20%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="2" colsep="0" rowsep="1" valign="bottom" id="t854h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t854h1">Column I</entry><entry colsep="0" rowsep="0" valign="bottom" id="t854h2">Column II</entry><entry colsep="0" rowsep="0" valign="bottom" id="t854h3">Column III</entry><entry colsep="0" rowsep="0" valign="bottom" id="t854h4">Column IV</entry></row><row valign="bottom" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t854h5" headers="t854h1">Minimum X-ray Tube Voltage (Kilovolts (Peak Value))</entry><entry namest="3" nameend="5" colsep="0" rowsep="1" align="center" valign="bottom" id="t854h6" headers="t854h2 t854h3 t854h4">Minimum Irradiation Time</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t854h7" headers="t854h2 t854h6">(Seconds)</entry><entry colsep="0" rowsep="1" valign="bottom" id="t854h8" headers="t854h3 t854h6">(Cycles)</entry><entry colsep="0" rowsep="1" valign="bottom" id="t854h9" headers="t854h4 t854h6">(Milliampere-seconds)</entry></row></thead><tbody><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t854h0">1.</entry><entry rowheader="yes" colname="2" colsep="0" rowsep="0" valign="top" id="t854h11" headers="t854h1 t854h5">Up to 70</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h11 t854h2 t854h6 t854h7">1/20</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h11 t854h3 t854h6 t854h8">3</entry><entry colname="5" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h11 t854h4 t854h6 t854h9">0.75</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t854h0">2.</entry><entry rowheader="yes" colname="2" colsep="0" rowsep="0" valign="top" id="t854h16" headers="t854h1 t854h5">71 to 80</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h16 t854h2 t854h6 t854h7">1/30</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h16 t854h3 t854h6 t854h8">2</entry><entry colname="5" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h16 t854h4 t854h6 t854h9">0.5</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t854h0">3.</entry><entry rowheader="yes" colname="2" colsep="0" rowsep="0" valign="top" id="t854h21" headers="t854h1 t854h5">81 or more</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h21 t854h2 t854h6 t854h7">1/60</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h21 t854h3 t854h6 t854h8">1</entry><entry colname="5" colsep="0" rowsep="0" valign="top" headers="t854h1 t854h5 t854h21 t854h4 t854h6 t854h9">0.25</entry></row></tbody></tgroup></table></TableGroup></Clause><Clause><Label>(B)</Label><Text>in the case of a timer that has a scale of irradiation times or milliampere-second values, is such that the ratio of no two consecutive settings exceeds 1.25:1, except for times not greater than 1/20 second, 3 cycles or the equivalent milliampere-second values, and</Text></Clause><Clause><Label>(C)</Label><Text>has a maximum irradiation time of no longer than 5 seconds or the time required to deliver 50 milliampere-seconds, whichever is the shorter,</Text></Clause></Subparagraph><Subparagraph><Label>(iii)</Label><Text>where the equipment is designed for operation in cephalometric mode, but not for conventional dental examinations, is such that the minimum value at which it can be set is equal to or less than the longest of</Text><Clause><Label>(A)</Label><Text>1/10 second,</Text></Clause><Clause><Label>(B)</Label><Text>6 cycles, and</Text></Clause><Clause><Label>(C)</Label><Text>the time required to deliver 3 milliampere-seconds, and</Text></Clause></Subparagraph><Subparagraph><Label>(iv)</Label><Text>where the equipment is designed for operation in panoramic mode, has a maximum irradiation time of not longer than 25 seconds or the time required to deliver 250 milliampere-seconds, whichever is the shorter;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>for any selected X-ray tube voltage within the range of values of operating X-ray tube voltages specified for the equipment and for any irradiation time equal to or greater than the longest of the minimum irradiation times set out in columns II to IV of the table to clause 4(1)(<Emphasis style="italic">e</Emphasis>)(ii)(A) for the minimum X-ray tube voltage shown in column I of the table, the following relation shall hold:</Text><FormulaGroup><Formula><FormulaText>X<Sub>l</Sub> - X<Sub>2</Sub> &lt; 0.1 (X<Sub>1</Sub> + X<Sub>2</Sub>)</FormulaText></Formula><FormulaConnector>where </FormulaConnector><FormulaDefinition><FormulaTerm>X<Sub>l</Sub> and X<Sub>2</Sub></FormulaTerm><Text>are the average exposure values (kerma) per second, per pulse or per milliampere-second obtained</Text><FormulaParagraph indent-level="3"><Label>(i)</Label><Text>where the X-ray tube current is fixed, at each two settings of the controlling timer that do not differ by more than a factor of two, and</Text></FormulaParagraph><FormulaParagraph indent-level="3"><Label>(ii)</Label><Text>where the irradiation time is fixed, at each two X-ray tube current settings that do not differ by more than a factor of two; and</Text></FormulaParagraph></FormulaDefinition></FormulaGroup></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>the leakage radiation from the X-ray tube housing, when measured at a distance of 1 m from the target and averaged over an area of 100 cm<Sup>2</Sup> having no linear dimension greater than 20 cm, does not exceed 25.8 microcoulombs per kilogram (100 milliroentgens) or 0.87 milligrays in any one-hour period under any combination of loading factors within the rated limits of use of the equipment.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Any testing of dental X-ray equipment with an extra-oral source that is carried out to verify its compliance with the functioning standards set out in subsection (1) shall be conducted under the following conditions:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the unloaded line voltage must remain within 1 per cent of its nominal value; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the line voltage must be regulated in such a manner that it does not vary by more than 6 per cent when the line is fully loaded at the maximum rated line current of the equipment.</Text></Paragraph></Subsection></Section><Heading level="2"><TitleText>Labelling and Information</TitleText></Heading><Section><Label>5.</Label><Subsection><Label>(1)</Label><Text>Dental X-ray equipment with an extra-oral source shall bear</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>an X-radiation warning symbol that</Text><Subparagraph><Label>(i)</Label><Text>is securely affixed to the equipment control panel,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is displayed in two contrasting colours,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>is clearly visible and readily discernable from a distance of 1 m,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>has no outer dimension that is less than 2 cm,</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>bears the words “CAUTION: X-RAYS — ATTENTION : RAYONS X”, and</Text></Subparagraph><Subparagraph><Label>(vi)</Label><Text>conforms to the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “CAUTION: X-RAYS — ATTENTION : RAYONS X”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef001.png" /></ImageGroup></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a warning sign that</Text><Subparagraph><Label>(i)</Label><Text>is clearly visible and legible to the operator,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>indicates the possibility of hazardous radiation emission when the equipment is in operation, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>states that any unauthorized use is prohibited;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>on the external surface of the equipment control panel, a clearly visible and readily discernable permanent mark or label that indicates, with respect to the equipment,</Text><Subparagraph><Label>(i)</Label><Text>the name of the manufacturer,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the model designation,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the serial number,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the date of manufacture, and</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>the country of manufacture; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>on the external surface of the X-ray tube assembly, a clearly visible and readily discernable permanent mark or label that indicates, with respect to the X-ray tube assembly,</Text><Subparagraph><Label>(i)</Label><Text>the name of the manufacturer,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the model designation,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the serial number,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the date of installation of the X-ray tube in the X-ray tube housing, and</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>the country of manufacture.</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Dental X-ray equipment with an extra-oral source shall be accompanied by the following materials, which shall be furnished by the manufacturer:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>operating instructions that provide the information necessary for the safe and proper operation of the equipment; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the following information respecting the functioning of the equipment:</Text><Subparagraph><Label>(i)</Label><Text>the maximum allowable deviation from the specified X-ray tube current and voltage,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the accuracy of the controlling timer, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the specific conditions on which the information referred to in subparagraphs (i) and (ii) is based.</Text></Subparagraph></Paragraph></Subsection></Section><Heading level="1"><Label>PART III</Label><TitleText>MICROWAVE OVENS</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part and in items 3 of Schedule I,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>cavity</DefinedTermEn> means a structure that encloses and confines a microwave field; (<DefinedTermFr>cavité</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>commercial microwave oven</DefinedTermEn> means a microwave oven for use</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>in a restaurant, cafeteria or other commercial establishment,</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>in an industrial establishment, or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>in or with a vending machine; (<DefinedTermFr>four à micro-ondes commercial</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>control panel</DefinedTermEn> means the portion of the external surface of a microwave oven on which the user controls are mounted; (<DefinedTermFr>panneau de commande</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>conveyor</DefinedTermEn> means a device that transports material into or within a cavity; (<DefinedTermFr>convoyeur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>door</DefinedTermEn>, with respect to a cavity, means a movable or removable structure that in the closed position is designed to prevent access to the cavity; (<DefinedTermFr>porte</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>effective aperture</DefinedTermEn><Repealed>[Revoked, SOR/84-930, s. 1]</Repealed></Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>external surface</DefinedTermEn>, with respect to a microwave oven, means the outside surface of the cabinet or other enclosure of the oven and includes the plane of any exit or entry port for conveyorized ovens; (<DefinedTermFr>surface externe</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>interlock</DefinedTermEn> means a component or set of components that prevents the generation of microwave power when access to a cavity is possible; (<DefinedTermFr>enclenchement</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>leakage radiation</DefinedTermEn> means any radiation transmitted outside the external surface; (<DefinedTermFr>rayonnement de fuite</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>microwave</DefinedTermEn> means an electromagnetic wave with frequency in the range 0.010 GHz to 300 GHz; (<DefinedTermFr>micro-onde</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>outer enclosure</DefinedTermEn> means a metal or plastic cover that encloses the mechanical and electronic parts of a microwave oven that, under normal conditions of use, are not accessible to the user of the oven; (<DefinedTermFr>enceinte externe</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>response time</DefinedTermEn> means the time period in which a radiation meter indicator reaches ninety per cent of its final steady state reading when subjected to a stepped input signal; (<DefinedTermFr>temps de réponse</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>service control</DefinedTermEn> means a control that is provided by the manufacturer for the purpose of adjustment of the microwave oven and that, under normal conditions of use, is not accessible to the user of the oven; (<DefinedTermFr>commande interne</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>stirrer</DefinedTermEn> means the structure designed to distribute the microwave energy within a cavity; (<DefinedTermFr>agitateur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>user control</DefinedTermEn> means a control that is provided by the manufacturer for the purpose of operation of the microwave oven and that, under normal conditions of use, is accessible to the user of the oven; (<DefinedTermFr>commande externe</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>water equivalent</DefinedTermEn><Repealed>[Revoked, SOR/79-920, s. 1]</Repealed></Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>waveguide</DefinedTermEn> means a metal tube or duct for transmitting microwave energy. (<DefinedTermFr>guide d’ondes</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Subsection><Label>(1)</Label><Text>Every microwave oven shall be designed and constructed in such a manner that, under the conditions of use specified by the manufacturer, it functions in accordance with section 4 with its original components or replacement components recommended by the manufacturer for at least</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>200 000 use cycles or openings and closings of the oven door, in the case of a commercial microwave oven; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>100 000 use cycles or openings and closings of the oven door, in the case of any other microwave oven.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Every microwave oven shall be designed and constructed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>for each microwave power source, a device or indicator that provides a visible indication of the status of operation of the oven;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a device to monitor one or more of the interlocks required by paragraph (<Emphasis style="italic">g</Emphasis>) that renders the oven inoperable when a monitored interlock fails or is otherwise rendered inoperable;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>where the power can be varied by a user control, an indicator to show the level of microwave power applied to the cavity;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>where a total microwave power generating capacity of 25 kilowatts or more is used, a lock on the control panel requiring the insertion of a key before microwave power can be generated;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>where access to the cavity is not by a conveyor, a door constructed and positioned so as to ensure that any leakage radiation does not exceed the limits prescribed by section 4;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>a covering or baffle arrangement over any viewing screen, vent or access port in the cavity wall, other than any opening through which conveyor borne material enters or leaves the cavity, that prevents the insertion of any object into the cavity while the microwave power source is in operation;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>where the oven is equipped with a door as specified in paragraph (<Emphasis style="italic">e</Emphasis>), at least two electrically and mechanically independent interlocks positioned so as to ensure that</Text><Subparagraph><Label>(i)</Label><Text>the door cannot be opened until the microwave power generating component has been turned off, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the microwave power generating component cannot be turned on while the door is open; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>components and shields constructed and positioned so that adjustments to the service controls and user controls to yield maximum possible output do not produce leakage radiation in excess of the limits prescribed by section 4.</Text></Paragraph></Subsection><Subsection><Label>(3)</Label><Text>Every microwave oven shall have permanently affixed to and clearly visible on its external surface the following information and warning sign:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the name of the manufacturer and the model number, serial number and place of manufacture of the oven;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the type of microwave power generating component and the normal operating voltage, operating frequency and normal maximum output power thereof;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a description of the test load prescribed by paragraph 4(3)(<Emphasis style="italic">a</Emphasis>);</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the year and month of manufacture of the oven;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>the sign described in section 3; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>where the oven is not a commercial microwave oven, the words “NOT FOR COMMERCIAL USE — NON DESTINÉ À UN USAGE COMMERCIAL”.</Text></Paragraph></Subsection><Subsection><Label>(4)</Label><Text>Where a microwave oven is equipped with a conveyorized system, a warning sign described in section 3 shall be permanently affixed to its external surface adjacent to each entry and exit port.</Text></Subsection><Subsection><Label>(5)</Label><Text>Where the generation of X-rays within a microwave in excess of 2.5 milliroentgens per hour averaged over 10 square centimetres is possible while the oven is functioning in accordance with subsection 4(1), an X-radiation warning sign that is clearly visible while the microwave oven is being serviced shall be permanently affixed to the microwave power generating component.</Text></Subsection></Section><Heading level="2"><TitleText>Warning Sign Specifications</TitleText></Heading><Section><Label>3.</Label><Text>The warning sign referred to in subsections 2(3) and (4) is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than 2 centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION — MICROWAVES” and “ATTENTION — MICRO-ONDES”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>A Warning sign, bearing the words “Caution - Microwaves”, and “Attention - Micro-ondes”, described by an inverted triangle containing a satellite dish emitting a series of lines</AlternateText><Image source="CRC1370img\CRC1370_ef002.png" /></ImageGroup></Paragraph></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>4.</Label><Subsection><Label>(1)</Label><Text>Every microwave oven shall, when fully assembled and operating with its service controls and user controls adjusted to yield the maximum output, function in such a manner that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the leakage radiation, measured with the instrument prescribed by paragraph (3)(<Emphasis style="italic">b</Emphasis>), at all points at least 5 cm from the external surface of the oven, does not exceed</Text><Subparagraph><Label>(i)</Label><Text>1.0 mW/cm<Sup>2</Sup> with the test load prescribed by paragraph (3)(<Emphasis style="italic">a</Emphasis>) placed</Text><Clause><Label>(A)</Label><Text>in the centre of the shelf in the cavity, in the case of an oven that is designed for cooking and that has a total microwave power generating capacity not greater than 1.5 kW, and</Text></Clause><Clause><Label>(B)</Label><Text>as specified by the manufacturer, in the case of an oven other than an oven described in clause (A), and</Text></Clause></Subparagraph><Subparagraph><Label>(ii)</Label><Text>5.0 mW/cm<Sup>2</Sup> without a test load, where the oven is operable in such conditions; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the intensity of X-ray exposure, at 5 cm from the external surface of the oven, does not exceed 0.5 mR per hour spread over an area of 10 cm<Sup>2</Sup>.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Every microwave oven shall, when the outer enclosure is removed and it is operating with its service controls and user controls adjusted to yield the maximum output, function in such a manner that the leakage radiation, measured with the instrument prescribed by paragraph (3)(<Emphasis style="italic">b</Emphasis>) and with the test load prescribed by paragraph (3)(<Emphasis style="italic">a</Emphasis>) in the cavity, at all points at least 5 cm from every mechanical or electronic part of the oven that is accessible to the user of the oven including, but not limited to, the waveguide, cavity, cavity seam, magnetron and magnetron to waveguide connection, does not exceed 5.0 mW/cm<Sup>2</Sup>.</Text></Subsection><Subsection><Label>(3)</Label><Text>For the purposes of subsections (1) and (2),</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the test load shall be</Text><Subparagraph><Label>(i)</Label><Text>in the case of an oven that is designed for cooking and that has a total microwave power generating capacity not greater than 1.5 kW, 275 ± 15 ml of water at an initial temperature of 20 ± 5°C, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>in the case of an oven other than an oven described in subparagraph (i), the substance and amount thereof specified by the manufacturer as the load to be used for testing the oven; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the instrument used to measure leakage radiation shall</Text><Subparagraph><Label>(i)</Label><Text>be capable of measuring a power density of 1.0 mW/cm<Sup>2</Sup> with an accuracy of 2 dB or better, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>have an indicator with a response time not greater than 3 seconds.</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(4)</Label><Text>Failure of any single component of a microwave oven shall not cause the interlock system to be inoperative.</Text></Subsection><Subsection><Label>(5)</Label><Text>The device required by paragraph 2(2)(<Emphasis style="italic">a</Emphasis>) shall have a rated lifetime that is not less than 5,000 hours.</Text></Subsection><Subsection><Label>(6)</Label><Text>Each interlock required by paragraph 2(2)(<Emphasis style="italic">g</Emphasis>) shall have a rated lifetime that is not less than</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>200 000 on-off cycles, in the case of a commercial microwave oven; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>100 000 on-off cycles, in the case of any other microwave oven.</Text></Paragraph></Subsection></Section><Heading level="1"><Label>PART IV</Label><TitleText>BAGGAGE INSPECTION X-RAY DEVICES</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>detector</DefinedTermEn> means the image receptor or other devices that interacts with the X-rays to produce a signal corresponding to the intensity of the X-rays incident on it; (<DefinedTermFr>détecteur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>model designation</DefinedTermEn> means any combination of letters or figures or both letters and figures by which a device that bears that designation is identified as having characteristics and design features that are uniform; (<DefinedTermFr>désignation du modèle</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>primary X-ray beam</DefinedTermEn> means that X-radiation emitted directly from the target of the X-ray tube and emerging through the window of the X-ray generator; (<DefinedTermFr>faisceau primaire de rayons X</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>X-ray generator</DefinedTermEn> means an assembly of components, including an X-ray tube and its housing and shielding, designed and constructed for the controlled generation of X-rays. (<DefinedTermFr>producteur de rayons X</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Subsection><Label>(1)</Label><Text>A baggage inspection X-ray device shall be designed and constructed in such a way that it functions in accordance with the standards of functioning described in section 3 for as long as the device has its original components or has replacement components recommended by the manufacturer.</Text></Subsection><Subsection><Label>(2)</Label><Text>Notwithstanding subsection (1), a baggage inspection X-ray device shall have sufficient shielding to enable the device to comply with the standards of functioning described in section 3.</Text></Subsection><Subsection><Label>(3)</Label><Text>A baggage inspection X-ray device shall be designed and constructed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>doors or panels over all access openings that are designed for insertion or removal of baggage, unless the devise is designed to prevent the insertion of any part of the human body into the primary X-ray beam through those access openings;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>interlocking of all doors or panels referred to in paragraph (<Emphasis style="italic">a</Emphasis>), with not less than two independent safety interlocks so that, if any of those doors or panels are opened, X-rays cannot be generated;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>interlocking of all doors or panels that allow access to areas where the exposure to X-rays may exceed the level specified in section 3, except those doors or panels referred to in paragraph (<Emphasis style="italic">a</Emphasis>) so that, if any of those doors or panels are opened or removed, X-rays cannot be generated;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>separate warning lights or other indicators</Text><Subparagraph><Label>(i)</Label><Text>that clearly indicate to the operator when the device is powered and when X-rays are being generated,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>designed so that if a pulsed X-ray system is used, the X-rays “ON” lights or indicators remain on for at least 1/2 second, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>that either</Text><Clause><Label>(A)</Label><Text>contain built-in duplication so that the requirements of subparagraphs (i) and (ii) are met, if one of the redundant components fails, or</Text></Clause><Clause><Label>(B)</Label><Text>are interlocked so that, if they malfunction, X-rays cannot be generated;</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>a lock that requires the insertion of a key before X-rays can be produced and that terminates the exposure when the key is removed;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>subject to subsection 2(4), a control or controls to initiate the generation of X-rays, requiring separate operator action for each exposure;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>the automatic alignment of the X-ray generator with the detector when the device is assembled;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>an X-radiation warning sign described in section 4 that</Text><Subparagraph><Label>(i)</Label><Text>is readily discernible and in clear view of the operator, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is permanently affixed to the device at all access openings where baggage is inserted or removed;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>a permanent mark or label that is readily discernible and clearly visible on</Text><Subparagraph><Label>(i)</Label><Text>the external surface of the device, under normal conditions of use, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>on the external surface of the X-ray generator, under conditions of servicing,</Text></Subparagraph><ContinuedParagraph><Text>and that will identify the manufacturer, model designation, serial number, and date and place of manufacture of the device.</Text></ContinuedParagraph></Paragraph></Subsection><Subsection><Label>(4)</Label><Text>Paragraph (3)(<Emphasis style="italic">f</Emphasis>) does not apply to a baggage inspection X-ray device that contains a conveyor system if the X-ray exposure, or a sequence of X-ray exposures, is initiated automatically and the device</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>contains a photocell or other baggage sensing device that initiates X-ray exposures or a sequence of X-ray exposures automatically; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is designed and constructed to include the following additional safety features:</Text><Subparagraph><Label>(i)</Label><Text>a control or switch of a type that</Text><Clause><Label>(A)</Label><Text>requires continuous pressure by the operator to maintain the automatic operation of the device, and</Text></Clause><Clause><Label>(B)</Label><Text>stops the conveyor and terminates the X-ray exposure, or sequence of X-ray exposures, when released, and</Text></Clause></Subparagraph><Subparagraph><Label>(ii)</Label><Text>a conveyor of sufficient length to prevent insertion of any part of the human body into any area where the exposure to X-rays exceeds the level specified in section 3.</Text></Subparagraph></Paragraph></Subsection></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>3.</Label><Text>All baggage inspection X-ray devices shall function in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>at the maximum possible baggage handling rate specified by the manufacturer for the device, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>under all operating conditions of X-ray generation,</Text></Paragraph><ContinuedSectionSubsection><Text>the average exposure rate of X-rays, averaged over a period that is not less than 5 minutes, to an object having a 10 square centimetre cross section and centered at 5 centimetres, from any accessible external surface of the device or from the imaginary plane surface that is drawn to close openings of the device, where baggage is inserted or removed, does not exceed 0.5 milliroentgen per hour.</Text></ContinuedSectionSubsection></Section><Heading level="2"><TitleText>Warning Sign</TitleText></Heading><Section><Label>4.</Label><Text>The X-radiation warning sign referred to in paragraph 2(3)(<Emphasis style="italic">h</Emphasis>) is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than 2 centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION — X-RAYS” and “ATTENTION — RAYONS X”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “CAUTION: X-RAYS — ATTENTION : RAYONS X”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef003.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART V</Label><TitleText>DEMONSTRATION-TYPE GAS DISCHARGE DEVICES</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>cabinet</DefinedTermEn> of a device means a structure that encloses and confines the X-ray source of the device and the material to be irradiated in the device; (<DefinedTermFr>coffret</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>device</DefinedTermEn> means a demonstration-type gas discharge device; (<DefinedTermFr>dispositif</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>gas discharge tube</DefinedTermEn> means an electronic tube in which glow discharges or X-rays or both may be produced by the acceleration of electrons or ions; (<DefinedTermFr>tube à décharge</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>model designation</DefinedTermEn> means any combination of letters or figures or both letters and figures by which a device that bears that designation is claimed to have characteristics and design features that are uniform. (<DefinedTermFr>désignation du modèle</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Text>Every device that contains a gas discharge tube not specifically designed to generate X-rays shall be designed and constructed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a permanent mark or label that</Text><Subparagraph><Label>(i)</Label><Text>is clearly visible under conditions of normal use,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>identifies the device by setting out the name of the manufacturer, model designation, and date and place of manufacture,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>indicates the intended polarity of each terminal of the device, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>indicates the power supply or maximum voltage to be used with the device; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>shielding that</Text><Subparagraph><Label>(i)</Label><Text>is sufficient to enable the device to comply with the standards of functioning set out in section 5, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is either non-removable or is so constructed that its removal renders the device inoperable.</Text></Subparagraph></Paragraph></Section><Section><Label>3.</Label><Text>Subject to section 4, where a device contains a gas discharge tube specifically designed to generate X-rays, that tube shall be enclosed in a cabinet that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>has permanently affixed to its external surface</Text><Subparagraph><Label>(i)</Label><Text>the X-radiation warning sign described in section 6,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>a readily discernible mark or label that</Text><Clause><Label>(A)</Label><Text>identifies the manufacturer, model designation and date and place of manufacture of the device, and</Text></Clause><Clause><Label>(B)</Label><Text>warns that the tube generates X-rays when energized, and</Text></Clause></Subparagraph><Subparagraph><Label>(iii)</Label><Text>where the power source for the device is not an integral part of the device, a mark or label that</Text><Clause><Label>(A)</Label><Text>is clearly visible under normal conditions of use of the device, and</Text></Clause><Clause><Label>(B)</Label><Text>indicates the intended polarity of the terminals of the device and the power supply or maximum voltage to be used with the device;</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is provided with</Text><Subparagraph><Label>(i)</Label><Text>two warning lights that are in clear view of the operator of the device, one of which indicates when the device is powered and the other of which indicates when X-rays are being generated, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>shielding that</Text><Clause><Label>(A)</Label><Text>is sufficient to enable the device to comply with section 5, and</Text></Clause><Clause><Label>(B)</Label><Text>is not removable or is so constructed that its removal renders the device inoperable; and</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>is constructed so that X-rays cannot be generated by the device when any door or panel in the cabinet that allows access to its interior is opened or removed.</Text></Paragraph></Section><Section><Label>4.</Label><Subsection><Label>(1)</Label><Text>Where the power source is not an integral part of the device, the warning light required by subparagraph 3(<Emphasis style="italic">b</Emphasis>)(i) to indicate when the device is powered shall be located on the power source.</Text></Subsection><Subsection><Label>(2)</Label><Text>The device referred to in section 3 shall be constructed so that, where the warning light required by subparagraph 3(<Emphasis style="italic">b</Emphasis>)(i) to indicate when X-rays are being generated fails or malfunctions,</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a duplicate warning light operates when X-rays are being generated; or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>X-rays cannot be generated by the device.</Text></Paragraph></Subsection></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>5.</Label><Text>Every device shall function in such a way that the emission of X-rays therefrom, under all possible conditions of operation and for as long as the device has its original components or has replacement components recommended by the manufacturer, is such that the average exposure rate of X-rays to an object having a 10 square centimetre cross section and centred at 5 centimetres from any accessible external surface of the device does not exceed 0.5 milliroentgen per hour.</Text></Section><Heading level="2"><TitleText>Warning Sign</TitleText></Heading><Section><Label>6.</Label><Text>The X-radiation warning sign referred to in subparagraph (3)(<Emphasis style="italic">a</Emphasis>)(i) is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than 2 centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION — X-RAYS” and “ATTENTION — RAYONS X”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “CAUTION: X-RAYS — ATTENTION : RAYONS X ”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef004.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART VI</Label><TitleText>PHOTOFLUOROGRAPHIC X-RAY EQUIPMENT</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>attenuation</DefinedTermEn> means a decrease in radiation intensity caused by absorption and scattering in a medium; (<DefinedTermFr>atténuation</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>coefficient of variation</DefinedTermEn> means the ratio of the standard deviation to the mean value of a series of exposure measurements calculated by using the following equation:</Text><FormulaGroup><Formula><ImageGroup position="center"><AlternateText>The coefficient of variation, calculated by the equation “C equals one divided by x bar multiplied by, open square bracket, sum from i equals one to n, open parenthesis, x sub i minus x bar, close parenthesis, squared, divided by n minus one, close square bracket, to the power of one half”</AlternateText><Image source="CRC1370img\CRC1370_e014.png" /></ImageGroup></Formula><FormulaConnector>when</FormulaConnector><FormulaDefinition><FormulaTerm>C</FormulaTerm><Text>= coefficient of variation</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>X<Sub>i</Sub></FormulaTerm><Text>= i<Sup>th</Sup> exposure measurement</Text></FormulaDefinition><FormulaDefinition><FormulaTerm><Emphasis style="overbar">X</Emphasis></FormulaTerm><Text>= mean value of the exposure measurements</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>n</FormulaTerm><Text>= number of exposure measurements;</Text></FormulaDefinition></FormulaGroup></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>collimator</DefinedTermEn> means a device or mechanism that limits the shape and size of the useful beam; (<DefinedTermFr>collimateur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>effective focal spot</DefinedTermEn> means the projection of the focal spot on the plane that is perpendicular to the X-ray beam axis and that passes through the centre of the focal spot; (<DefinedTermFr>tache focale efficace</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>filter</DefinedTermEn> means material placed in the useful beam to attenuate preferentially the lower energy radiations; (<DefinedTermFr>filtre</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>focal spot</DefinedTermEn> means the section at which the anode of an X-ray tube intercepts the electron beam; (<DefinedTermFr>tache focale</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>leakage radiation</DefinedTermEn> means all radiation, except the useful beam, coming from within the housing of an energized X-ray generating tube; (<DefinedTermFr>rayonnement de fuite</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>useful beam</DefinedTermEn> means the radiation passing through the aperture, cone or collimator of the housing of an X-ray generating tube. (<DefinedTermFr>faisceau utile</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design</TitleText></Heading><Section><Label>2.</Label><Text>Photofluorographic X-ray equipment shall be designed in such a way that all controls, meters, lights or other indicators are readily discernible and clearly labelled to indicate their function.</Text></Section><Section><Label>3.</Label><Text>Photofluorographic X-ray equipment shall be designed to include the following features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>on the control panel, the X-radiation warning sign described in section 6;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a warning sign that</Text><Subparagraph><Label>(i)</Label><Text>is in clear view of and readily discernible by the operator,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is permanently affixed to the X-ray control panel,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>indicates the possibility of hazardous radiation emission when the equipment is in operation, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>prohibits unauthorized use;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>electrical meters or other indicators that</Text><Subparagraph><Label>(i)</Label><Text>are in clear view of the operator,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>are permanently affixed to the X-ray control panel, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>show</Text><Clause><Label>(A)</Label><Text>the preset operating kilovoltage of the equipment when used in the phototiming mode,</Text></Clause><Clause><Label>(B)</Label><Text>the preset operating kilovoltage and milliamperage of the equipment when not used in the phototiming mode, and</Text></Clause><Clause><Label>(C)</Label><Text>for battery-operated equipment, the state of charge of the battery;</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>separate aural or visual indicators on the control panel that</Text><Subparagraph><Label>(i)</Label><Text>are clearly discernible by the operator, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>indicate when the X-ray machine is powered and when X-rays are being produced;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>where more than one X-ray tube is controlled by one control panel, it shall not be possible to energize more than one X-ray tube at the same time and there shall be</Text><Subparagraph><Label>(i)</Label><Text>at each X-ray tube housing, a visible indicator that indicates when the tube is connected and ready to be energized,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>at the control panel, a visible indicator that indicates which tube is connected and ready to be energized, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>at the normal position of the operator, visible indicators that provide the information required by subparagraph (<Emphasis style="italic">c</Emphasis>)(iii);</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>an exposure switch on the control panel of a kind that requires continuous pressure by the operator to complete the circuit;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>a timer so designed that</Text><Subparagraph><Label>(i)</Label><Text>when the production of X-rays is automatically terminated after a preset time interval,</Text><Clause><Label>(A)</Label><Text>the preset time interval is clearly indicated,</Text></Clause><Clause><Label>(B)</Label><Text>an exposure cannot be initiated with the timer set to zero or in the OFF position, and</Text></Clause><Clause><Label>(C)</Label><Text>the minimum exposure time is not greater than 1/60 second,</Text></Clause></Subparagraph><Subparagraph><Label>(ii)</Label><Text>when the production of X-rays is automatically terminated after the integrated radiation exposure to a photocell or similar component behind the fluorescent screen is measured, there is incorporated in the timer’s electrical circuitry</Text><Clause><Label>(A)</Label><Text>an aural or visual indicator that indicates clearly to the operator when the phototimer has failed to terminate the exposure, and</Text></Clause><Clause><Label>(B)</Label><Text>a back-up or safety timer, with a maximum setting of 1 second, that will terminate the exposure in the event of phototimer failure, and</Text></Clause></Subparagraph><Subparagraph><Label>(iii)</Label><Text>when the production of X-rays is automatically terminated after a preset milliampere-second interval, the minimum milliampere-second interval is no greater than 1 milliampere-second;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>a collimator that renders the machine inoperative if a part of the useful beam at the plane of the fluorescent screen extends beyond the useful portion of the fluorescent screen and</Text><Subparagraph><Label>(i)</Label><Text>where the photofluorographic X-ray equipment is designed for a variable image receptor size and the target to fluorescent screen distance is variable,</Text><Clause><Label>(A)</Label><Text>the collimator shall be adjustable, and give a rectangular beam,</Text></Clause><Clause><Label>(B)</Label><Text>the alignment of the centre of the useful beam with the centre of the fluorescent screen shall be within two per cent of the target to the fluorescent screen distance,</Text></Clause><Clause><Label>(C)</Label><Text>there shall be an apron (providing attenuation equivalent to 0.25 millimetre lead for a 100 kilovolts (peak) X-ray beam of half value layer 2.7 millimetres of aluminium) on a bracket attached to the bottom of the fluorescent screen mounting that can be swiveled into place to protect the patient’s gonad area from the useful beam,</Text></Clause><Clause><Label>(D)</Label><Text>means shall be provided to ensure that the axis of the X-ray beam can only be perpendicular to the plane of the fluorescent screen, and</Text></Clause><Clause><Label>(E)</Label><Text>a light beam generator shall be incorporated into the collimator assembly</Text><Subclause><Label>(I)</Label><Text>that visually defines the outline of the useful beam, and</Text></Subclause><Subclause><Label>(II)</Label><Text>that does not permit a misalignment of the visually defined field with the X-ray field along either the length or width of the X-ray field to exceed two per cent of the target to fluorescent screen distance, and</Text></Subclause></Clause></Subparagraph><Subparagraph><Label>(ii)</Label><Text>where the photofluoraphic X-ray equipment is designed for a constant image receptor size and the target to fluorescent screen distance is fixed,</Text><Clause><Label>(A)</Label><Text>there shall be a shaped lower leaf or an apron on a bracket as described in clause (<Emphasis style="italic">h</Emphasis>)(i)(C) designed to protect the patient’s gonad area from the useful beam,</Text></Clause><Clause><Label>(B)</Label><Text>the alignment of the centre of the useful beam with the centre of the fluorescent screen shall be within two per cent of the target to fluorescent screen distance, and</Text></Clause><Clause><Label>(C)</Label><Text>means shall be provided to ensure that the axis of the X-ray beam can only be perpendicular to the plane of the fluorescent screen;</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>a filter that</Text><Subparagraph><Label>(i)</Label><Text>is securely installed in the path of the useful beam, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>provides attenuation, including inherent filtration, at least equivalent to that afforded by 2.5 millimetres of aluminium at 100 kilovolts (peak);</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">j</Emphasis>)</Label><Text>a readily visible mark on the X-ray tube housing indicating, to within 2 millimetres, the location on the tube housing of the projection, at right angles to the beam axis, of the position of the focal spot on the target;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">k</Emphasis>)</Label><Text>an effective focal spot size of not greater than 1.5 square millimetres for units designed to operate at not more than 150 kilovolts (peak);</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">l</Emphasis>)</Label><Text>a lock, on the X-ray machine control panel, of a type that requires the insertion of a key before X-rays can be produced and the removal thereof causes termination of the production of X-rays;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">m</Emphasis>)</Label><Text>an appliance or other means that</Text><Subparagraph><Label>(i)</Label><Text>ensure that the machine is not able to be operated if the target to fluorescent screen distance is less than 100 centimetres, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>clearly indicate the target to fluorescent screen distance to within two per cent;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">n</Emphasis>)</Label><Text>means that allow data to be recorded automatically on to each film;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">o</Emphasis>)</Label><Text>an interlock on the film holder mechanism that</Text><Subparagraph><Label>(i)</Label><Text>prevents more than one film exposure being recorded without resetting; and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>prevents the same film frame receiving more than one exposure;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">p</Emphasis>)</Label><Text>means that prevent the use of films of size smaller than 70 millimetres × 70 millimetres; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">q</Emphasis>)</Label><Text>on or near the power-input socket, a statement of power requirements at the maximum line current and, for battery operated units, a statement of charge-use frequency.</Text></Paragraph></Section><Heading level="2"><TitleText>Standards of Construction</TitleText></Heading><Section><Label>4.</Label><Subsection><Label>(1)</Label><Text>Photofluorographic X-ray equipment shall be constructed in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the X-ray tube is securely fixed and correctly aligned within the tube housing;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the X-ray tube head maintains its exposure position without drift, tipping or vibration during operation;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the equipment, under normal conditions of use, functions in accordance with the standards described in section 5 for as long as the X-ray machine has its original components or replacement components recommended by the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>where there is provision for insertion of a grid, there is a holder containing a label that indicates the grid ratio of the grid used;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>mirrors, lenses or other optical components within the camera hood are securely fastened so that optical misalignment cannot occur with normal use;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>the camera hood, containing the whole optical system, does not admit light that can fog the film;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>any material between the film side of the fluorescent screen and the first optical component does not reduce the light intensity or alter the spectrum of the light emitted from the fluorescent screen;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>the front panel at the camera hood, located between the patient and the fluorescent screen, has less than 0.5 millimetre aluminium equivalence measured with an X-ray beam of potential 100 kilovolts (peak) with a half value layer of 2.7 millimetres of aluminium;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>the device bears a replaceable label mounted on the camera hood in clear view of the operator, indicating the type and date of insertion of the fluorescent screen; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">j</Emphasis>)</Label><Text>when the recommended schedule of maintenance for the device is followed, the device will function in accordance with section 5 for the normal lifetime of the device.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Photofluorographic X-ray equipment shall bear, on the tube housing and on the control panel, a permanent mark or label that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is clearly visible; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>carries, with reference to the X-ray machine and X-ray tube, the name of the manufacturer, the model number, the serial number and the date and place of manufacture.</Text></Paragraph></Subsection></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>5.</Label><Text>Photofluorographic X-ray equipment when fully assembled for use shall function under normal conditions of use in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>it is not possible to preset the kilovoltage below 70 kilovolts (peak);</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the preset kilovoltage of the X-ray tube is maintained or is adjustable to within 3 kilovolts when the line voltage supply varies by plus or minus 10 per cent of its nominal value;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>for any combination of kilovoltage, current and time, the coefficient of variation of any 10 consecutive radiation exposures, taken at the same distance within a time period of 1 hour, is no greater than 0.05;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the entrance exposure to a sheet of 99.9 per cent pure copper of dimensions 45 centimetres × 45 centimetres × 1.41 millimetres, placed over the front panel, and phototimed with the exposure taken at 85 kilovolts (peak), does not exceed 50 milliroentgens per exposure to obtain at the centre of the film an optical density above gross fog of not less than 0.5;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>at any fixed kilovoltage, the average ratios of milliroentgen exposure to milliampere-seconds obtained at any two consecutive tube current settings do not differ by more than 0.10 times their sum, namely,</Text><FormulaGroup><Formula><FormulaText><Emphasis style="overbar">X</Emphasis><Sub>1</Sub> - <Emphasis style="overbar">X</Emphasis><Sub>2</Sub> ≤ 0.10(<Emphasis style="overbar">X</Emphasis><Sub>1</Sub> + <Emphasis style="overbar">X</Emphasis><Sub>2</Sub>),</FormulaText></Formula><FormulaConnector>where </FormulaConnector><FormulaDefinition><FormulaTerm><Emphasis style="overbar">X</Emphasis><Sub>1</Sub> and <Emphasis style="overbar">X</Emphasis><Sub>2</Sub></FormulaTerm><Text>are the average milliroentgen per milliampere-second values obtained at each of two consecutive tube current settings;</Text></FormulaDefinition></FormulaGroup></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>the leakage radiation from the X-ray tube assembly, when the machine is operated at any kilovoltage and current within the rating of the X-ray tube, is such that the average exposure rate of X-rays to any object having a 100 square centimetre cross-section and centred at 1 metre from the focal spot of the X-ray tube does not exceed 100 milliroentgens in 1 hour; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>emission of ionizing radiation by any component other than the X-ray tube assembly, when the machine is operated at any kilovoltage and current within the rating of the X-ray tube, is such that the average exposure rate of X-rays to any object having a 10 square centimetre cross-section and centred at 5 centimetres from any accessible surface of the device, does not exceed 2.5 milliroentgens per hour.</Text></Paragraph></Section><Heading level="2"><TitleText>Warning Sign</TitleText></Heading><Section><Label>6.</Label><Text>The X-radiation warning sign referred to in section 3 of this Part is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than 2 centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION X-RAYS” and “ATTENTION RAYONS X”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “CAUTION: X-RAYS — ATTENTION : RAYONS”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef005.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART VII</Label><TitleText>LASER SCANNERS</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part and in item 7 of Schedule I,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>accessible location</DefinedTermEn> means any point that can be reached by any part of the human body; (<DefinedTermFr>point accessible</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>exit aperture</DefinedTermEn> means an opening or window in the protective enclosure of a laser scanner that is designed to allow laser radiation to be transmitted to the outside; (<DefinedTermFr>ouverture de sortie</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>integrated irradiance</DefinedTermEn> means the radiant energy incident per unit area of surface expressed as joules per square centimetre (J cm<Sup>-2</Sup>); (<DefinedTermFr>exposition énergétique (dose)</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>irradiance</DefinedTermEn> means radiant power incident per unit area expressed as watts per square centimetre (W cm<Sup>-2</Sup>); (<DefinedTermFr>éclairement énergétique</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>laser</DefinedTermEn> means any device that can be made to produce light primarily by the process of stimulated emission; (<DefinedTermFr>laser</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>laser radiation</DefinedTermEn> means all electromagnetic radiation generated by a laser that is coherent and propagates collinearly through space; (<DefinedTermFr>rayonnement laser</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>protective enclosure</DefinedTermEn> means a structure that encloses a laser scanner and its accessory components and restricts the emission of laser radiation to one or more exit apertures; (<DefinedTermFr>parois protectrices</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>protective housing</DefinedTermEn> means a structure that encloses the components of a laser and prevents the emission of laser radiation except through an exit aperture; (<DefinedTermFr>logement protecteur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>pulse</DefinedTermEn> means an intermittent emission of laser radiation for a duration of less than 0.25 second; (<DefinedTermFr>impulsion</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>pulse duration</DefinedTermEn> means the time interval measured between the half-peak power points on the leading and trailing edges of a pulse; (<DefinedTermFr>durée d’impulsion</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>safety interlock</DefinedTermEn> means a mechanism that prevents the generation of radiation when any portion of the protective enclosure is removed or displaced; (<DefinedTermFr>enclenchement de sécurité</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>scanned laser radiation</DefinedTermEn> means laser radiation having a time varying direction, origin or pattern of propagation with respect to a stationary frame of reference; (<DefinedTermFr>rayonnement laser balayeur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>service controls</DefinedTermEn> means controls that are provided by the manufacturer for the purpose of adjustment of the laser scanner and that, under normal conditions of use, are not accessible to the user of the laser scanner; (<DefinedTermFr>commandes internes</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>user controls</DefinedTermEn> means controls that are provided by the manufacturer for the purpose of operation of the laser scanner and that, under normal conditions of use, are accessible to the user of the laser scanner; (<DefinedTermFr>commandes externes</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>wavelength</DefinedTermEn> means a wavelength <Language xml:lang="la">in vacuo</Language>. (<DefinedTermFr>longueur d’onde</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Subsection><Label>(1)</Label><Text>Every laser scanner shall be designed and constructed in such a manner that, under the conditions of use specified by the manufacturer, it functions in accordance with section 4 so long as its original components or replacement components recommended by the manufacturer are in use.</Text></Subsection><Subsection><Label>(2)</Label><Text>Every laser scanner shall be designed in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>all marks, labels and signs are permanently affixed and clearly visible; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>all controls, meters, lights or other indicators are readily discernible and clearly labelled to indicate their function.</Text></Paragraph></Subsection><Subsection><Label>(3)</Label><Text>Every laser scanner shall have on its external surface the following information:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the name and address of the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the name and address of the distributor, if the distributor is other than the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the model designation, the serial number and the month and year of manufacture;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>for pulsed lasers, the output power; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>the name and address of the manufacturer of the laser or lasers used in the device.</Text></Paragraph></Subsection><Subsection><Label>(4)</Label><Text>Every laser scanner shall be designed and constructed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a switch or other control by which the laser scanner may be turned ON and OFF;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a protective housing;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a protective enclosure;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>a safety interlock or interlocks; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>if means are provided to defeat or bypass interlocks for maintenance purposes,</Text><Subparagraph><Label>(i)</Label><Text>a visual or aural indication when any interlock is defeated or bypassed, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the replacement of any removed or displaced portion of the protective enclosure is not possible when the interlock or interlocks are defeated or bypassed.</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(5)</Label><Text>Every laser scanner shall be designed and constructed in such a way that failure or malfunction of any component of the scanner does not result in leakage of laser radiation in excess of the limits specified in section 4.</Text></Subsection><Subsection><Label>(6)</Label><Text>Every laser scanner shall have the laser radiation warning sign described in section 5 permanently affixed to appropriate surfaces inside the scanner so as to be clearly visible under conditions of removal or displacement of each removable or displaceable portion of the protective enclosure.</Text></Subsection></Section><Section><Label>3.</Label><Text>Every laser scanner shall be equipped with</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>an operation manual that contains instructions for</Text><Subparagraph><Label>(i)</Label><Text>the installation,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the operation, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the detection of any malfunction of the laser scanner; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a servicing manual that contains</Text><Subparagraph><Label>(i)</Label><Text>details of the electronic and mechanical control systems,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>instructions for service adjustments and service procedures including warnings or precautions to be taken to avoid possible exposure to laser radiation or other electromagnetic radiations, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>a schedule of maintenance requirements that, if followed, will maintain the safety features indicated in paragraphs 2(4)(<Emphasis style="italic">a</Emphasis>) to (<Emphasis style="italic">e</Emphasis>) and keep the scanner functioning in accordance with section 4 during the normal operation and normal lifetime of the scanner.</Text></Subparagraph></Paragraph></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>4.</Label><Subsection><Label>(1)</Label><Text>Every laser scanner, when fully assembled and operating with its service controls and user controls adjusted to yield the maximum emission, shall function in such a manner that the intensity of laser radiation at all accessible locations, when measured within a stationary circular area of 0.385 square centimetres and averaged over that area does not exceed the following limits:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>during any time interval of less than 1.8 × 10<Sup>-5</Sup> seconds, an integrated irradiance of 5.0 × 10<Sup>-7</Sup> joules per square centimetre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>during any time interval, t seconds, that is greater than 1.8 × 10<Sup>-5</Sup> seconds but less than or equal to 10 seconds, an integrated irradiance of 1.8 × 10<Sup>-3</Sup> t<Sup>0.75</Sup> joules per square centimetre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>during any time interval of greater than 10 seconds but less than or equal to 1.0 × 10<Sup>-4</Sup> seconds, an integrated irradiance of 1.0 × 10<Sup>-2</Sup> joules per square centimetre; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>during any time interval of greater than 1.0 × 10<Sup>-4</Sup> seconds, an irradiance of 1.0 × 10<Sup>-6</Sup> watts per square centimetre.</Text></Paragraph></Subsection></Section><Heading level="2"><TitleText>Warning Sign Specifications</TitleText></Heading><Section><Label>5.</Label><Text>The laser radiation warning sign referred to in subsection 2(6) is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimension less than 2 centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION — HAZARDOUS LASER AND ELECTROMAGNETIC RADIATION WHEN OPEN AND INTERLOCK DEFEATED” — “ATTENTION — RAYONNEMENT LASER ET ELECTROMAGNÉTIQUE DANGEREUX SI OUVERT AVEC L’ENCLENCHEMENT DE SÉCURITÉ ANNULÉ”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “Caution - Hazardous laser and electromagnetic radiation when open and interlock defeated” — “Attention - Rayonnement laser et electromagnétique dangereux si ouvert avec l’enclenchement de sécurité annulé”, consisting of an inverted triangle containing a spark with a line attached going to the top right corner.</AlternateText><Image source="CRC1370img\CRC1370_ef006.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART VIII</Label><TitleText>DEMONSTRATION LASERS</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part and in item 8 of Schedule I,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>beam</DefinedTermEn> means a collection of rays that may be parallel, convergent or divergent; (<DefinedTermFr>faisceau</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>diameter</DefinedTermEn>, with respect to a laser beam, means the minimum diameter of a circular aperture that, when placed to intercept the beam with the plane of the circular aperture perpendicular to the direction of propagation of the beam, will permit 0.865 of the total beam power to be transmitted; (<DefinedTermFr>diamètre</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>divergence</DefinedTermEn>, with respect to a laser beam, means the full angle of spread of the beam; (<DefinedTermFr>divergence</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>exit aperture</DefinedTermEn> means an opening or window in the protective housing of a demonstration laser that is designed to allow laser radiation to be transmitted to the outside; (<DefinedTermFr>ouverture de sortie</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>kit</DefinedTermEn>, with respect to a demonstration laser, means a set of components that, when assembled in accordance with the manufacturer’s instructions, will result in a demonstration laser; (<DefinedTermFr>ensemble en pièces détachées</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>laser</DefinedTermEn> means any device that can be made to produce light primarily by the process of stimulated emission; (<DefinedTermFr>laser</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>laser radiation</DefinedTermEn> means all electromagnetic radiation generated by a laser that is coherent and propagates collinearly through space; (<DefinedTermFr>rayonnement laser</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>lasing medium</DefinedTermEn> means a material that emits laser radiation by virtue of stimulated transitions between specific electronic or molecular energy levels; (<DefinedTermFr>matière active</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>protective housing</DefinedTermEn> means a structure that encloses the components of a demonstration laser and prevents the emission of laser radiation except through an exit aperture; (<DefinedTermFr>logement protecteur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>shutter</DefinedTermEn> means a mechanism that, in its closed position, intercepts the laser beam and prevents the emission of laser radiation from the demonstration laser; (<DefinedTermFr>obturateur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>wavelength</DefinedTermEn> means a wavelength <Language xml:lang="la">in vacuo</Language>. (<DefinedTermFr>longueur d’onde</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Subsection><Label>(1)</Label><Text>Every demonstration laser shall be designed and constructed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a switch or control by which the device may be turned ON and OFF;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a protective housing the removal of which renders the laser inoperable;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>an indicator, other than the laser beam, that indicates visually or aurally when</Text><Subparagraph><Label>(i)</Label><Text>the laser is in operation, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the shutter required by paragraph (<Emphasis style="italic">d</Emphasis>) is open; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>a permanently attached shutter.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Every demonstration laser shall have permanently affixed and clearly visible on its external surface the following information:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the name and address of the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the model designation, serial number, and the year and month of manufacture;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the name and address of the distributor, if the distributor is other than the manufacturer of the device;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the type of lasing medium;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>the maximum laser radiation output power, in milliwatts;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>the wavelength of the laser radiation, in nanometres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>the diameter of the laser beam at the exit aperture, in millimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>the divergence of the laser beam, in milliradians; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>the laser radiation warning sign described in section 5.</Text></Paragraph></Subsection></Section><Section><Label>3.</Label><Text>Every demonstration laser shall be equipped with</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>instructions for the safe operation and use of the device including warnings or precautions to be taken to avoid hazardous exposure to laser radiation;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a schedule of maintenance requirements that, if followed, will maintain the safety features indicated in paragraphs 2(1)(<Emphasis style="italic">a</Emphasis>) to (<Emphasis style="italic">d</Emphasis>) in a good state of repair and keep the device functioning in accordance with section 4; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>where the device is in the form of a kit, a set of instructions setting out the step-by-step procedure for assembly and testing to ensure that all the requirements of these Regulations are met.</Text></Paragraph></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>4.</Label><Subsection><Label>(1)</Label><Text>Every demonstration laser shall function in such a way that the laser radiation emission</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>does not exceed 1 milliwatt under all possible conditions of use; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is within the wavelength range from 400 nanometres to 780 nanometres.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>For the purpose of subsection (1), the instrument used for the measurement of laser radiation emission shall have a circular aperture with a diameter of at least 7 millimetres.</Text></Subsection></Section><Heading level="2"><TitleText>Warning Sign Specifications</TitleText></Heading><Section><Label>5.</Label><Text>For the purpose of paragraph 2(2)(<Emphasis style="italic">i</Emphasis>), the laser radiation warning sign described in this section is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than 2 centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION LASER, TO AVOID EYE DAMAGE DO NOT LOOK INTO BEAM — ATTENTION LASER, POUR ÉVITER DES DOMMAGES AUX YEUX NE PAS REGARDER DANS LE FAISCEAU”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “CAUTION LASER, TO AVOID EYE DAMAGE DO NOT LOOK INTO BEAM — ATTENTION LASER, POUR ÉVITER DES DOMMAGES AUX YEUX NE PAS REGARDER DANS LE FAISCEAU”, consisting of an inverted triangle containing a spark with a line attached going to the top right corner.</AlternateText><Image source="CRC1370img\CRC1370_ef007.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART IX</Label><TitleText>LOW ENERGY ELECTRON MICROSCOPES</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>service control</DefinedTermEn> means a control that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is installed by the manufacturer on a low energy electron microscope or on any accessory thereof for the purpose of adjustment, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>under normal conditions of use, is not accessible to the user; (<DefinedTermFr>commande interne</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>user control</DefinedTermEn> means a control that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is provided by the manufacturer on a low energy electron microscope or on any accessory thereof for the purpose of adjustment, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>in the case of a fully assembled low energy electron microscope under normal conditions of use, is accessible to the user. (<DefinedTermFr>commande externe</DefinedTermFr>)</Text></Paragraph></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Text>Every low energy electron microscope shall be designed in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>all marks, labels and signs are permanently affixed and clearly visible;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>all user controls, meters, lights or other indicators are readily discernible and clearly labelled as to function; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>it bears on the external surface</Text><Subparagraph><Label>(i)</Label><Text>a mark or label that sets out with respect to the device</Text><Clause><Label>(A)</Label><Text>the name and address of the manufacturer,</Text></Clause><Clause><Label>(B)</Label><Text>the model designation,</Text></Clause><Clause><Label>(C)</Label><Text>the serial number, and</Text></Clause><Clause><Label>(D)</Label><Text>the date of manufacture, and</Text></Clause></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the X-radiation warning sign described in section 6.</Text></Subparagraph></Paragraph></Section><Section><Label>3.</Label><Text>Every low energy electron microscope shall be designed and constructed in such a way that it functions in accordance with the standards of functioning described in section 5 for as long as the device has its original components or has replacement components recommended by the manufacturer.</Text></Section><Section><Label>4.</Label><Text>Every low energy electron microscope shall be designed and constructed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>shielding, if necessary, that</Text><Subparagraph><Label>(i)</Label><Text>is sufficient to enable the device to comply with the standards of functioning set out in section 5, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is either non-removable or is so constructed or connected that its removal renders the device inoperable;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>any door or panel that allows access to areas where the exposure to X-rays may exceed the level specified in section 5 is connected in such a way that, if that door or panel is opened or removed, the device cannot function;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>accessory ports, if any, are provided with shielded covers that are interlocked in such a way that, if a cover is removed, the device cannot function unless the accessory for which the port is intended is in place;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>any accessory provided with or intended for use with the device is shielded in accordance with paragraph (<Emphasis style="italic">a</Emphasis>).</Text></Paragraph></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>5.</Label><Text>Every low energy electron microscope shall function in such a way that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>at all settings of the user controls and service controls, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>under all possible modes of operation,</Text></Paragraph><ContinuedSectionSubsection><Text>the exposure rate of X-rays to an object having a 10 square centimetre cross section and centred at 5 centimetres from any accessible external surface of</Text></ContinuedSectionSubsection><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the device, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>any accessory provided for use with the device,</Text></Paragraph><ContinuedSectionSubsection><Text>does not exceed 0.5 milliroentgen per hour.</Text></ContinuedSectionSubsection></Section><Heading level="2"><TitleText>Warning Sign Specification</TitleText></Heading><Section><Label>6.</Label><Text>The X-radiation warning sign referred to in paragraph 2(<Emphasis style="italic">c</Emphasis>) is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than 2 centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION, X-RAYS” and “ATTENTION, RAYONS X”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “Caution: X-Rays — Attention: Rayons X”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef008.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART X</Label><TitleText>HIGH INTENSITY MERCURY VAPOUR DISCHARGE LAMPS</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>cumulative operating time</DefinedTermEn> means the sum of the periods of time during which electric current passes through the high-pressure arc discharge tube of a lamp; (<DefinedTermFr>durée cumulée de fonctionnement</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>lamp</DefinedTermEn> means any high intensity mercury vapour discharge lamp; (<DefinedTermFr>lampe</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>lamp container</DefinedTermEn> means a carton, wrapper or other package in which a lamp is shipped, stored or displayed, and that bears information concerning the identification or use of the lamp it contains; (<DefinedTermFr>conteneur de lampe</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>outer envelope</DefinedTermEn> means the exterior shell, usually glass, surrounding a high-pressure arc discharge tube that attenuates the emission of shortwave ultraviolet radiation when intact; (<DefinedTermFr>enveloppe extérieure</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>self-extinguishing lamp</DefinedTermEn> means a lamp</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>manufactured after March 7, 1980, that, when tested in accordance with section 6, ceases operation within a cumulative operating time not exceeding 15 minutes following the puncturing or the complete breakage of the outer envelope except where fragments of the breakage extend 50 mm or less from the base shell, or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>manufactured after September 7, 1981, that, when tested in accordance with section 6, ceases operation within a cumulative operating time not exceeding 15 minutes following the puncturing of the outer envelope or the breakage of at least 3 square cm of the outer envelope; (<DefinedTermFr>lampe à auto-extinction</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>shortwave ultraviolet radiation</DefinedTermEn> means ultraviolet radiation having wavelengths shorter than 320 nanometres. (<DefinedTermFr>rayonnement ultraviolet court</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Text>Every lamp shall have the name of the manufacturer and the month and year of its manufacture</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>imprinted thereon, or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>set out in a label permanently attached thereto</Text></Paragraph><ContinuedSectionSubsection><Text>in such a manner that the information is legible whether the lamp is intact or its outer envelope is broken or removed.</Text></ContinuedSectionSubsection></Section><Section><Label>3.</Label><Text>Every self-extinguishing lamp shall bear the letter “T” clearly marked on its outer envelope and on its base.</Text></Section><Section><Label>4.</Label><Text>Every lamp other than a self-extinguishing lamp shall bear the letter “R” clearly marked on its outer envelope and on its base.</Text></Section><Section><Label>5.</Label><Subsection><Label>(1)</Label><Text>Every lamp container of a self-extinguishing lamp shall bear, in a clear and prominent manner,</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the letter “T”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a statement in English and in French to the following effect:</Text><Provision format-ref="indent-1-1" language-align="no" list-item="no"><Text>“This lamp should self-extinguish within 15 minutes after the outer envelope is broken or punctured. If such damage to the outer envelope occurs, TURN OFF AND REMOVE LAMP to avoid possible injury from hazardous shortwave ultraviolet radiation.”.</Text></Provision></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Every lamp container of a lamp other than a self-extinguishing lamp shall bear, in a clear and prominent manner,</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the letter “R”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the words “WARNING — MISE EN GARDE” followed by a statement in English and French to the following effect:</Text><Provision format-ref="indent-1-1" language-align="no" list-item="no"><Text>“This lamp can cause serious skin burn and eye inflammation from shortwave ultraviolet radiation if the outer envelope of the lamp is broken or punctured. Do not use where people will remain for more than a few minutes unless adequate shielding or other safety precautions are used. Lamps that will automatically extinguish when the outer envelope is broken or punctured are commercially available.”.</Text></Provision></Paragraph></Subsection></Section><Heading level="2"><TitleText>Testing Conditions</TitleText></Heading><Section><Label>6.</Label><Text>To determine whether a lamp is a self-extinguishing lamp, the lamp shall be tested as follows:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the lamp voltage, current and fixture shall be those indicated or recommended by the manufacturer for the operation of the lamp while undamaged;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>where there is no indication or recommendation by the manufacturer referred to in paragraph (<Emphasis style="italic">a</Emphasis>) respecting the fixture, the lamp shall be tested with no reflector or other surrounding material;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the lamp shall be started and operated on a reference ballast;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the lamp shall be started in air that has a temperature of 25± 5°C and any heating or movement of the air surrounding the lamp after the test has commenced shall be that produced by the lamp and ballast alone; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>if any test is performed in an enclosure, the enclosure shall not be less than 0.227 cubic metre.</Text></Paragraph></Section><Heading level="1"><Label>PART XI</Label><TitleText>TANNING EQUIPMENT</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>The following definitions apply in this Part.</Text><Definition generate-in-text="no"><Text><DefinedTermEn>double-contact medium screw lampholder</DefinedTermEn> means a lampholder described in <XRefExternal reference-type="standard">American National Standard for Lampholders for Electric Lamps</XRefExternal>, ANSI C81.62-1991, Standard Sheet 2-158-1, entitled <XRefExternal reference-type="standard">Double-Contact Medium Screw Lampholder</XRefExternal>, published by the American National Standards Institute and approved on July 15, 1991. (<DefinedTermFr>douille à contact double pour vis moyenne</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>erythema reference action spectrum</DefinedTermEn> means the erythema action spectrum set out in section 5.2 of CIE Standard CIE S 007/E-1998 entitled <XRefExternal reference-type="standard">Erythema Reference Action Spectrum and Standard Erythema Dose</XRefExternal>, published in 1998 by the Commission internationale de l’éclairage. (<DefinedTermFr>spectre d’action érythémale de référence</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>exposure position</DefinedTermEn> means any place, orientation or distance relative to the ultraviolet-radiating surface of tanning equipment at which it is recommended by the manufacturer that the user be exposed. (<DefinedTermFr>position pendant l’exposition</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>exposure schedule</DefinedTermEn> means a program of exposure recommended by the manufacturer of tanning equipment that takes into account exposure times, intervals between exposures and the degree of sensitivity for each skin type. (<DefinedTermFr>programme d’expositions</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>irradiance</DefinedTermEn> means radiant power incident per unit area, expressed in watts per square metre (W/m<Sup>2</Sup>). (<DefinedTermFr>éclairement énergétique</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>maximum exposure time</DefinedTermEn> means the longest period for continuous exposure recommended by the manufacturer of tanning equipment. (<DefinedTermFr>durée maximale d’exposition</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>protective eyewear</DefinedTermEn> means a device that is worn by the user of tanning equipment to reduce the ultraviolet radiation reaching their eyes either directly or indirectly. (<DefinedTermFr>dispositif de protection des yeux</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>single-contact medium screw lampholder</DefinedTermEn> means a lampholder described in <XRefExternal reference-type="standard">American National Standard for Lampholders for Electric Lamps</XRefExternal>, ANSI C81.62-1991, Standard Sheet 2-157-1, entitled <XRefExternal reference-type="standard">Single-Contact Medium Screw Lampholder</XRefExternal>, published by the American National Standards Institute and approved on July 15, 1991. (<DefinedTermFr>douille à contact unique pour vis moyenne</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>spectral irradiance</DefinedTermEn> means the irradiance that results from radiation within an infinitesimally small wavelength range, expressed in watts per square metre per nanometre (W/m<Sup>2</Sup>/nm). (<DefinedTermFr>éclairement énergétique spectral</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>spectral transmittance</DefinedTermEn> means the ratio of the spectral irradiance that is transmitted through protective eyewear to the spectral irradiance that is incident and normal to the surface of the eyewear. (<DefinedTermFr>transmittance spectrale</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>tanning equipment</DefinedTermEn> means a device that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>can be equipped with one or more ultraviolet lamps; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>induces skin tanning or other cosmetic effects.</Text></Paragraph><ContinuedDefinition><Text>It does not include any such device that is used in the production of therapeutic effects for medical purposes. (<DefinedTermFr>appareil de bronzage</DefinedTermFr>)</Text></ContinuedDefinition></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>timer</DefinedTermEn> means a device that is capable of ending the emission of ultraviolet radiation from tanning equipment after a preset period. (<DefinedTermFr>minuterie</DefinedTermFr>) </Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>ultraviolet lamp</DefinedTermEn> means a device that produces ultraviolet radiation in the wavelength range from 200 nm to 400 nm and is used in tanning equipment. (<DefinedTermFr>lampe à rayonnements ultraviolets</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>wavelength</DefinedTermEn> means a wavelength as measured in air. (<DefinedTermFr>longueur d’onde</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Information and Labelling</TitleText></Heading><Heading level="3"><TitleText>General</TitleText></Heading><Section><Label>2.</Label><Text>The information and labels required by this Part must be provided in both official languages.</Text></Section><Heading level="3"><TitleText>Information</TitleText></Heading><Section><Label>3.</Label><Text>The following information must accompany each piece of tanning equipment:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>instructions for its operation and safe use that include</Text><Subparagraph><Label>(i)</Label><Text>detailed directions for determining the exposure positions,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the maximum exposure time,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the minimum interval between consecutive exposures recommended by the manufacturer,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the maximum number of persons who may, at the same time, be exposed to ultraviolet radiation from the tanning equipment, as recommended by the manufacturer, and</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>the ultraviolet radiation warning labels described in section 5;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>instructions for obtaining repairs and the recommended replacement components and accessories that comply with the requirements of these Regulations; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a warning to always follow the instructions that accompany the equipment so as to avoid injury.</Text></Paragraph></Section><Heading level="3"><TitleText>Labelling</TitleText></Heading><Section><Label>4.</Label><Text>Every piece of tanning equipment must have permanently affixed to its external surface the following information, clearly legible and readily accessible to view by the user immediately before use:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the manufacturer’s name and address;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the model designation, serial number and month and year of manufacture;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>detailed directions for determining the exposure positions and a warning that the use of any other position may result in overexposure;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the recommended exposure time, as calculated in seconds using the formula</Text><FormulaGroup><Formula><FormulaText>X / (Σ V<Sub>λ</Sub> R<Sub>λ</Sub>)</FormulaText></Formula><FormulaConnector>and converted into and expressed in minutes, where</FormulaConnector><FormulaDefinition><FormulaTerm>X</FormulaTerm><Text>is a dose not greater than 100 J/m<Sup>2</Sup> for the first exposure session for untanned skin, gradually increasing over the following sessions to a maximum of 625 J/m<Sup>2</Sup> per session,</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>λ</FormulaTerm><Text>is the wavelength in nanometers,</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>R<Sub>λ</Sub></FormulaTerm><Text>is the irradiance of the tanning equipment, measured at the minimum exposure distance, and</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>V<Sub>λ</Sub></FormulaTerm><Text>is the weighting factor determined in accordance with the erythema reference action spectrum;</Text></FormulaDefinition></FormulaGroup></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>the minimum interval between consecutive exposures;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>the maximum number of minutes of exposure per year, as recommended by the manufacturer based on a maximum annual dose of 15 kJ/m<Sup>2</Sup>, weighted in accordance with the erythema reference action spectrum and taking into account the recommended exposure schedule;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>the model designation for each type of ultraviolet lamp that is to be used in the tanning equipment; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>the ultraviolet radiation warning labels designed in accordance with section 5.</Text></Paragraph></Section><Section><Label>5.</Label><Text>The ultraviolet radiation warning labels must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>be reproduced from the electronic file provided by the Minister;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>include in the French version of the label illustrated in Figure 1 of paragraph (<Emphasis style="italic">e</Emphasis>), enclosed within a black border,</Text><Subparagraph><Label>(i)</Label><Text>in the upper portion, on a white background, the signal word “Danger” in red with the hazard symbol to its right,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>in the middle portion, the primary hazard statement “Rayonnements ultraviolets” in yellow on a black background, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>in the lower portion, the following message in black on a white background:</Text><ContinuedSubparagraph><Text>“La surexposition provoque des brûlures aux yeux et à la peau. Porter le dispositif de protection des yeux. Suivre les instructions. Médicaments et cosmétiques peuvent augmenter les effets des UV. L’exposition aux UV peut avoir des effets nocifs sur la santé et contribuer, à long terme, au vieillissement prématuré et au cancer de la peau. Ces effets sont cumulatifs. Plus l’exposition régulière commence tôt, plus les risques qui y sont associés sont élevés.”;</Text></ContinuedSubparagraph></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>include in the English version of the label illustrated in Figure 2 of paragraph (<Emphasis style="italic">e</Emphasis>), enclosed within a black border,</Text><Subparagraph><Label>(i)</Label><Text>in the upper portion, on a white background, the signal word “Danger” in red with the hazard symbol to its right,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>in the middle portion, the primary hazard statement “Ultraviolet Radiation” in yellow on a black background, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>in the lower portion, the following message in black on a white background:</Text><ContinuedSubparagraph><Text>“Overexposure causes skin and eye burns. Use protective eyewear. Follow instructions. Drugs and cosmetics may increase UV effects. UV exposure can be hazardous to your health and in the long term can contribute to premature skin ageing and skin cancer. UV effects are cumulative. Greater risks are associated with early and repeated exposure.”;</Text></ContinuedSubparagraph></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>measure</Text><Subparagraph><Label>(i)</Label><Text>75 mm high and 200 mm wide, in the case of tanning equipment used for full- or half-body exposure, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>50 mm high and 100 mm wide, in any other case; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>conform to the following diagrams:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>The image is the French version of the ultraviolet radiation warning label as described in section 5. The image also includes a hazard symbol in the upper right corner, which is an octagonal sign depicting the outline of a human form being irradiated by a source of ultraviolet radiation represented by lines coming from a point above and to the right of the human form. In the upper left corner is a Canadian flag with the words “Health Canada” and “Santé Canada” to its right. In the lower right corner is the word “Canada” with a Canadian flag above the last letter.</AlternateText><Caption position="over" justification="center">Figure 1</Caption><Image source="CRC1370img\CRC1370_ef016.png" /></ImageGroup><ImageGroup bilingual="yes" position="center"><AlternateText>The image is the English version of the ultraviolet radiation warning label as described in section 5. The image also includes a hazard symbol in the upper right corner, which is an octagonal sign depicting the outline of a human form being irradiated by a source of ultraviolet radiation represented by lines coming from a point above and to the right of the human form. In the upper left corner is a Canadian flag with the words “Health Canada” and “Santé Canada” to its right. In the lower right corner is the word “Canada” with a Canadian flag above the last letter.</AlternateText><Caption position="over" justification="center">Figure 2</Caption><Image source="CRC1370img\CRC1370_ef017.png" /></ImageGroup></Paragraph></Section><Section><Label>6.</Label><Subsection><Label>(1)</Label><Text>Subject to subsection (2), all advertising material in relation to tanning equipment must include, in a clearly legible manner, the signal word “Danger”, the primary hazard statements “Ultraviolet Radiation / Rayonnements ultraviolets” and the messages set out in subparagraphs 5(<Emphasis style="italic">b</Emphasis>)(iii) and (<Emphasis style="italic">c</Emphasis>)(iii).</Text></Subsection><Subsection><Label>(2)</Label><Text>Advertising material that is in only English or French must include, in a clearly legible manner,</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>if it is only in French, the signal word “Danger”, the primary hazard statement “Rayonnements ultraviolets” and the message set out in subparagraph 5(<Emphasis style="italic">b</Emphasis>)(iii); and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>if it is only in English, the signal word “Danger”, the primary hazard statement “Ultraviolet Radiation” and the message set out in subparagraph 5(<Emphasis style="italic">c</Emphasis>)(iii).</Text></Paragraph></Subsection></Section><Section><Label>7.</Label><Text>Every ultraviolet lamp must have a tag, tape or card affixed to it that sets out</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>its model designation; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the warning “DANGER — Ultraviolet radiation. Follow instructions. Use only in fixtures equipped with a timer. / DANGER — Rayonnements ultraviolets. Suivre les instructions. À n’utiliser qu’avec un dispositif pourvu d’une minuterie.”.</Text></Paragraph></Section><Heading level="2"><TitleText>Construction Standards</TitleText></Heading><Heading level="3"><TitleText>General</TitleText></Heading><Section><Label>8.</Label><Text>All controls, meters, lights or other indicators of a piece of tanning equipment must be readily identifiable and clearly labelled to indicate their function.</Text></Section><Heading level="3"><TitleText>Safety Features</TitleText></Heading><Section><Label>9.</Label><Text>Every piece of tanning equipment must have the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a control by which the person being exposed may easily turn off the tanning equipment at any time without disconnecting the electrical plug or removing the ultraviolet lamps; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a timer that meets the functioning standards set out in section 16.</Text></Paragraph></Section><Section><Label>10.</Label><Subsection><Label>(1)</Label><Text>Every piece of tanning equipment must have a physical barrier between the ultraviolet lamps and the user that prevents any direct physical contact between the user and the lamps.</Text></Subsection><Subsection><Label>(2)</Label><Text>In the case of tanning beds, the physical barrier must be constructed of plexiglass or an equivalent material.</Text></Subsection></Section><Heading level="3"><TitleText>Components and Accessories</TitleText></Heading><Section><Label>11.</Label><Text>Every ultraviolet lamp that is used in tanning equipment must be constructed so that it cannot be inserted and operated in a single-contact medium screw lampholder or a double-contact medium screw lampholder.</Text></Section><Section><Label>12.</Label><Text>Every piece of tanning equipment must be accompanied by a number of sets of protective eyewear at least equal to the maximum number of persons who may, at the same time, be exposed to ultraviolet radiation from the tanning equipment, as recommended by the manufacturer of the equipment.</Text></Section><Heading level="2"><TitleText>Functioning Standards</TitleText></Heading><Section><Label>13.</Label><Text>Every piece of tanning equipment, whether it has its original components or replacement components recommended by the manufacturer, must, under the conditions of use specified by the manufacturer, meet the functioning standards set out in this Part.</Text></Section><Section><Label>14.</Label><Text>Every ultraviolet lamp that is used in tanning equipment must function so that, at any distance and in any direction from the radiation source, the irradiance within the wavelength range from 200 nm to less than 260 nm does not exceed 0.003 of the irradiance within the wavelength range from 260 nm to 320 nm.</Text></Section><Section><Label>15.</Label><Text>Every replacement ultraviolet lamp must function so that the maximum exposure time remains within 10% of the maximum exposure time originally recommended by the manufacturer.</Text></Section><Section><Label>16.</Label><Text>The timer must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>be adjustable to preset times and have a maximum timer setting not greater than the maximum exposure time recommended by the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>have a margin of error not greater than 10% of the maximum timer setting; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>not automatically reset when the tanning equipment emissions have been ended by the timer.</Text></Paragraph></Section><Section><Label>17.</Label><Text>Protective eyewear must have a spectral transmittance that is</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>not more than 0.001 over the wavelength range from 200 nm to 320 nm;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>not more than 0.01 over the wavelength range from 320 nm to 400 nm; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>sufficient over wavelengths greater than 400 nm to enable the user to read the labels and use the control specified in paragraph 9(<Emphasis style="italic">a</Emphasis>).</Text></Paragraph></Section><Heading level="1"><Label>PART XII</Label><TitleText>DIAGNOSTIC X-RAY EQUIPMENT</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Subsection><Label>(1)</Label><Text>The definitions in this subsection apply in this Part.</Text><Definition generate-in-text="no"><Text><DefinedTermEn>aluminum</DefinedTermEn> means aluminum that has a degree of purity of 99.9% or higher and a density of 2.70 g/cm<Sup>3</Sup>. (<DefinedTermFr>aluminium</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>aluminum equivalent</DefinedTermEn> means the attenuation equivalent of an object expressed in thickness of aluminum. (<DefinedTermFr>équivalent en aluminium</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>field emission device</DefinedTermEn> means a device in which the emission of electrons from the cathode is due solely to the action of an electric field. (<DefinedTermFr>dispositif d’émission par effet de champ</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>general purpose radiographic equipment</DefinedTermEn> means any stationary equipment other than that used solely for the examination of specific anatomical regions. (<DefinedTermFr>appareil de radiographie pour usage général</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>loading factor</DefinedTermEn> means a factor the value of which influences the X-ray tube load, and includes</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>for diagnostic X-ray equipment, if the X-ray beam is produced by the discharge of the capacitor through an X-ray tube, the X-ray tube voltage and the amount of capacitor charge;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>for a field emission device, the X-ray tube voltage and the number of pulses; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>for any other diagnostic X-ray equipment, the X-ray tube voltage and</Text><Subparagraph><Label>(i)</Label><Text>the X-ray tube current and irradiation time, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the current time product. (<DefinedTermFr>paramètre de charge</DefinedTermFr>)</Text></Subparagraph></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>mammography equipment</DefinedTermEn> means diagnostic X-ray equipment that is used for the examination of breast tissue. (<DefinedTermFr>appareil à mammographie</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>mobile equipment</DefinedTermEn> means, with respect to diagnostic X-ray equipment, equipment that is moved between incidents of use. (<DefinedTermFr>appareil mobile</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>radiographic equipment</DefinedTermEn> means diagnostic X-ray equipment that implements a technique in which the information contained in the X-ray pattern is obtained, recorded and optionally processed. (<DefinedTermFr>appareil de radiographie</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>radioscopic equipment</DefinedTermEn> means diagnostic X-ray equipment that implements a technique in which continuous or periodic sequences of X-ray patterns are produced and simultaneously and continuously displayed in the form of visible images. (<DefinedTermFr>appareil de radioscopie</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>radioscopic imaging assembly</DefinedTermEn> means the combination of components in radioscopic equipment that uses X-ray photons to produce a radioscopic image. These components usually consist of the X-ray image receptor, X-ray image intensifier, equipment housings, interlocks and protective shielding. (<DefinedTermFr>système d’imagerie radioscopique</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>rectification type</DefinedTermEn> means, with respect to diagnostic X-ray equipment, the process by which the X-ray generator converts high voltage to X-ray tube voltage. (<DefinedTermFr>type deredressement</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>stationary equipment</DefinedTermEn> means, with respect to diagnostic X-ray equipment, equipment that is never moved between incidents of use. (<DefinedTermFr>appareil fixe</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>X-ray image receptor</DefinedTermEn> means a device that converts incident X-rays into a visible image or into a form that can be made into a visible image by further transformation. (<DefinedTermFr>récepteur d’image radiologique</DefinedTermFr>)</Text></Definition></Subsection><Subsection><Label>(2)</Label><Text>Unless otherwise defined, words and expressions used in this Part have the same meaning as in the International Electrotechnical Commission Standard entitled <XRefExternal reference-type="other">Medical radiology — Terminology</XRefExternal>, Publication 788, First edition, 1984.</Text></Subsection></Section><Heading level="2"><TitleText>Information and Labelling</TitleText></Heading><Heading level="3"><TitleText>Information</TitleText></Heading><Section><Label>2.</Label><Text>The manufacturer must ensure that the following information accompanies each piece of diagnostic X-ray equipment:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the installation instructions;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the address of the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>any radiological safety procedures and additional precautions that are necessary because of any unique features of the equipment;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the maintenance instructions necessary to keep the equipment in compliance with the requirements of this Part;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>the rated line voltage, the maximum line current and the line voltage regulation for the operation of the equipment at the maximum line current;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>the loading factors that constitute the maximum line current condition for the X-ray generator;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>for each X-ray tube assembly,</Text><Subparagraph><Label>(i)</Label><Text>the nominal focal spot sizes and the method of their determination,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the cooling curves for the anode and for the X-ray tube housing,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the X-ray tube rating charts, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the method by which the focal spot to image receptor distance can be determined using the indicator specified in subparagraph 3(<Emphasis style="italic">c</Emphasis>)(i);</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>its duty cycles, rectification type and generator rating;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>if the equipment is battery powered, the minimum state of charge necessary for it to operate;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">j</Emphasis>)</Label><Text>the operating range of X-ray tube voltages and the maximum deviation for any selected X-ray tube voltage within that range of values;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">k</Emphasis>)</Label><Text>if the equipment is not operated exclusively in automatic exposure control mode, the accuracy limits of</Text><Subparagraph><Label>(i)</Label><Text>the controlling timer,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the X-ray tube current, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the current time product;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">l</Emphasis>)</Label><Text>where the equipment operates under automatic exposure control, the accuracy limits of that control; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">m</Emphasis>)</Label><Text>the conditions under which the information provided under paragraphs (<Emphasis style="italic">j</Emphasis>) to (<Emphasis style="italic">l</Emphasis>) is valid.</Text></Paragraph></Section><Heading level="3"><TitleText>Labelling</TitleText></Heading><Section><Label>3.</Label><Text>Diagnostic X-ray equipment must display the following information in a manner that is legible, permanent and visible on the specified surfaces:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>on the external surface of the main control panel</Text><Subparagraph><Label>(i)</Label><Text>a statement prohibiting unauthorized use and warning that hazardous X-rays are emitted when the equipment is in operation,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the X-ray warning symbol described in section 4, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>with respect to the X-ray generator,</Text><Clause><Label>(A)</Label><Text>the name of the manufacturer,</Text></Clause><Clause><Label>(B)</Label><Text>the model designation,</Text></Clause><Clause><Label>(C)</Label><Text>the serial number,</Text></Clause><Clause><Label>(D)</Label><Text>the date of manufacture, and</Text></Clause><Clause><Label>(E)</Label><Text>the country of manufacture;</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>on the external surface of the X-ray tube housing, with respect to the X-ray tube assembly,</Text><Subparagraph><Label>(i)</Label><Text>the name of the manufacturer,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the model designation,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the serial number,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the date of installation of the X-ray tube in the X-ray tube housing,</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>the country of manufacture, and</Text></Subparagraph><Subparagraph><Label>(vi)</Label><Text>the minimum permanent inherent filtration of the X-ray beam emitted from the X-ray tube assembly, expressed in millimetres of aluminum equivalent at a specified X-ray tube voltage;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>on the external surface of the X-ray tube housing or another suitable structure permanently attached to the X-ray tube housing</Text><Subparagraph><Label>(i)</Label><Text>an indicator that enables the focal spot to image receptor distance to be determined to within 2% of that distance, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>if the X-ray tube and the X-ray generator are not located within a common enclosure, marks that clearly indicate the anode and cathode terminals on the X-ray tube housing and on the high-voltage generator; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>on the external surface of any beam limiting device that adds filtration to the X-ray beam, the total permanent filtration deliverable by the beam limiting device, expressed in millimetres of aluminum equivalent at a specified X-ray tube voltage.</Text></Paragraph></Section><Section><Label>4.</Label><Text>The X-ray warning symbol shall</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>be displayed in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>be visible and identifiable from a distance of 1 m;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>be at least 2 cm high and at least 2 cm wide;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bear the words “CAUTION: X-RAYS — ATTENTION : RAYONS X”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>conform to</Text><Subparagraph><Label>(i)</Label><Text>the following diagram:</Text><ImageGroup position="center"><AlternateText>Warning sign, bearing the words “Caution: X-Rays — Attention: Rayons X”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef009.png" /></ImageGroup></Subparagraph><ContinuedParagraph><Text>or</Text></ContinuedParagraph><Subparagraph><Label>(ii)</Label><Text>symbol 03-03 in the report of the International Electrotechnical Commission entitled <XRefExternal reference-type="other">Graphical symbols for electrical equipment in medical practice</XRefExternal>, Publication 878, 1988, illustrated as follows:</Text><ImageGroup position="center"><AlternateText>Warning sign, bearing the graphic of three blades around a circle which represents symbol 03-03 in the report of the International Electrotehcnical Commission entitled Graphical symbols for electrical equipment in medical practice, Publication 878, 1988.</AlternateText><Image source="CRC1370img\CRC1370_ef010.png" /></ImageGroup></Subparagraph></Paragraph></Section><Section><Label>5.</Label><Text>All controls, meters, warning lights and other indicators required by this Part must be clearly labelled as to their function.</Text></Section><Heading level="2"><TitleText>Construction Standards</TitleText></Heading><Heading level="3"><TitleText>General Requirements</TitleText></Heading><Section><Label>6.</Label><Text>Diagnostic X-ray equipment must have</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>if more than one X-ray tube is controlled by one control panel,</Text><Subparagraph><Label>(i)</Label><Text>a visual indicator on or near each X-ray tube housing that shows that the X-ray tube to which the indicator applies is connected and ready to be energized, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>a visual indicator on the control panel that shows which of the X-ray tubes are connected and ready to be energized;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a means, appropriate to the rectification type of the equipment, to compensate for variations in X-ray tube voltage caused by line voltage fluctuations;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a visual or audible indicator that warns the operator when the variation in line voltage exceeds the rate set out in subsection 23(2) or a mechanism that, in that event, prevents X-rays from being emitted;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>on the control panel</Text><Subparagraph><Label>(i)</Label><Text>a warning light that indicates when the equipment is ready to be energized,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>a second warning light that indicates when X-rays are being emitted,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>if an automatic exposure control is provided, a visual indicator showing when that mode of operation is selected, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>if the automatic exposure control mode is not selected or does not exist, controls and visual indicators that enable the operator to select the loading factors before an irradiation;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>if the equipment is battery powered, a visual indicator on the control panel showing whether the battery is adequately charged for the proper operation of the equipment;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>a mechanism to initiate and terminate an irradiation;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>an audible signal to indicate the termination of an irradiation;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>in the case of equipment other than mammography equipment, if an X-ray source assembly has a permanent inherent filtration of 0.5 mm aluminum equivalent or less, a means to add additional filtration;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>if the equipment moves around a patient by remote control, an emergency stop switch that immediately terminates both the motion of the equipment and the emission of X-rays;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">j</Emphasis>)</Label><Text>a beam limiting device; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">k</Emphasis>)</Label><Text>for equipment that operates within a range set out in column 1 of an item of the table to this paragraph, radiation filters that result in a measured half-value layer of aluminum of not less than</Text><Subparagraph><Label>(i)</Label><Text>for each X-ray tube voltage set out in column 2 of that item, the half-value layer set out in column 3 of that item, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>in any other case, the half-value layer obtained by linear interpolation or extrapolation from that table.</Text></Subparagraph><TableGroup pointsize="10" spanlanguages="yes" topmarginspacing="4"><Caption position="over" justification="center">TABLE TO PARAGRAPH 6 (<Emphasis style="italic">k</Emphasis>)</Caption><table frame="topbot"><title justification="center">MINIMUM HALF-VALUE LAYER OF ALUMINUM</title><tgroup cols="4"><colspec colname="1" colwidth="47.07*" htmlwidth="10%" /><colspec colname="2" colwidth="161.97*" htmlwidth="35%" /><colspec colname="3" colwidth="147.61*" htmlwidth="32%" /><colspec colname="4" colwidth="111.30*" htmlwidth="24%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t855h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t855h1">Column 1</entry><entry colsep="0" rowsep="0" valign="bottom" id="t855h2">Column 2</entry><entry colsep="0" rowsep="0" valign="bottom" id="t855h3">Column 3</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t855h4" headers="t855h1">Operating Range for Normal Use (kV)</entry><entry colsep="0" rowsep="1" valign="bottom" id="t855h5" headers="t855h2">X-ray Tube Voltage (kV)</entry><entry colsep="0" rowsep="1" valign="bottom" id="t855h6" headers="t855h3">Half-value Layer of Aluminum (mm)</entry></row></thead><tbody><row><entry colname="1" morerows="2" colsep="0" rowsep="0" valign="top" headers="t855h0">1.</entry><entry rowheader="yes" colname="2" morerows="2" colsep="0" rowsep="0" valign="top" id="t855h8" headers="t855h1 t855h4">50 or less</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h8 t855h2 t855h5">(<Emphasis style="italic">a</Emphasis>) 30</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h8 t855h3 t855h6">0.3</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h8 t855h2 t855h5">(<Emphasis style="italic">b</Emphasis>) 40</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h8 t855h3 t855h6">0.4</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h8 t855h2 t855h5">(<Emphasis style="italic">c</Emphasis>) 50</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h8 t855h3 t855h6">0.5</entry></row><row><entry colname="1" morerows="2" colsep="0" rowsep="0" valign="top" headers="t855h0">2.</entry><entry rowheader="yes" colname="2" morerows="2" colsep="0" rowsep="0" valign="top" id="t855h16" headers="t855h1 t855h4">50 or more but not more than 70</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h16 t855h2 t855h5">(<Emphasis style="italic">a</Emphasis>) 50</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h16 t855h3 t855h6">1.2</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h16 t855h2 t855h5">(<Emphasis style="italic">b</Emphasis>) 60</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h16 t855h3 t855h6">1.3</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h16 t855h2 t855h5">(<Emphasis style="italic">c</Emphasis>) 70</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h16 t855h3 t855h6">1.5</entry></row><row><entry colname="1" morerows="8" colsep="0" rowsep="0" valign="top" headers="t855h0">3.</entry><entry rowheader="yes" colname="2" morerows="8" colsep="0" rowsep="0" valign="top" id="t855h24" headers="t855h1 t855h4">70 or more</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">a</Emphasis>) 70</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">2.1</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">b</Emphasis>) 80</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">2.3</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">c</Emphasis>) 90</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">2.5</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">d</Emphasis>) 100</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">2.7</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">e</Emphasis>) 110</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">3.0</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">f</Emphasis>) 120</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">3.2</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">g</Emphasis>) 130</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">3.5</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">h</Emphasis>) 140</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">3.8</entry></row><row><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h2 t855h5">(<Emphasis style="italic">i</Emphasis>) 150</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t855h1 t855h4 t855h24 t855h3 t855h6">4.1</entry></row></tbody></tgroup></table></TableGroup></Paragraph></Section><Section><Label>7.</Label><Subsection><Label>(1)</Label><Text>An irradiation switch for diagnostic X-ray equipment must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>permit the emission of X-rays only when the operator exerts continuous pressure on the switch;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>in the case of a foot switch, prevent the emission of any unintended X-rays when it is overturned; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>in the case of mobile equipment, permit the operator to stand at least 3 m from the X-ray source when the X-ray tube is energized.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>The controlling timer for diagnostic X-ray equipment must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>automatically terminate an irradiation</Text><Subparagraph><Label>(i)</Label><Text>on completion of a preset irradiation time,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>on attainment of a preset current time product value, or</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>on completion of a preset number of X-ray pulses;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>permit the operator to terminate an irradiation at any time;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>automatically reset itself to its original setting or to zero on termination of an irradiation; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>prevent the initiation of irradiation when the timer is set at zero, at the “off” position or at an unmarked setting.</Text></Paragraph></Subsection></Section><Section><Label>8.</Label><Subsection><Label>(1)</Label><Text>In the case of diagnostic X-ray equipment, other than mammography equipment, when an object set out in column 1 of an item of the table to this subsection is positioned between the patient and the X-ray image receptor, the aluminum equivalent of the object shall not exceed the amount set out in column 2 of that item, as determined using an X-ray beam that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is generated at an X-ray tube voltage of 100 kV;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>has a maximum X-ray tube voltage ripple of 10%; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has a half-value layer of aluminum of 3.7 mm.</Text></Paragraph><TableGroup pointsize="10" spanlanguages="yes" topmarginspacing="4"><Caption position="over" justification="center">TABLE TO SUBSECTION 8(1)</Caption><table frame="topbot"><title justification="center">NON-MAMMOGRAPHY EQUIPMENT — MAXIMUM ALUMINUM EQUIVALENT OF INTERPOSITIONED OBJECTS</title><tgroup cols="3"><colspec colname="1" colwidth="45.0*" htmlwidth="10%" /><colspec colname="2" colwidth="316.20*" htmlwidth="68%" /><colspec colname="3" colwidth="106.8*" htmlwidth="23%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t856h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t856h1">Column 1</entry><entry colsep="0" rowsep="0" valign="bottom" id="t856h2">Column 2</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t856h3" headers="t856h1">Object</entry><entry colsep="0" rowsep="1" valign="bottom" id="t856h4" headers="t856h2">Maximum Aluminum Equivalent (mm)</entry></row></thead><tbody><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">1.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Front panel of cassette holder (total of all layers)</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">1.2</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">2.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Front panel of film changer (total of all layers)</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">1.2</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">3.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Cradle</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">2.3</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">4.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Fixed patient support without an articulated joint</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">1.2</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">5.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Movable patient support without an articulated joint, including any fixed layers</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">1.7</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">6.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Patient support with one articulated joint and a radiolucent panel and</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">1.7</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">7.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Patient support with two or more articulated joints and a radiolucent panel</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">2.3</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t856h0">8.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t856h1 t856h3">Cantilevered patient support</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t856h2 t856h4">2.3</entry></row></tbody></tgroup></table></TableGroup></Subsection><Subsection><Label>(2)</Label><Text>In the case of mammography equipment, when an object set out in column 1 of an item of the table to this subsection is positioned between the patient and the X-ray image receptor, the aluminum equivalent of the object shall not exceed the amount set out in column 2 of that item, as determined using an X-ray beam that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is generated at an X-ray tube voltage of 30 kV;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>has a maximum X-ray tube voltage ripple of 10%; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has a half-value layer of aluminum of 0.3 mm.</Text></Paragraph><TableGroup pointsize="10" spanlanguages="yes" topmarginspacing="4"><Caption position="over" justification="center">TABLE TO SUBSECTION 8(2)</Caption><table frame="topbot"><title justification="center">MAMMOGRAPHY EQUIPMENT — MAXIMUM ALUMINUM EQUIVALENT OF INTERPOSITIONED OBJECTS</title><tgroup cols="3" keep-together="yes"><colspec colname="1" colwidth="57.50*" htmlwidth="12%" /><colspec colname="2" colwidth="221.02*" htmlwidth="47%" /><colspec colname="3" colwidth="189.48*" htmlwidth="40%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t857h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t857h1">Column 1</entry><entry colsep="0" rowsep="0" valign="bottom" id="t857h2">Column 2</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t857h3" headers="t857h1">Object</entry><entry colsep="0" rowsep="1" valign="bottom" id="t857h4" headers="t857h2">Maximum Aluminum Equivalent (mm)</entry></row></thead><tbody><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t857h0">1.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t857h1 t857h3">Support table, including all layers</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t857h2 t857h4">0.3</entry></row></tbody></tgroup></table></TableGroup></Subsection><Subsection><Label>(3)</Label><Text>For the purposes of subsections (1) and (2), any sensor used in automatic exposure control is a part of the X-ray image receptor.</Text></Subsection></Section><Section><Label>9.</Label><Text>For diagnostic X-ray equipment,</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the X-ray tube must be securely affixed to and aligned within the X-ray tube housing;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the radiation filters must be securely affixed to the exit port of the X-ray tube housing or beam limiting device, or both; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the X-ray source assembly must maintain its required position or movement without drift or vibration during operation.</Text></Paragraph></Section><Heading level="3"><TitleText>Radiographic Equipment</TitleText></Heading><Section><Label>10.</Label><Text>Radiographic equipment that is equipped with an automatic exposure control must have</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>if the operating X-ray tube voltage is 50 kV or more, a minimum irradiation time capability that does not exceed</Text><Subparagraph><Label>(i)</Label><Text>in the case of a field emission device that operates in pulse mode, the time equivalent to two pulses, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>in the case of any other radiographic equipment, the greater of 1/60 s or the time required to deliver a current time product of 5 mAs;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a means to automatically terminate the irradiation when</Text><Subparagraph><Label>(i)</Label><Text>if the operating X-ray tube voltage is less than 50 kV, the current time product exceeds 1,200 mAs per irradiation, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>if the operating X-ray tube voltage is 50 kV or more,</Text><Clause><Label>(A)</Label><Text>the current time product exceeds 600 mAs per irradiation, or</Text></Clause><Clause><Label>(B)</Label><Text>the product of the X-ray tube voltage, X-ray tube current and irradiation time exceeds 60 kVAs per irradiation; and</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>when an irradiation under automatic exposure control terminates because the limits specified in paragraph (<Emphasis style="italic">b</Emphasis>) have been reached,</Text><Subparagraph><Label>(i)</Label><Text>a visual indicator or audible signal that warns the operator of the termination, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>a reset control that must be activated manually before another irradiation under automatic exposure control can be made.</Text></Subparagraph></Paragraph></Section><Section><Label>11.</Label><Subsection><Label>(1)</Label><Text>General purpose radiographic equipment must have</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a beam limiting device that</Text><Subparagraph><Label>(i)</Label><Text>permits stepless adjustment of the size of the X-ray field, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>when it is set at the smallest aperture and at a focal spot to image receptor distance of 1 m, has an X-ray field of 5 cm by 5 cm, or less;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>an X-ray field indicator that uses light to visually define the X-ray field so that the limits of the X-ray field are visible under the ambient lighting conditions in an X-ray room; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a means by which the operator may</Text><Subparagraph><Label>(i)</Label><Text>determine when the X-ray beam axis is perpendicular to the image receptor plane,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>determine the focal spot to image receptor distance to within 2% of that distance, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>align the centre of the X-ray field with the centre of the image reception area to within 2% of the focal spot to image receptor distance.</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>The X-ray field indicator referred to in paragraph (1)(<Emphasis style="italic">b</Emphasis>) must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>illuminate on average at a minimum of 100 lx when measured from the lesser of</Text><Subparagraph><Label>(i)</Label><Text>the distance of 1 m from the X-ray source, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the maximum focal spot to image receptor distance;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>be circumscribed by the beam limiting device;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>have as its perimeter the locus of points at which the illumination is one fourth of the maximum illumination in the area; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>when the X-ray beam axis is perpendicular to the image receptor plane, visually define the X-ray field within the following specifications, namely,</Text><Subparagraph><Label>(i)</Label><Text>the separation between the perimeter of the visually defined field and that of the X-ray field does not exceed 2% of the focal spot to image receptor distance, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the dimensions of the X-ray field are indicated and are accurate to within 2% of the focal spot to image receptor distance.</Text></Subparagraph></Paragraph></Subsection></Section><Section><Label>12.</Label><Subsection><Label>(1)</Label><Text>General purpose radiographic equipment that has a positive beam limiting system must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>permit stepless adjustment of the size of the X-ray field;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>when it is set at the smallest aperture and at a focal spot to image receptor distance of 1 m, have an X-ray field of 5 cm by 5 cm, or less;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>permit adjustment of the size of the X-ray field to dimensions that are smaller than those of the image reception area;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>under the conditions of operation specified in subsection (2),</Text><Subparagraph><Label>(i)</Label><Text>automatically adjust the dimensions of the X-ray field to the dimensions of the image reception area, or to a selected portion of that area, within 5 s after insertion of the image receptor, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>prevent the emission of X-rays until the beam limiting device is manually adjusted so that</Text><Clause><Label>(A)</Label><Text>the dimensions of the X-ray field do not exceed those of the image reception area, or the selected portion of that area, by more than 3% of the focal spot to image receptor distance, and</Text></Clause><Clause><Label>(B)</Label><Text>the sum of the absolute values of the differences in the dimensions of the X-ray field and the image reception area, or the selected portion of that area, does not exceed 4% of the focal spot to image receptor distance; and</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>automatically revert to one of the requirements set out in paragraph (<Emphasis style="italic">d</Emphasis>) on any change to</Text><Subparagraph><Label>(i)</Label><Text>the dimensions of the image reception area, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the focal spot to image receptor distance, if the change would result in failure to meet the requirements of paragraph (<Emphasis style="italic">d</Emphasis>).</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of paragraph (1)(<Emphasis style="italic">d</Emphasis>), the conditions of operation are as follows:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the image receptor is inserted into a permanently mounted cassette holder;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>neither the length nor the width of the image reception area exceeds 50 cm;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the X-ray beam axis is within 3° of the perpendicular to the image receptor plane;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the X-ray beam axis is within 3° of</Text><Subparagraph><Label>(i)</Label><Text>the horizontal plane, when the focal spot to image receptor distance is 90 cm or more but not more than 205 cm, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the vertical plane, when the focal spot to image receptor distance is 90 cm or more but not more than 130 cm; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>neither tomography nor stereoscopic radiography is being performed.</Text></Paragraph></Subsection></Section><Section><Label>13.</Label><Subsection><Label>(1)</Label><Text>Subject to section 14, radiographic equipment, other than general purpose radiographic or mammography equipment, must have a fixed-aperture beam limiting device that, for the combination of image reception area and focal spot to image receptor distance described in subsection (2),</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>permits the centre of the X-ray field to be aligned with the centre of the image reception area to within 2% of the focal spot to image receptor distance; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>prevents the X-ray field from extending beyond any edge of the image reception area.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>The fixed-aperture beam limiting device referred to in subsection (1) must display on its exterior surface a specified focal spot to image receptor distance and the dimensions of its image reception area at that distance.</Text></Subsection></Section><Section><Label>14.</Label><Text>Mobile radiographic equipment that does not meet the requirements of section 13 must have</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a beam limiting device that</Text><Subparagraph><Label>(i)</Label><Text>permits stepless adjustment of the size of the X-ray field, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>when it is set at the smallest aperture and at a focal spot to image receptor distance of 1 m, has an X-ray field of 5 cm by 5 cm, or less;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>an X-ray field indicator referred to in paragraph 11(1)(<Emphasis style="italic">b</Emphasis>) that, when the X-ray beam axis is perpendicular to the image receptor plane, permits the alignment of the edges of the illuminated field with the edges of the X-ray field so that the difference between their edges does not exceed 2% of the focal spot to image receptor distance; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a means to determine the focal spot to image receptor distance to an accuracy of 2% or less of that distance.</Text></Paragraph></Section><Section><Label>15.</Label><Subsection><Label>(1)</Label><Text>Mammography equipment must have</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a beam limiting device that limits the size of the X-ray beam to prevent the X-ray field, at any focal spot to image receptor distance at which the equipment operates, from extending</Text><Subparagraph><Label>(i)</Label><Text>more than 5 mm beyond the edge of the patient support next to the chest wall of the patient, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>more than a distance equivalent to 2% of the focal spot to image receptor distance beyond any other edge of the image reception area;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>an image receptor supporting device that</Text><Subparagraph><Label>(i)</Label><Text>has protective shielding that limits the residual radiation in accordance with section 26,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>extends to the patient’s chest wall, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>at every other edge, extends beyond the X-ray field by at least 1% of the focal spot to image receptor distance; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a breast compression device that</Text><Subparagraph><Label>(i)</Label><Text>is foot-actuated to start the compression,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>permits fine adjustment of motion during the compression,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>permits rapid decompression,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>has motion adjustment controls on both sides of the position for the patient, and</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>allows the portion of the compression plate in contact with the breast to be brought to within 10 mm of the surface of the patient support.</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>Mammography equipment that has a removable, fixed-aperture beam limiting device must display the following information on its external surface:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the dimensions of the image reception area; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the focal spot to image receptor distance at which the beam limiting device must be used.</Text></Paragraph></Subsection></Section><Section><Label>16.</Label><Text>Diagnostic X-ray equipment that has a spotfilm device must have</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>if the angle of the image receptor plane or of the X-ray source assembly is adjustable by the operator, a means to indicate when the X-ray beam axis is perpendicular to the image receptor plane;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>when the X-ray beam axis is perpendicular to the image receptor plane, a mechanism that</Text><Subparagraph><Label>(i)</Label><Text>when the X-ray field is larger than the selected portion of the image reception area, adjusts the dimensions of the field automatically to those of the selected portion of that area,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>when the X-ray field is smaller than the selected portion of the image reception area, permits the adjustment of the dimensions of the X-ray field,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>permits the perimeter of the X-ray field to be aligned with that of the selected portion of the image reception area so that</Text><Clause><Label>(A)</Label><Text>the dimensions of the X-ray field differ from the corresponding dimensions of the image reception area by a distance that does not exceed 3% of the focal spot to image receptor distance, and</Text></Clause><Clause><Label>(B)</Label><Text>the sum of the absolute values of the differences in the dimensions between the X-ray field size and the image reception area does not exceed 4% of the focal spot to image receptor distance, and</Text></Clause></Subparagraph><Subparagraph><Label>(iv)</Label><Text>permits the centre of the X-ray field to be aligned with the centre of the selected portion of the image reception area to within 2% of the focal spot to image receptor distance;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a mechanism for adjusting the X-ray field to dimensions that are smaller than those of the selected portion of the image reception area so that</Text><Subparagraph><Label>(i)</Label><Text>when the mechanism is set at the smallest aperture and at the longest focal spot to image receptor distance, the X-ray field is 5 cm by 5 cm, or less, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the requirement set out in subparagraph (<Emphasis style="italic">b</Emphasis>)(iv) is met when the X-ray beam axis is perpendicular to the image receptor plane; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>if a means is provided for the operator to override a failure of the automatic X-ray field size adjustment, a visual indicator at the operator’s position that shows when the override is activated.</Text></Paragraph></Section><Section><Label>17.</Label><Text>Radiographic equipment, other than equipment described in sections 11 to 16, must have a beam limiting device that, when the axis of the X-ray beam is perpendicular to the image receptor plane, permits</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the alignment of the centre of the X-ray field with the centre of the image reception area to within 2% of the focal spot to image receptor distance; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the adjustment of the perimeter of the X-ray field so that the perimeter does not extend beyond that of the image reception area by more than 2% of the focal spot to image receptor distance.</Text></Paragraph></Section><Heading level="3"><TitleText>Radioscopic Equipment</TitleText></Heading><Section><Label>18.</Label><Text>Radioscopic equipment must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>have an X-ray image intensifier that includes protective shielding that</Text><Subparagraph><Label>(i)</Label><Text>for any focal spot to image receptor distance, intercepts the entire cross section of the X-ray beam,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>intercepts the X-ray beam and scattered radiation from the image intensifier that would otherwise reach the operator,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>prevents the radioscopic X-ray tube from emitting X-rays unless the protective shielding is in place to intercept the X-ray beam, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>sufficiently attenuates the X-rays transmitted through or scattered from the entrance window of the radioscopic imaging assembly to meet the requirements of section 32;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>for mobile radioscopic equipment, have an X-ray image intensifier that is an integral part of the equipment or is interlocked in such a manner that its removal prevents X-rays from being produced;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>for stationary radioscopic equipment, prevent the X-ray tube from producing X-rays when there is no image receptor in place to intercept the X-ray beam;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>for stationary radioscopic equipment that is not equipped with remote control,</Text><Subparagraph><Label>(i)</Label><Text>have protective shielding of at least 0.25 mm lead equivalent at 100 kV, such as overlapping hinged or sliding panels or protective drapes, to intercept the scattered radiation that would otherwise reach the operator, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the capability to remove the protective shielding referred to in subparagraph (i) when it interferes with the performance of diagnostic procedures;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>have an irradiation switch that</Text><Subparagraph><Label>(i)</Label><Text>requires continuous pressure by the operator for the entire period of an irradiation, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>enables the operator to terminate the recording of serial radioscopic images at any time;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>have a chronometer that</Text><Subparagraph><Label>(i)</Label><Text>indicates the amount of time that the equipment has been emitting X-rays, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>can be reset to zero or any other selected value;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">g</Emphasis>)</Label><Text>have a positive beam limiting system that, when the X-ray beam axis is perpendicular to the image receptor plane, permits the alignment of the perimeter of the illuminated field with that of the X-ray field so that</Text><Subparagraph><Label>(i)</Label><Text>the difference between the perimeters does not exceed a distance equivalent to 3% of the focal spot to image receptor distance, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the sum of the excess length and excess width does not exceed 4% of the focal spot to image receptor distance;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">h</Emphasis>)</Label><Text>have visual indicators that continuously display the X-ray tube voltage and the X-ray tube current; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">i</Emphasis>)</Label><Text>have a device that limits the minimum focal spot to skin distance</Text><Subparagraph><Label>(i)</Label><Text>for mobile radioscopic equipment, to 30 cm,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>for stationary radioscopic equipment, to 38 cm,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>for radioscopic equipment fitted with an X-ray image intensifier and used for special applications that require shorter focal spot to skin distances than the focal spot to skin distances specified in subparagraphs (i) and (ii), to 20 cm, or</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>for small-format, low-intensity radioscopic equipment, to the distance at which the equipment delivers an air kerma rate of 50 mGy/min or an exposure rate of 5.75 R/min.</Text></Subparagraph></Paragraph></Section><Section><Label>19.</Label><Text>Radioscopic equipment that is used for cineradiography must have visual indicators that continuously display the X-ray tube voltage and the X-ray tube current.</Text></Section><Section><Label>20.</Label><Text>A high-level irradiation control for radioscopic equipment must</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>be activated by a separate means that requires continuous pressure by the operator for it to emit X-rays; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>when it is in use, emit a continuous audible signal or an intermittent signal with silent periods of no longer than one second.</Text></Paragraph></Section><Heading level="2"><TitleText>Functioning Standards</TitleText></Heading><Section><Label>21.</Label><Text>Diagnostic X-ray equipment must function in accordance with the requirements set out in sections 22 to 32 during its operation under normal conditions of use.</Text></Section><Section><Label>22.</Label><Subsection><Label>(1)</Label><Text>The definitions in this subsection apply in this section.</Text><Definition generate-in-text="no"><Text><DefinedTermEn>coefficient of variation</DefinedTermEn> means the ratio of the estimated standard deviation to the mean value of a series of measurements calculated using the equation:</Text><FormulaGroup><Formula><ImageGroup position="center"><AlternateText>The coefficient of variation, calculated by the equation “C equals S divided by X bar which is also equal to one divided by x bar multiplied by, open square bracket, sum from i equals one to n, open parenthesis, x sub i minus x bar, close parenthesis, squared, divided by n minus one, close square bracket, to the power of one half”</AlternateText><Image source="CRC1370img\CRC1370_e015.png" /></ImageGroup></Formula><FormulaConnector>where</FormulaConnector><FormulaDefinition><FormulaTerm><Emphasis style="italic">C</Emphasis></FormulaTerm><Text>is the coefficient of variation;</Text></FormulaDefinition><FormulaDefinition><FormulaTerm><Emphasis style="italic">S</Emphasis></FormulaTerm><Text>is the estimated standard deviation;</Text></FormulaDefinition><FormulaDefinition><FormulaTerm><Emphasis style="italic">X<Sub>i</Sub></Emphasis></FormulaTerm><Text>is the value of the i<Sup>th</Sup> measurement;</Text></FormulaDefinition><FormulaDefinition><FormulaTerm><Emphasis style="overbar">X</Emphasis></FormulaTerm><Text>is the mean value of the measurements; and</Text></FormulaDefinition><FormulaDefinition><FormulaTerm><Emphasis style="italic">n</Emphasis></FormulaTerm><Text>is the number of measurements. (<DefinedTermFr>coefficient de variation</DefinedTermFr>)</Text></FormulaDefinition></FormulaGroup></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>exposure to the X-ray image receptor</DefinedTermEn> means the amount of X-rays, registered by one or more detectors located in a fixed position in proximity to the X-ray image receptor, that is necessary to produce a radiogram of the overall density sought by the operator. (<DefinedTermFr>exposition au récepteur d’image radiologique</DefinedTermFr>)</Text></Definition></Subsection><Subsection><Label>(2)</Label><Text>For any combination of X-ray tube voltage, X-ray tube current and irradiation time, or for any selected exposure to the X-ray image receptor, when the line voltage for each measurement is accurate to within 1% of the mean line voltage value of all the measurements, and when all variable controls for the loading factors are adjusted to alternate settings and reset to the test setting before each measurement,</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the coefficient of variation of any 10 consecutive air kerma or exposure measurements, taken at the same point along the X-ray beam axis within a period of one hour, must be no greater than 0.05; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>each of the 10 air kerma or exposure measurements taken under paragraph (<Emphasis style="italic">a</Emphasis>) must be within 15% of the mean value of those measurements.</Text></Paragraph></Subsection><Subsection><Label>(3)</Label><Text>For the purposes of subsection (2), diagnostic X-ray equipment with an automatic exposure control must have attenuating material in the X-ray beam that is thick enough that the loading factors can be adjusted to provide single irradiations of at least</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>12 pulses, in the case of a field emission device that operates in pulse mode; or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>0.1 s, in the case of any other diagnostic X-ray equipment.</Text></Paragraph></Subsection></Section><Section><Label>23.</Label><Subsection><Label>(1)</Label><Text>This section applies in respect of diagnostic X-ray equipment that has</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a high-voltage generator that is not a stored energy high-voltage generator;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>loading factors that do not change automatically to compensate for unintentional variations in X-ray tube voltage; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>an irradiation time of at least 0.1 s and a current time product of at least 5 mAs.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>In the case of a line voltage regulation of 6% or less, the loading factor set out in column 1 of an item of the table to this subsection must not deviate from the selected value, for any combination of loading factors, by more than the quantity set out in column 2 of that item.</Text><TableGroup pointsize="10" spanlanguages="yes" topmarginspacing="4"><Caption position="over" justification="center">TABLE TO SUBSECTION 23(2)</Caption><table frame="topbot"><title justification="center">MAXIMUM DEVIATION OF LOADING FACTORS</title><tgroup cols="3"><colspec colname="1" colwidth="42.25*" htmlwidth="9%" /><colspec colname="2" colwidth="254.70*" htmlwidth="54%" /><colspec colname="3" colwidth="171.04*" htmlwidth="37%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t858h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t858h1">Column 1</entry><entry colsep="0" rowsep="0" valign="bottom" id="t858h2">Column 2</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t858h3" headers="t858h1">Loading Factor</entry><entry colsep="0" rowsep="1" valign="bottom" id="t858h4" headers="t858h2">Maximum Deviation from the Selected Value</entry></row></thead><tbody><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t858h0">1.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t858h1 t858h3">X-ray tube voltage of mammography equipment</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t858h2 t858h4">5%</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t858h0">2.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t858h1 t858h3">X-ray tube voltage of non-mammography equipment</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t858h2 t858h4">10%</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t858h0">3.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t858h1 t858h3">Irradiation time</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t858h2 t858h4">10% plus 1 ms</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t858h0">4.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t858h1 t858h3">X-ray tube current</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t858h2 t858h4">20%</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t858h0">5.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t858h1 t858h3">Current time product</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t858h2 t858h4">10% plus 0.2 mAs</entry></row></tbody></tgroup></table></TableGroup></Subsection></Section><Section><Label>24.</Label><Subsection><Label>(1)</Label><Text>The controlling timer or automatic exposure control device of diagnostic X-ray equipment must have a minimum irradiation time capability that does not exceed the greater of:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>1/60 s, or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the time required to deliver a current time product of 5 mAs.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>If the automatic exposure control of diagnostic X-ray equipment is selected, the variation in optical density set out in subsection (3) or (4) must be determined using objects that are made of human-tissue equivalent material and have thicknesses that are representative of the actual range of the body thicknesses of the patients.</Text></Subsection><Subsection><Label>(3)</Label><Text>The automatic exposure control device of diagnostic X-ray equipment, other than mammography equipment, when the X-ray tube voltage and the thickness of the objects described in subsection (2) are held constant or varied as specified in columns 1 and 2 of an item of the table to this subsection, must limit the variation in optical density of the resulting radiograms to the quantity set out in column 3 of that item.</Text><TableGroup pointsize="10" spanlanguages="yes" topmarginspacing="4"><Caption position="over" justification="center">TABLE TO SUBSECTION 24(3)</Caption><table frame="topbot"><title justification="center">MAXIMUM VARIATION IN OPTICAL DENSITY</title><tgroup cols="4"><colspec colname="1" colwidth="45.0*" htmlwidth="10%" /><colspec colname="2" colwidth="130.2*" htmlwidth="28%" /><colspec colname="3" colwidth="159.00*" htmlwidth="34%" /><colspec colname="4" colwidth="133.80*" htmlwidth="29%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t859h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t859h1">Column 1</entry><entry colsep="0" rowsep="0" valign="bottom" id="t859h2">Column 2</entry><entry colsep="0" rowsep="0" valign="bottom" id="t859h3">Column 3</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t859h4" headers="t859h1">X-ray Tube Voltage</entry><entry colsep="0" rowsep="1" valign="bottom" id="t859h5" headers="t859h2">Thickness of the Object</entry><entry colsep="0" rowsep="1" valign="bottom" id="t859h6" headers="t859h3">Maximum Variation in Optical Density</entry></row></thead><tbody><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t859h0">1.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t859h1 t859h4">Variable</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t859h2 t859h5">Constant</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t859h3 t859h6">0.15</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t859h0">2.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t859h1 t859h4">Constant</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t859h2 t859h5">Variable</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t859h3 t859h6">0.20</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t859h0">3.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t859h1 t859h4">Variable</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t859h2 t859h5">Variable</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t859h3 t859h6">0.20</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t859h0">4.</entry><entry colname="2" colsep="0" rowsep="0" valign="top" headers="t859h1 t859h4">Constant</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t859h2 t859h5">Constant</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t859h3 t859h6">0.10</entry></row></tbody></tgroup></table></TableGroup></Subsection><Subsection><Label>(4)</Label><Text>The automatic exposure control device of mammography equipment, when both the X-ray tube voltage and the thickness of the objects described in subsection (2) are varied, must limit the variation in optical density of the resulting radiograms to 0.15.</Text></Subsection></Section><Section><Label>25.</Label><Subsection><Label>(1)</Label><Text>For any selected value of X-ray tube voltage within a range determined in accordance with subsection (2), the quotients of the average air kerma or exposure measurement divided by the indicated current time product, obtained at the applicable settings specified in subsection (3), must not differ by more than 0.10 times their sum as determined by the formula</Text><Provision format-ref="indent-1-0" language-align="no" list-item="no"><FormulaGroup><Formula><FormulaText>|<Emphasis style="italic">X</Emphasis><Sub>1</Sub> - <Emphasis style="italic">X</Emphasis><Sub>2</Sub>| ≤ 0,1 (<Emphasis style="italic">X</Emphasis><Sub>1</Sub> + <Emphasis style="italic">X</Emphasis><Sub>2</Sub>)</FormulaText></Formula><FormulaConnector>where</FormulaConnector><FormulaDefinition><FormulaTerm><Emphasis style="italic">X</Emphasis><Sub>1</Sub></FormulaTerm><Text>is the quotient of the average air kerma or exposure</Text></FormulaDefinition><FormulaDefinition><FormulaTerm><Emphasis style="italic">X</Emphasis><Sub>2</Sub></FormulaTerm><Text>is the quotient of the average air kerma or exposure measurement divided by the current time product determined at the second of the two applicable settings specified in subsection (3).</Text></FormulaDefinition></FormulaGroup></Provision></Subsection><Subsection><Label>(2)</Label><Text>The range referred to in subsection (1) is the smaller of</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>40% to 90% of the maximum available X-ray tube voltage, or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the range of X-ray tube voltages specified for the diagnostic X-ray equipment by the manufacturer.</Text></Paragraph></Subsection><Subsection><Label>(3)</Label><Text>The quotients referred to in subsection (1) must be determined at</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>if the X-ray tube current is selected in discrete steps, any two consecutive X-ray tube current settings;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>if the X-ray tube current selection is continuous, any two X-ray tube current settings that differ by a factor of 2 or less;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>if the current time product is selected in discrete steps, any two consecutive current time product settings; or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>if the current time product selection is continuous, any two current time product settings that differ by a factor of 2 or less.</Text></Paragraph></Subsection><Subsection><Label>(4)</Label><Text>If diagnostic X-ray equipment has more than one focal spot, the quotients referred to in subsection (1) must be determined for all combinations of two focal spots that have a nominal focal spot size greater than 0.45 mm, and all combinations of two focal spots that have a nominal focal spot size equal to or less than 0.45 mm at the applicable settings set out in subsection (3).</Text></Subsection></Section><Section><Label>26.</Label><Subsection><Label>(1)</Label><Text>For mammography equipment, the residual radiation behind the image receptor supporting device must not exceed an air kerma measurement of 1.0 µGy or an exposure measurement of 0.115 mR per irradiation when the equipment is operated at</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>its maximum X-ray field and minimum focal spot to image receptor distance; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>its maximum X-ray tube voltage and maximum current time product.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of subsection (1), the air kerma or exposure measurement must be averaged over a detection area that is 100 cm<Sup>2</Sup>, of which no linear dimension is greater than 20 cm, centred at 5 cm from any accessible surface beyond the image receptor supporting device.</Text></Subsection></Section><Section><Label>27.</Label><Subsection><Label>(1)</Label><Text>Mammography equipment must have a minimum rate of radiation output of 7.0 mGy/s or 802 mR/s when the equipment is operated</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>with a molybdenum anode and molybdenum filter;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>with the breast compression device in place between the X-ray source and the detector; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>at an X-ray tube voltage of 28 kV in standard mammography mode at any focal spot to image receptor distance.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of subsection (1), the minimum rate of radiation output must be</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>measured at a position that is 4.5 cm above the patient support; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>averaged over a period of irradiation of 3.0 s.</Text></Paragraph></Subsection></Section><Section><Label>28.</Label><Subsection><Label>(1)</Label><Text>Radioscopic equipment that has a feature described in column 1 of an item of the table to this subsection, other than when radioscopic images are being recorded, must not operate at any combination of X-ray tube voltage and X-ray tube current that results in an air kerma rate that exceeds that set out in column 2 of that item or an exposure rate that exceeds that set out in column 3 of that item:</Text><TableGroup pointsize="10" spanlanguages="yes" topmarginspacing="4"><Caption position="over" justification="center">TABLE TO SUBSECTION 28(1)</Caption><table frame="topbot"><title justification="center">MAXIMUM AIR KERMA OR EXPOSURE RATE OF RADIOSCOPIC EQUIPMENT, OTHER THAN WHEN IT IS RECORDING IMAGES</title><tgroup cols="4" keep-together="yes"><colspec colname="1" colwidth="50.4*" htmlwidth="11%" /><colspec colname="2" colwidth="216.00*" htmlwidth="46%" /><colspec colname="3" colwidth="108.00*" htmlwidth="23%" /><colspec colname="4" colwidth="93.60*" htmlwidth="20%" /><thead><row valign="bottom" topdouble="yes" topmarginspacing="4"><entry morerows="1" colsep="0" rowsep="1" valign="bottom" id="t860h0">Item</entry><entry colsep="0" rowsep="0" valign="bottom" id="t860h1">Column 1</entry><entry colsep="0" rowsep="0" valign="bottom" id="t860h2">Column 2</entry><entry colsep="0" rowsep="0" valign="bottom" id="t860h3">Column 3</entry></row><row valign="bottom" topmarginspacing="4"><entry colsep="0" rowsep="1" valign="bottom" id="t860h4" headers="t860h1">Feature</entry><entry colsep="0" rowsep="1" valign="bottom" id="t860h5" headers="t860h2">Maximum Air Kerma Rate</entry><entry colsep="0" rowsep="1" valign="bottom" id="t860h6" headers="t860h3">Maximum Exposure Rate</entry></row></thead><tbody><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t860h0">1.</entry><entry rowheader="yes" colname="2" colsep="0" rowsep="0" valign="top" id="t860h8" headers="t860h1 t860h4">Not equipped with an automatic intensity control</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t860h1 t860h4 t860h8 t860h2 t860h5">50 mGy/min</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t860h1 t860h4 t860h8 t860h3 t860h6">5.75 R/min</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t860h0">2.</entry><entry rowheader="yes" colname="2" colsep="0" rowsep="0" valign="top" id="t860h12" headers="t860h1 t860h4">Equipped with an automatic intensity control</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t860h1 t860h4 t860h12 t860h2 t860h5">100 mGy/min</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t860h1 t860h4 t860h12 t860h3 t860h6">11.5 R/min</entry></row><row><entry colname="1" colsep="0" rowsep="0" valign="top" headers="t860h0">3.</entry><entry rowheader="yes" colname="2" colsep="0" rowsep="0" valign="top" id="t860h16" headers="t860h1 t860h4">Equipped with both an automatic intensity control and a high-level irradiation control when the latter is activated</entry><entry colname="3" colsep="0" rowsep="0" valign="top" headers="t860h1 t860h4 t860h16 t860h2 t860h5">150 mGy/min</entry><entry colname="4" colsep="0" rowsep="0" valign="top" headers="t860h1 t860h4 t860h16 t860h3 t860h6">17.25 R/min</entry></row></tbody></tgroup></table></TableGroup></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of subsection (1), the air kerma or exposure rate must be determined at a location along the X-ray beam axis that is</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>if the X-ray source is below the table, 1 cm above the table;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>if the X-ray source is above the table, 30 cm above the table and with the X-ray source assembly positioned as closely as possible to the location of the measurement;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>if the equipment is a C-arm radioscope, 30 cm from the input surface of the radioscopic imaging assembly; or</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>if the equipment is a lateral type radioscope, 15 cm from the centre line of the table in the direction of the X-ray source and with the X-ray source assembly positioned as closely as possible to the location of the measurement for all positions of the table.</Text></Paragraph></Subsection></Section><Section><Label>29.</Label><Subsection><Label>(1)</Label><Text>The leakage radiation from the X-ray source assembly of diagnostic X-ray equipment must not exceed an air kerma rate of 1.0 mGy/h or an exposure rate of 115 mR/h when the equipment is operated at the nominal X-ray tube conditions of loading that correspond to the maximum specified energy input in one hour.</Text></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of subsection (1), the rate must be averaged over a detection area of 100 cm<Sup>2</Sup>, of which no linear dimension is greater than 20 cm, that is centred at 1 m from the focal point.</Text></Subsection></Section><Section><Label>30.</Label><Subsection><Label>(1)</Label><Text>If high voltage can appear across the X-ray tube of the diagnostic X-ray equipment, then the radiation emitting from the X-ray source assembly of the equipment must not exceed an air kerma rate of 20.0 µGy/h or an exposure rate of 2.3 mR/h when</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the equipment is operated with its beam limiting device fully open; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the automatic exposure control or the irradiation switch has not been activated.</Text></Paragraph></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of subsection (1), the rate must be averaged over a detection area of 10 cm<Sup>2</Sup>, of which no linear dimension is greater than 5 cm, that is centred at 5 cm from any accessible surface of the X-ray source assembly.</Text></Subsection></Section><Section><Label>31.</Label><Subsection><Label>(1)</Label><Text>Under any operating condition, the radiation from any component of diagnostic X-ray equipment, other than the X-ray source assembly, must not exceed an air kerma rate of 20.0 µGy/h or an exposure rate of 2.3 mR/h.</Text></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of subsection 1, the rate must be averaged over a detection area of 10 cm<Sup>2</Sup>, of which no linear dimension is greater than 5 cm, that is centred at 5 cm from any accessible surface of the component.</Text></Subsection></Section><Section><Label>32.</Label><Subsection><Label>(1)</Label><Text>In the case of radioscopic equipment, the radiation resulting from the transmission of the X-ray beam through, or scattered from, the entrance window of the radioscopic imaging assembly must not exceed an air kerma rate of 2 mGy/h for an entrance air kerma rate of 1 Gy/min or an exposure rate of 2 mR/h for an entrance exposure rate of 1 R/min.</Text></Subsection><Subsection><Label>(2)</Label><Text>For the purposes of subsection (1), the rate must be</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>taken at the applicable location specified in subsection 28(2) for the type or configuration of the equipment;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>taken with an attenuation block made of aluminum and having the dimensions 20 cm by 20 cm by 3.8 cm that is positioned between the point of entrance of the radiation and the image reception area of the radioscopic imaging assembly; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>averaged over a detection area of 100 cm <Sup>2</Sup> that is centred 10 cm from any accessible surface of the radioscopic imaging assembly and in a plane beyond the image receptor.</Text></Paragraph></Subsection></Section><Heading level="1"><Label>PART XIII</Label><TitleText>ULTRASOUND THERAPY DEVICES</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part and in item 13 of Schedule I,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>amplitude modulated wave</DefinedTermEn> means a wave in which the ratio of the temporal maximum pressure amplitude to the root-mean square pressure amplitude, each spatially averaged over the effective radiating surface, is greater than 1.05; (<DefinedTermFr>onde modulée en amplitude</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>applicator</DefinedTermEn> means the part of the ultrasound therapy device designed to transmit ultrasonic power from the transducer to a patient and includes the transducer and any associated housing; (<DefinedTermFr>applicateur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>continuous wave</DefinedTermEn> means a wave in which the ratio of the temporal maximum pressure amplitude to the root-mean square pressure amplitude, each spatially averaged over the effective radiating surface, is less than or equal to 1.05; (<DefinedTermFr>onde entretenue</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>control panel</DefinedTermEn> means that portion of the external surface of the housing on which the user controls are mounted; (<DefinedTermFr>tableau de commande</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>effective intensity</DefinedTermEn><Repealed>[Revoked, SOR/84-930, s. 2]</Repealed></Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>effective radiating area</DefinedTermEn> means the area of the effective radiating surface that encompasses all points at which the ultrasonic intensity is equal to or greater than five per cent of the maximum spatial ultrasonic intensity at the effective radiating surface, expressed in square centimetres (cm<Sup>2</Sup>); (<DefinedTermFr>aire émettrice utile</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>effective radiating surface</DefinedTermEn> means the surface covering all points that are 5 mm from the applicator face; (<DefinedTermFr>surface émettrice utile</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>effective ultrasonic intensity</DefinedTermEn>, with respect to ultrasonic power, means</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>in the case of a focussing applicator, the quotient obtained by dividing the amount of the ultrasonic power by the amount of the focal area, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>in any other case, the quotient obtained by dividing the amount of the ultrasonic power by the amount of the effective radiating area; (<DefinedTermFr>intensité ultrasonore utile</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>focal area</DefinedTermEn> means the area of the focal surface expressed in square centimetres (cm<Sup>2</Sup>); (<DefinedTermFr>aire focale</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>focal length</DefinedTermEn> in respect of a focussing applicator, means the distance in centimetres between</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>that point of the effective radiating surface the coordinates of which are the mean values of the coordinates of the points of that surface, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>that point of the focal surface the coordinates of which are the mean values of the coordinates of the points of that surface; (<DefinedTermFr>longueur focale</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>focal surface</DefinedTermEn> in respect of a focussing applicator, means the beam cross-section with the smallest area; (<DefinedTermFr>surface focale</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>focussing applicator</DefinedTermEn> means an applicator in which the quotient of the area of the surface having the smallest area, in any plane consisting of the points at which the ultrasonic intensity is greater than five per cent of the maximum intensity in that plane, and the effective radiating area is less than one-half; (<DefinedTermFr>applicateur focalisant</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>housing</DefinedTermEn> means a structure that encloses</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>that portion of a fully assembled ultrasound therapy device that supplies the electrical energy to the applicator, and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the electrical circuitry of the device; (<DefinedTermFr>logement</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>maximum intensity</DefinedTermEn><Repealed>[Revoked, SOR/84-930, s. 2]</Repealed></Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>pressure amplitude</DefinedTermEn> means the instantaneous value of the modulating waveform in the following equation:</Text><FormulaGroup><Formula><FormulaText>p(t)=p<Sub>1</Sub>(t)p<Sub>2</Sub>(t)</FormulaText></Formula><FormulaConnector>where </FormulaConnector><FormulaDefinition><FormulaTerm>p(t) </FormulaTerm><Text>is the instantaneous pressure,</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>p<Sub>1</Sub>(t) </FormulaTerm><Text>is the modulating wave form,</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>p<Sub>2</Sub>(t) </FormulaTerm><Text>is the relative amplitude of the carrier wave normalized to a peak height of one,</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>p(t), p<Sub>1</Sub>(t) and p<Sub>2</Sub>(t)</FormulaTerm><Text>are periodic functions of time, t, at any point in space, and</Text></FormulaDefinition><FormulaDefinition><FormulaTerm>the period of p<Sub>1</Sub>(t)</FormulaTerm><Text>is greater than the period p<Sub>2</Sub>(t);</Text></FormulaDefinition></FormulaGroup><Provision format-ref="indent-1-1" language-align="no" list-item="no"><Text>(<DefinedTermFr>amplitude de pression</DefinedTermFr>)</Text></Provision></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>pulse duration</DefinedTermEn> means a time interval, expressed in seconds, beginning at the first time the pressure amplitude exceeds the value that equals the sum of the minimum pressure amplitude and 10 per cent of the difference between the maximum and minimum pressure amplitudes and ending at the last time the pressure amplitude returns to that value; (<DefinedTermFr>durée des impulsions</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>pulse repetition rate</DefinedTermEn> means the repetition frequency of the wave shape that modulates the ultrasonic radiation carrier wave expressed in pulses per second (pps); (<DefinedTermFr>taux de répétition des impulsions</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>service controls</DefinedTermEn> means controls provided by the manufacturer for adjusting the ultrasound therapy device that, under normal conditions of use, are not accessible to the user; (<DefinedTermFr>commandes internes</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>transducer</DefinedTermEn><Repealed>[Revoked, SOR/84-930, s. 2]</Repealed></Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>ultrasonic frequency</DefinedTermEn> means the frequency of the ultrasonic power wave, expressed in kilohertz (kHz) or megahertz (mHz); (<DefinedTermFr>fréquence</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>ultrasonic power</DefinedTermEn> means the total power that is emitted in the form of ultrasonic radiation by the applicator and that is averaged over each cycle of the ultrasonic radiation carrier wave, expressed in watts (W); (<DefinedTermFr>puissance acoustique</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>user controls</DefinedTermEn> means controls provided by the manufacturer for operating the ultrasound therapy device that, under normal conditions of use, are accessible to the user. (<DefinedTermFr>commandes externes</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Subsection><Label>(1)</Label><Text>Every ultrasound therapy device shall be designed and constructed in such a manner that, under the conditions of use specified by the manufacturer, it functions in accordance with section 3 for as long as the device has its original components or replacement components recommended by the manufacturer.</Text></Subsection><Subsection><Label>(2)</Label><Text>Every ultrasound therapy device shall be designed in such a manner that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>all marks, labels and signs are permanently affixed thereon and clearly visible; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>all user controls, meters, lights or other indicators are clearly visible, readily discernible and clearly labelled to indicate their function.</Text></Paragraph></Subsection><Subsection><Label>(3)</Label><Text>Every ultrasound therapy device shall bear</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>on the external surface of its housing</Text><Subparagraph><Label>(i)</Label><Text>the name and address of the manufacturer,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the name and address of the distributor, if the distributor is not the manufacturer,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the type and model designation,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the serial number,</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>the month and year of manufacture,</Text></Subparagraph><Subparagraph><Label>(vi)</Label><Text>the ultrasonic frequencies in kilohertz (kHz) or megahertz (mHz),</Text></Subparagraph><Subparagraph><Label>(vii)</Label><Text>a statement indicating whether the wave produced by the device is a continuous wave or an amplitude modulated wave,</Text></Subparagraph><Subparagraph><Label>(viii)</Label><Text>in the case of a device that produces an amplitude modulated wave,</Text><Clause><Label>(A)</Label><Text>the pulse repetition rate, the pulse duration, the ratio of the temporal maximum effective ultrasonic intensity to the temporal average effective ultrasonic intensity, and a description of the wave shape, where these parameters do not vary depending on the power, and</Text></Clause><Clause><Label>(B)</Label><Text>the pulse repetition rate, the pulse duration, the ratio of the temporal maximum effective ultrasonic intensity to the temporal average effective ultrasonic intensity, and a description of the wave shape, all at temporal maximum ultrasonic power where these parameters do vary depending on the power,</Text></Clause></Subparagraph><Subparagraph><Label>(ix)</Label><Text>the line voltage used for normal operation, and</Text></Subparagraph><Subparagraph><Label>(x)</Label><Text>the ultrasound radiation warning sign described in section 4; and</Text></Subparagraph><Subparagraph><Label>(xi) to (xiii)</Label><Text><Repealed>[Revoked, SOR/84-930, s. 3]</Repealed></Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>on the external surface of each applicator</Text><Subparagraph><Label>(i)</Label><Text>the identification of the type and model of the ultrasound therapy device for which it is designed,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>where an applicator is a focussing applicator, the focal length and the focal area,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>a unique serial number or other unique identification, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the effective radiating area in square centimetres (cm<Sup>2</Sup>).</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(4)</Label><Text>Every ultrasound therapy device shall be designed and constructed to include the following safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>on the control panel, separate indicator lights or other equivalent indicators that have an expected lifetime of at least 5000 hours,</Text><Subparagraph><Label>(i)</Label><Text>to show when the line voltage is “ON” or “OFF”, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>to show when the ultrasonic power is being applied to the applicator;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a power indicator that</Text><Subparagraph><Label>(i)</Label><Text>in the case of a device that produces a continuous wave, shows by a direct reading the level of the temporal average ultrasonic power and the temporal average effective ultrasonic intensity,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>in the case of a device that produces an amplitude modulated wave, shows by a direct reading the level of the temporal maximum ultrasonic power and the temporal maximum effective ultrasonic intensity, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>functions in accordance with section 3;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a clear and reliable indicator of the range used, if the power indicator described in paragraph (<Emphasis style="italic">b</Emphasis>) utilizes two or more different ranges of measurement; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>a timer that</Text><Subparagraph><Label>(i)</Label><Text>terminates the generation of ultrasound after a preset time interval and then returns to zero,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>does not allow the generation of ultrasound with the timer set at zero,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>is adjustable to settings in increments not greater than one minute, and</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>functions in accordance with section 3.</Text></Subparagraph></Paragraph></Subsection><Subsection><Label>(5)</Label><Text>Where an ultrasound therapy device is equipped with an ultrasonic power control, that control shall</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>allow the adjustment of ultrasonic power;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>have a minimum and maximum adjustment that directly relates to the ultrasonic power level indicator; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>function in accordance with section 3.</Text></Paragraph></Subsection></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>3.</Label><Subsection><Label>(1)</Label><Text>Every ultrasound therapy device shall function in such a manner that when the device is operating with its user controls adjusted to yield the maximum temporal average-spatial average effective ultrasonic intensity, such intensity shall not exceed 3 W/cm<Sup>2</Sup>, when measured in accordance with subsection (2).</Text></Subsection><Subsection><Label>(2)</Label><Text>The method used to measure the effective ultrasonic intensity for the purposes of subsection (1) shall produce a result that is at least as accurate as the result that would be produced by using</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>an ultrasound balance radiometer to measure the ultrasonic power; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>an ultrasound detector of dimensions less than one wavelength in water to measure the pulse repetition rate, the pulse duration and the effective radiating area.</Text></Paragraph></Subsection><Subsection><Label>(3)</Label><Text>The power indicator referred to in paragraph 2(4)(<Emphasis style="italic">b</Emphasis>) shall show on the scale of the ultrasonic power control or on the output power meter the ultrasonic power with an accuracy of ±20 per cent when the output is greater than 10 per cent of the maximum ultrasonic power.</Text></Subsection><Subsection><Label>(4)</Label><Text>The timer referred to in paragraph 2(4)(<Emphasis style="italic">d</Emphasis>) shall be accurate to within 30 seconds for settings less than five minutes, to within 10 per cent for settings from five to 10 minutes, and to within one minute for settings greater than 10 minutes.</Text></Subsection><Subsection><Label>(5)</Label><Text>The ultrasonic power output shall remain constant within ±20 per cent of its initial value during one hour of continuous operation, at maximum output and at rated supply line voltage, in water at a temperature of 22°C ± 3°C.</Text></Subsection><Subsection><Label>(6)</Label><Text>The actual ultrasonic frequency of an ultrasound therapy device shall not differ more than ±5% from the ultrasonic frequency of the device that is stated on the external surface of the housing of the device pursuant to subparagraph 2(3)(<Emphasis style="italic">a</Emphasis>)(vi).</Text></Subsection><Subsection><Label>(7)</Label><Text>The effective radiating area shall be kept within ±20 per cent of the rated value given by the manufacturer.</Text></Subsection></Section><Heading level="2"><TitleText>Warning Sign Specifications</TitleText></Heading><Section><Label>4.</Label><Text>The warning sign referred to in subparagraph 2(3)(<Emphasis style="italic">a</Emphasis>)(x) is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of 1 m;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than 2 cm;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION-ULTRASOUND, ATTENTION-ULTRASONS”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “CAUTION-ULTRASOUND, ATTENTION-ULTRASONS”, described by an inverted triangle containing a small rectangle with a curled line attached in the right corner emitting a series of lines</AlternateText><Image source="CRC1370img\CRC1370_ef011.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART XIV</Label><TitleText>ANALYTICAL X-RAY EQUIPMENT</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>The definitions in this section apply in this Part.</Text><Definition generate-in-text="no"><Text><DefinedTermEn>accessory</DefinedTermEn> means any component used on analytical X-ray equipment in order to enable the equipment to perform its intended use, to adapt the equipment to some special purpose, to facilitate use of the equipment, to enhance the performance of the equipment or to enable the function of the equipment to be integrated with that of other equipment. (<DefinedTermFr>accessoire</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>beam-limiting device</DefinedTermEn> means a device that limits the dimensions of a radiation beam. (<DefinedTermFr>dispositif de limitation du faisceau</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>control panel</DefinedTermEn> means that part of analytical X-ray equipment on which are mounted</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>one or more manually-operated controls that regulate all, or some, of the functions of the equipment; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>meters, lights or other indicators that disclose operating factors and conditions. (<DefinedTermFr>poste de commande</DefinedTermFr>)</Text></Paragraph></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>high-voltage generator</DefinedTermEn> means a combination of the components necessary for the production and control of the electrical energy to be supplied to an X-ray tube, which components usually consist of a high-voltage transformer assembly and a control assembly. (<DefinedTermFr>générateur radiologique</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>interlock</DefinedTermEn> means a system that prevents the start or the continued operation of equipment unless certain predetermined conditions prevail. (<DefinedTermFr>verrouillage</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>protective shielding</DefinedTermEn> means material that limits the extent of a radiation beam or attenuates stray radiation. (<DefinedTermFr>barrière de protection radiologique</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>radiation aperture</DefinedTermEn> means an aperture in the protective shielding of a radiation source or a beam-limiting device, that is intended to give passage to the radiation beam. (<DefinedTermFr>fenêtre</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>radiation beam</DefinedTermEn> means a spatial region that is limited in solid angle and that contains a flux of ionizing radiation which originates from a radiation source that is considered a point source. (<DefinedTermFr>faisceau de rayonnement</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>shutter</DefinedTermEn> means a mechanism that opens or closes a radiation aperture to enable or prevent the passage of a radiation beam. (<DefinedTermFr>obturateur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>stray radiation</DefinedTermEn> means all ionizing radiation except that of the specified radiation beam under consideration, but includes that part of the radiation beam which emerges from the material being irradiated. (<DefinedTermFr>rayonnement parasite</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>timing device</DefinedTermEn> means a device that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>integrates or presents the time elapsed during an equipment function; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>changes the state of operation at the end of a predetermined time interval. (<DefinedTermFr>intégrateur de temps</DefinedTermFr>)</Text></Paragraph></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Text>Analytical X-ray equipment shall be designed and constructed in such a manner that, when installed, operated and maintained in accordance with the instructions referred to in paragraph 3(<Emphasis style="italic">a</Emphasis>), it functions in accordance with section 10.</Text></Section><Section><Label>3.</Label><Text>Analytical X-ray equipment shall be accompanied by</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>instructions from the manufacturer as to the installation, interconnection, testing, operation and maintenance of the equipment and its accessories and replacement components; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>information from the manufacturer on the recommended accessories and replacement components.</Text></Paragraph></Section><Section><Label>4.</Label><Text>Analytical X-ray equipment shall bear on the control panel, next to any one switch that turns on an X-ray tube, each of the following labels:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a label bearing an X-radiation warning symbol that</Text><Subparagraph><Label>(i)</Label><Text>is displayed in two contrasting colours,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is clearly visible and readily discernible from a distance of 1 m,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>has no outer dimension that is less than 2 cm,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>bears the words “CAUTION: X-RADIATION — ATTENTION: RAYONNEMENTS X”, and</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>conforms to the following diagram, namely,</Text><ImageGroup position="center"><AlternateText>Warning sign, bearing the words “Caution: X-Rays — Attention: Rayons X”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef012.png" /></ImageGroup></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>a label bearing the words</Text><Provision format-ref="indent-2-2" language-align="no" list-item="no"><Text>“CAUTION — X-RADIATION.  This Equipment Produces High Intensity X-radiation When Energized.  To be Operated Only by Qualified Personnel.</Text></Provision><Provision format-ref="indent-2-2" language-align="no" list-item="no"><Text>ATTENTION — RAYONNEMENTS X.  Cet appareil produit des rayonnements X haute intensité lorsqu’il est sous tension.  Son utilisation est réservée au personnel compétent;” and</Text></Provision></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>a label setting out</Text><Subparagraph><Label>(i)</Label><Text>the name and address of the manufacturer,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the model designation,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>the serial number,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>the date of manufacture, and</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>the country of manufacture.</Text></Subparagraph></Paragraph></Section><Section><Label>5.</Label><Text>Analytical X-ray equipment shall be designed and constructed in such a manner that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>all labels required by this Part are securely affixed to the equipment and clearly visible;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>all controls, meters, lights and other indicators required by this Part are readily discernible and clearly labelled as to function;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the X-ray tube housing, shutters, beam-limiting devices, couplings and protective shielding will not deform or their shielding properties deteriorate;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>when the equipment is assembled, the radiation beam is contained within protective shielding that is equipped with an interlock; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>where a radiation aperture must be open in order to set up or align the equipment, a visual and an aural indicator warns the operator that the radiation aperture is open for the duration that it is open.</Text></Paragraph></Section><Section><Label>6.</Label><Text>Analytical X-ray equipment shall be designed and constructed to include</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a control panel that regulates one or more X-ray tubes and that includes</Text><Subparagraph><Label>(i)</Label><Text>an “ON/OFF” switch equipped with warning lights that indicate when the control panel is energized,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>a key-actuated control that prevents X-radiation production with the key removed,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>an “ON/OFF” X-radiation switch for each X-ray tube, equipped with a warning light that indicates when X-radiation is being produced,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>indicators or meters that identify which X-ray tube is being used and the corresponding applied X-ray tube voltage and X-ray tube current, unless this information is provided on an accessory,</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>a warning light for each X-ray tube that indicates when any shutter on that tube is open, and</Text></Subparagraph><Subparagraph><Label>(vi)</Label><Text>a timing device for each X-ray tube, unless the device is a component of an accessory;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>for each radiation aperture, a shutter that</Text><Subparagraph><Label>(i)</Label><Text>remains closed when not in use,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>requires the operator to open it manually or by means of a computer controlled function,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>is equipped with an interlock in its operating mechanism, unless an alignment device or accessory is integrated into the radiation aperture and that alignment device or accessory is equipped with an interlock,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>can be removed only with special tools, and</Text></Subparagraph><Subparagraph><Label>(v)</Label><Text>is designed with protective shielding;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>labyrinth-type joints, couplings or interfaces between radiation apertures, shutters, protective shielding, beam-limiting devices and accessories; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>where the equipment is designed for use with interchangeable filters,</Text><Subparagraph><Label>(i)</Label><Text>interlocks or other means to ensure that insertion or removal of those filters is possible only when X-radiation is not being produced, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>filter slots that are covered by protective shielding when they are not in use.</Text></Subparagraph></Paragraph></Section><Section><Label>7.</Label><Text>Every accessory supplied by a manufacturer shall be accompanied by information that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>sets out the specific analytical X-ray equipment for which the accessory is designed; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>indicates the design and specifications of the couplings, fittings, interfaces and parts that are necessary for the installation and functioning of that accessory.</Text></Paragraph></Section><Section><Label>8.</Label><Text>Every accessory associated with the production, collimation, transmission or detection of X-radiation shall bear on its external surface a label that sets out, in respect of that accessory,</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>the name and address of the manufacturer;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>the model designation;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>the serial number;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>the date of manufacture; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>the country of manufacture.</Text></Paragraph></Section><Section><Label>9.</Label><Text>Every accessory associated with the production, collimation, transmission or detection of X-radiation shall be designed and constructed in such a manner that it is contained within protective shielding that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>prevents the radiation beam from touching any part of the operator’s body; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is equipped with an interlock.</Text></Paragraph></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>10.</Label><Text>Assembled analytical X-ray equipment, when operated at any of its designed X-ray tube voltage and X-ray tube current specifications, shall function in such a way that the X-radiation emitted does not exceed 0.5 milliroentgen (≈4.38 microgray air kerma) in one hour at a distance of 5 cm from</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>any accessible external surfaces of any X-ray tube housing, beam-limiting device, protective shielding or accessory;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>any radiation aperture, shutter or filter slot that is in the closed position; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>any point on the housing of the high-voltage generator.</Text></Paragraph></Section><Heading level="1"><Label>PART XV</Label><TitleText>CABINET X-RAY EQUIPMENT</TitleText></Heading><Heading level="2"><TitleText>Interpretation</TitleText></Heading><Section><Label>1.</Label><Text>In this Part,</Text><Definition generate-in-text="no"><Text><DefinedTermEn>cabinet</DefinedTermEn> means an enclosure that, independent of existing structures, contains the X-ray generator, detectors and space for the material to be examined, provides radiation attenuation and prevents access to the X-ray beam; (<DefinedTermFr>enceinte</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>central processing unit</DefinedTermEn> means that part of the electronic circuitry that receives signals from the X-ray detection system and subsequently causes an action to be taken on the material being examined; (<DefinedTermFr>unité centrale</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>control panel</DefinedTermEn> means the control panel referred to in paragraph 7(<Emphasis style="italic">a</Emphasis>) of this Part; (<DefinedTermFr>tableau de commande</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>detector</DefinedTermEn> means the image receptor or other device that intereacts with the X-rays to produce a signal corresponding to the intensity of the X-rays incident on it; (<DefinedTermFr>détecteur</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>material</DefinedTermEn> means any substance, object or product subjected to X-ray irradiation for the purpose of obtaining information about that substance, object or product; (<DefinedTermFr>matière</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>model designation</DefinedTermEn> means any combination of letters or figures or both letters and figures by which a device that bears that designation is identified as having characteristics and design features that are uniform; (<DefinedTermFr>désignation du modèle</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>primary X-ray beam</DefinedTermEn> means that X-radiation emitted directly from the target of the X-ray tube and emerging from the port of the X-ray tube housing; (<DefinedTermFr>faisceau primaire de rayons X</DefinedTermFr>)</Text></Definition><Definition generate-in-text="no"><Text><DefinedTermEn>X-ray generator</DefinedTermEn> means an assembly of components, including an X-ray tube and its housing and shielding, designed and constructed for the controlled generation of X-rays. (<DefinedTermFr>producteur de rayons X</DefinedTermFr>)</Text></Definition></Section><Heading level="2"><TitleText>Standards of Design and Construction</TitleText></Heading><Section><Label>2.</Label><Text>Cabinet X-ray equipment shall be designed and constructed in such a manner that, when installed and maintained in accordance with the instructions referred to in section 3, it functions within the standards of functioning described in section 10 for as long as the device has its original components or has replacement components recommended by the manufacturer.</Text></Section><Section><Label>3.</Label><Text>Cabinet X-ray equipment shall be equipped with installation and maintenance instructions furnished by the manufacturer of the device that, if followed, will enable the device to function within the requirements of this Part.</Text></Section><Section><Label>4.</Label><Text>Cabinet X-ray equipment shall bear</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>on the external surface of the X-ray control panel,</Text><Subparagraph><Label>(i)</Label><Text>a warning sign, next to the X-rays “ON” switch, that</Text><Clause><Label>(A)</Label><Text>indicates the possibility of hazardous radiation emission when the device is in operation, and</Text></Clause><Clause><Label>(B)</Label><Text>prohibits unauthorized use,</Text></Clause></Subparagraph><Subparagraph><Label>(ii)</Label><Text>the X-radiation warning sign described in section 11, and</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>a permanent mark or label that sets out, in respect of the device,</Text><Clause><Label>(A)</Label><Text>the name of the manufacturer,</Text></Clause><Clause><Label>(B)</Label><Text>the model designation,</Text></Clause><Clause><Label>(C)</Label><Text>the serial number,</Text></Clause><Clause><Label>(D)</Label><Text>the date of manufacture, and</Text></Clause><Clause><Label>(E)</Label><Text>the city and country of manufacture; and</Text></Clause></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>on the external surface of all doors or panels through which material to be X-rayed is inserted into or removed from the device, the X-radiation warning sign described in section 11.</Text></Paragraph></Section><Section><Label>5.</Label><Text>Cabinet X-ray equipment shall be designed and constructed in such a manner that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>all marks, labels and signs required by this Part are securely affixed to the device and are clearly visible and readily discernible; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>all controls, meters, lights or other indicators required by this Part are readily discernible and clearly labelled or marked with respect to function.</Text></Paragraph></Section><Section><Label>6.</Label><Text>Cabinet X-ray equipment shall have sufficient shielding to enable the device to function within the standards of functioning described in section 10.</Text></Section><Section><Label>7.</Label><Text>Cabinet X-ray equipment shall be designed and constructed to include the following:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a control panel having the following safety features:</Text><Subparagraph><Label>(i)</Label><Text>a power “ON/OFF” switch,</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>a lock of a type that requires the insertion of a key before X-rays can be produced and the removal of the key to terminate production of X-rays,</Text></Subparagraph><Subparagraph><Label>(iii)</Label><Text>a warning light that indicates when the power is “ON”,</Text></Subparagraph><Subparagraph><Label>(iv)</Label><Text>a warning light, in addition to the warning light referred to in subparagraph (iii), that</Text><Clause><Label>(A)</Label><Text>indicates when X-rays are being generated,</Text></Clause><Clause><Label>(B)</Label><Text>where the device is designed to operate with a pulsed X-ray beam, is activated for at least one-half second, and</Text></Clause><Clause><Label>(C)</Label><Text>unless there is a warning light or other indicator referred to in subparagraph (v) incorporated on the control panel, is connected and interlocked in such a manner that X-rays cannot be produced if the warning light malfunctions, and</Text></Clause></Subparagraph><Subparagraph><Label>(v)</Label><Text>unless the warning light referred to in subparagraph (iv) is connected and interlocked in the manner described in clause (iv)(C), a warning light or other indicator that duplicates the functions of the warning light referred to in subparagraph (iv) whose function is not affected if the warning light referred to in that subparagraph malfunctions;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>shielded doors or panels over all access openings designed for insertion or removal of any material to be examined except where the design of the device prevents the insertion of any part of the human body into the primary X-ray beam;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>safety features connecting and interlocking each of the doors and panels referred to in paragraph (<Emphasis style="italic">b</Emphasis>) with at least two independent interlocks to prevent the generation of X-rays if any door or panel is open;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>interlocks on all doors and panels, other than those doors and panels referred to in paragraph (<Emphasis style="italic">b</Emphasis>), that allow access to areas inside the cabinet where the exposure rate to X-rays may exceed 0.5 milliroentgen per hour to prevent the generation of X-rays if any door or panel is opened or removed;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>subject to section 8, one or more controls that require separate operator action to initiate each X-ray exposure; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">f</Emphasis>)</Label><Text>a beam limiting device that</Text><Subparagraph><Label>(i)</Label><Text>ensures that the primary X-ray beam is aligned with the X-ray detector, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>restricts the size of the primary X-ray beam, at the plane of the X-ray detector, so that it does not exceed the maximum size of the detector.</Text></Subparagraph></Paragraph></Section><Section><Label>8.</Label><Text>Paragraph 7(<Emphasis style="italic">e</Emphasis>) does not apply to cabinet X-ray equipment that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>contains a conveyor or other automatic feed system for insertion and removal of material to be examined where the X-ray exposure or sequence of X-ray exposures is initiated automatically by means of a photocell or other material sensing device; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is designed and constructed to include the following additional safety features:</Text><Subparagraph><Label>(i)</Label><Text>where the device requires the continuous presence of an operator during any examination of material, a control or switch of a type that</Text><Clause><Label>(A)</Label><Text>requires continuous pressure by the operator to maintain automatic operation of the device, and</Text></Clause><Clause><Label>(B)</Label><Text>when released, terminates the X-ray exposure or sequence of exposures and stops the conveyor or other automatic feed system,</Text></Clause></Subparagraph><Subparagraph><Label>(ii)</Label><Text>where the device is controlled by a central processing unit and does not require the continuous presence of an operator during any examination of material, provision in the circuitry of the device for the connection of</Text><Clause><Label>(A)</Label><Text>a remote warning light to indicate when X-rays are being generated, and</Text></Clause><Clause><Label>(B)</Label><Text>a control or switch to stop the conveyor or other automatic feed system and terminate the X-ray exposure or sequence of X-ray exposures from a location other than the control panel, and</Text></Clause></Subparagraph><Subparagraph><Label>(iii)</Label><Text>a conveyor or other automatic feed system of sufficient length to prevent the insertion of any part of the human body into any area of the device where the exposure to X-rays exceeds 0.5 milliroentgen per hour.</Text></Subparagraph></Paragraph></Section><Section><Label>9.</Label><Text>Cabinet X-ray equipment that is designed to permit the entry of a person shall be designed and constructed to include the following additional safety features:</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>a switch or control within the cabinet that, when activated,</Text><Subparagraph><Label>(i)</Label><Text>prevents X-rays from being generated, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>cannot be reset, by-passed or overriden from the control panel;</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>controls for initiating X-ray generation that are external to the cabinet and that make it impossible for the initiation of X-ray generation from within the cabinet;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>where any optical or image-monitoring equipment is used within the cabinet and fluoroscopy alignment is required, a control device external to the cabinet designed to remotely focus the device;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>an audible and visible warning signal within the cabinet that</Text><Subparagraph><Label>(i)</Label><Text>is activated for at least ten seconds immediately prior to initiation of X-ray generation after the closing of any door that is designed to permit human access into the cabinet, and</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>is connected in such a way that failure or malfunction of any single component in the device does not cause a simultaneous failure of both the audible and visible warning signal; and</Text></Subparagraph></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>a warning light within the cabinet that, when X-rays are being produced,</Text><Subparagraph><Label>(i)</Label><Text>remains on continuously, or</Text></Subparagraph><Subparagraph><Label>(ii)</Label><Text>where the X-ray generation period is one-half second or less, remains on for at least one-half second.</Text></Subparagraph></Paragraph></Section><Heading level="2"><TitleText>Standards of Functioning</TitleText></Heading><Section><Label>10.</Label><Text>Cabinet X-ray equipment, when fully assembled for use, shall function in such a way that, under all possible operating conditions of X-ray generation and at the maximum possible material handling rate specified by the manufacturer, the exposure rate from leakage radiation, averaged over a time period that is not less than five minutes, does not exceed 0.5 milliroentgen per hour at a distance of five centimetres from</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>any accessible external surface of the device, including the shielded doors or panels referred to in paragraph 7(<Emphasis style="italic">b</Emphasis>), and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>where the design of the device prevents the insertion of any part of the human body into the primary X-ray beam, an imaginary plane surface that is drawn to close those access openings referred to in paragraph 7(<Emphasis style="italic">b</Emphasis>),</Text></Paragraph><ContinuedSectionSubsection><Text>when averaged over a detection area of ten square centimetres.</Text></ContinuedSectionSubsection></Section><Heading level="2"><TitleText>Warning Sign Specifications</TitleText></Heading><Section><Label>11.</Label><Text>The X-radiation warning sign referred to in section 4 is a sign that</Text><Paragraph><Label>(<Emphasis style="italic">a</Emphasis>)</Label><Text>is shown in two contrasting colours;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">b</Emphasis>)</Label><Text>is clearly visible and identifiable from a distance of one metre;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">c</Emphasis>)</Label><Text>has no outer dimensions less than two centimetres;</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">d</Emphasis>)</Label><Text>bears the words “CAUTION, X-RAYS” and “ATTENTION, RAYONS X”; and</Text></Paragraph><Paragraph><Label>(<Emphasis style="italic">e</Emphasis>)</Label><Text>is designed in accordance with the following diagram:</Text><ImageGroup bilingual="yes" position="center"><AlternateText>Warning sign, bearing the words “Caution: X-Rays — Attention: Rayons X”, described by an inverted triangle containing a tube with a circle in the middle emitting dashed lines</AlternateText><Image source="CRC1370img\CRC1370_ef013.png" /></ImageGroup></Paragraph></Section><Heading level="1"><Label>PART XVI</Label><TitleText><Repealed>[Revoked, SOR/88-471, s. 4]</Repealed></TitleText></Heading></RegulationPiece><HistoricalNote><ul><li>SOR/78-407, s. 2;</li><li> SOR/79-229, ss. 2(E), 3;</li><li> SOR/79-920, ss. 1 to 3;</li><li> SOR/80-381, s. 2;</li><li> SOR/80-464, s. 2;</li><li> SOR/81-23, s. 2;</li><li> SOR/81-286, s. 2;</li><li> SOR/81-545, s. 2;</li><li> SOR/82-542, s. 1;</li><li> SOR/82-981, s. 2;</li><li> SOR/83-495, s. 2;</li><li> SOR/84-930, ss. 1 to 5;</li><li> SOR/85-705, ss. 1, 2(F), 3(F), 4(F);</li><li> SOR/85-756, s. 1;</li><li> SOR/85-757, ss. 1 to 3;</li><li> SOR/88-471, ss. 2(E), 3(E), 4;</li><li> SOR/91-408, s. 2(F);</li><li> SOR/93-201, s. 2;</li><li> SOR/94-40, s. 1;</li><li> SOR/97-511, s. 2;</li><li> SOR/2001-252, s. 2;</li><li> SOR/2005-33, s. 2;</li><li> SOR/2006-122, s. 1(F), 2(F), 3(F), 4(F), 5(F), 6(F), 7, 8(F), 9(F) and 10.</li></ul></HistoricalNote><a startdate="20060322">Previous Version</a></Schedule></Regulation>