  1. Weights and Measures Regulations - C.R.C., c. 1605 (Section 270)
    •  (1) In this section,

      VCF

      VCF  means, in respect of a mechanical or electronic automatic temperature compensator, the applicable volume correction factor

      • (a) at the temperature of the measured liquid during each test of the compensator, in the column corresponding to the density at 15°C in kilograms per cubic metre for which the compensator is marked or adjusted, or to the density corresponding to the cubical coefficient of thermal expansion, relative density or API gravity for which the compensator is marked or adjusted, that is set out

        [...]

      • (b) at the temperature of the measured liquid during each test of the compensator, in the column corresponding to the agreed or experimentally determined cubical coefficient of thermal expansion at 15°C for that liquid and for which the compensator is marked or adjusted, that is set out, in the case of special petroleum liquids to which Tables 54A, 54B and 54D of the API Standard are not applicable, in Table 54C of the API Standard, or

      • (c) at the temperature of the measured liquid during each test of the compensator, in the column corresponding to the density at 15°C in grams per litre for which the compensator is marked or adjusted, or to the density corresponding to the cubical coefficient of thermal expansion, relative density or API gravity for which the compensator is marked or adjusted, that is set out, in the case of liquefied petroleum gases, natural gas liquids and asphalts, in Table 54 of the ASTM-IP Petroleum Measurement Tables, with designation ASTM D1250, 1953 Metric Edition. (FCV)

    • (2) Subject to subsections (4) to (6), the limits of error set out in the table to this subsection apply in respect of mechanical automatic temperature compensators used in conjunction with meters in the measurement of liquids.

      TABLE

      Column I Column II Column III
      Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error
      1 Test volume shall be in accordance with specifications established pursuant to section 13 for mechanical automatic temperature compensators for use with any device or class, type or design of device The ratio NET/GROSS shall lie in the range from

      [...]

      The ratio NET/GROSS shall lie in the range from

      [...]

    • (3) In the table to subsection (2) and in subsections (5) and (7),

      • [...]

      • (b)  ΔT is the difference in degrees Celsius

        • (i) between the test temperature and 15°C where the compensator is used with a meter designed to measure liquids at ambient temperature, or

        • (ii) between the test temperature and the mid-range temperature of the compensator, where the compensator is used with a meter designed to measure liquids at temperatures other than ambient temperature, and for the purposes of calculating the limits of error, “ΔT” shall not exceed 15°C except in the case of a compensator used with a meter designed to measure liquids at ambient temperature.

    • [...]

    • (5) In addition to being within applicable limits of error, the NET/GROSS ratios shall agree within 0.001 on at least two of three consecutive tests at the same temperature under controlled conditions.

    • (6) The limits of error for mechanical automatic temperature compensators submitted for approval pursuant to section 3 of the Act shall be one half the limits of error set out in the table to subsection (2).

    • (7) The limits of error set out in the table to this subsection apply in respect of electronic automatic temperature compensators used in conjunction with meters in the measurement of liquids.

      TABLE

      Column I Column II Column III
      Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error
      1 Test volume shall be in accordance with specifications established pursuant to section 13 for electronic automatic temperature compensators for use with any device or class, type or design of device The VCF applied to the gross registration shall be not more than the VCF corresponding to a temperature that is 0.5°C less than the test temperature, and not less than the VCF corresponding to a temperature that is 0.5°C more than the test temperature The VCF applied to the gross registration shall be not more than the VCF corresponding to a temperature that is 1°C less than the test temperature, and not less than the VCF corresponding to a temperature that is 1°C more than the test temperature
    • (8) Where an automatic temperature compensator referred to in subsection (7) is not marked as described in the definition VCF in subsection (1), the printed ticket, card, bill of lading or other printed record of a transaction shall specify the density, relative density, API gravity or cubical coefficient of thermal expansion, as appropriate, for which the compensator is adjusted, for each delivery by the meter.

    [...]


  2. Weights and Measures Regulations - C.R.C., c. 1605 (Section 178)
    •  (1) An electronic machine shall perform within the applicable limits of error, when tested after being calibrated and zeroed, under relative humidity conditions between 10 per cent and 95 per cent, and under one of the following operating conditions:

      • (a) the ambient temperature range is -10°C to +40°C, unless expressly specified otherwise on the approval application and marked on the machine;

      • (b) if the specified minimum temperature is lower than -10°C or the specified maximum temperature is more than +40°C, the machine shall bear markings specifying the minimum and maximum temperatures and shall be approved for use within this expanded temperature range, if Measurement Canada laboratories are equipped to test at those temperatures;

      • (c) when a machine meets the conditions described in paragraphs (a) and (b), it may be used beyond the approved minimum and maximum temperatures if it continues to perform within allowable limits of error; or

      • (d) if the specified minimum temperature is higher than -10°C or the specified maximum temperature is less than +40°C, the machine shall bear markings specifying the minimum and maximum temperatures and shall only be approved for use within this reduced temperature range.

    • (2) The difference between the specified minimum and maximum temperatures for an electronic machine shall not be less than

      [...]

    • (3) An electronic machine shall perform within the applicable limits of error when tested at a constant ambient temperature that does not vary more than 5°C.

    [...]


  3. Weights and Measures Regulations - C.R.C., c. 1605 (SCHEDULE IV : Tolerances for Local Standards)

    [...]

    [...]

    PART XIX

    Tolerances for Standards of Temperature

    Column I Column II
    Item Nominal Value of Local Standard Tolerance
    1 Any temperature from -40°C to +50°C 0.25°C
    2 Any temperature greater than +50°C to +175°C 0.6°C

    [...]


  4. Weights and Measures Regulations - C.R.C., c. 1605 (Section 51)
    •  (1) When an examination is made of a standard quantity commodity that is in a liquid or near liquid state at a temperature of 20°C and has its quantity stated in units of volume, the quantity of the commodity shall be determined at a temperature of 20°C.

    • (2) When an examination is made of a standard quantity commodity that is in a frozen state and has its quantity stated in units of volume, the quantity of the commodity shall be determined at the temperature at which it is customarily offered for sale.

    [...]


  5. Weights and Measures Regulations - C.R.C., c. 1605 (Section 53)

     In this Part and in Schedule IV,

    nominal value

    nominal value  means

    • [...]

    • (e) in the case of a master meter standard, gravimetric prover standard, pressure hydrometer standard, standard of temperature or electricity measuring standard, the value shown by that standard for any particular test, and

    • (f) in the case of a pipe prover standard, the volume of the standard stabilized at a temperature of 15°C and 101.325 kPa internal pressure as shown on the certificate issued for that standard pursuant to section 13 of the Act; (valeur nominale)

    [...]