PART VSpecifications Relating to Devices (continued)
DIVISION XIVolumetric Liquid Meters (continued)
Performance (continued)
- SOR/2005-297, s. 42(F)
266 (1) Subject to subsections (4) and (5), the limits of error set out in the tables to subsections (2) and (3) apply to all positive displacement bulk volumetric liquid meters, including vehicle-mounted meters, that are designed to measure volumes of 180 litres or more, or 40 gallons or more, of refined petroleum products such as gasoline and fuel oils, foodstuffs such as milk, liquid fertilizers of the solution type such as aqua ammonia, or other liquids of low viscosity.
(2) The limits of error set out in the following table apply to meters, generally of a size 65 mm or 2½ inches or smaller that are designed to measure liquids of low viscosity:
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error 1 Any volume of 225 litres or more up to 900 litres or of 50 gallons or more up to 200 gallons 3/16% of known test quantity ¼% of known test quantity (3) The limits of error set out in the following table apply to meters, generally of a size 75 mm or three inches or larger that are designed to measure liquids of low viscosity:
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error 1 Any volume of 1 350 litres or more or of 300 gallons or more 1/8% of known test quantity ¼% of known test quantity (4) The acceptance limits of error and in-service limits of error set out in the tables to subsections (2) and (3) apply to known test quantities equal to or greater than that delivered by the meter in one minute at the maximum operating rate.
(5) The acceptance limits of error set out in Column II of the tables to subsections (2) and (3) apply only to tests made under controlled conditions such as bench tests and, for tests for acceptance limits of error under other conditions, the limits of error set out in Column III of those tables shall apply.
- SOR/89-570, s. 6(F)
- SOR/2017-17, s. 16
- SOR/2018-252, s. 5
267 (1) Subject to subsections (4) and (5), the limits of error set out in the tables to subsections (2) and (3) apply to all positive displacement bulk volumetric liquid meters, including vehicle-mounted meters, that are designed to measure volumes of 180 litres or more or 40 gallons or more of heavy petroleum products, such as bunker oils, foodstuffs such as syrups, liquid cattle feed, fertilizers of high specific gravities or other liquids of high viscosity.
(2) The limits of error set out in the following table apply to meters, generally of a size 65 mm or 2½ inches or smaller, that are used to measure liquids of high viscosity:
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error 1 Any volume of 225 litres or more up to 900 litres or of 50 gallons or more up to 200 gallons ¼% of known test quantity ½% of known test quantity (3) The limits of error set out in the following table apply to meters, generally of a size 75 mm or three inches or larger, that are used to measure liquids of high viscosity:
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error 1 Any volume of 1 350 litres or more, or of 300 gallons or more 3/16% of known test quantity 3/8% of known test quantity (4) The acceptance limits of error and in-service limits of error set out in the tables to subsections (2) and (3) apply to known test quantities equal to or greater than that delivered by the meter in one minute at maximum operating rate.
(5) The acceptance limits of error set out in Column II of the tables to subsections (2) and (3) apply only to tests under controlled conditions such as bench tests, and for tests for acceptance limits of error under other conditions, the limits of error set out in Column III of those tables shall apply.
- SOR/89-570, s. 6(F)
- SOR/2017-17, s. 17
- SOR/2018-252, s. 6(F)
268 (1) Subject to subsection (7), the limits of error set out in the tables to subsections (2) to (6) apply in respect of all meters that are designed to measure liquefied gases at ambient temperatures.
(2) The limits of error set out in the table to this subsection apply in respect of meters of a size from 25 mm to 50 mm or one inch to two inches, when tested using the liquefied gases the meters are intended to measure and using a narrow neck metal standard of the vapour displacement type.
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error Litres Gallons 1 225 or more 50 or more 0.5% of known test quantity 1% of known test quantity (3) The limits of error set out in the table to this subsection apply in respect of meters of a size 25 mm or larger or one inch or larger, except meters referred to in paragraph (4)(a), when tested using the liquefied gases the meters are intended to measure and using a pipe prover standard or master meter standard.
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error Litres Gallons 1 225 or more 50 or more 0.25% of known test quantity 0.5% of known test quantity (4) The limits of error set out in the table to this subsection apply in respect of
(a) meters used to dispense liquid propane to motor vehicles, when tested using the liquefied gases the meters are intended to measure and using a pipe prover standard or master meter standard; and
(b) meters of a size smaller than 25 mm or one inch, when tested using the liquefied gases the meters are intended to measure and using a pipe prover standard or master meter standard.
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error Litres Gallons 1 5 1 60 mL or 2 fluid ounces 120 mL or 4 fluid ounces 2 10 2 85 mL or 3 fluid ounces 170 mL or 6 fluid ounces 3 20 or more 4 or more 0.5% of known test quantity 1% of known test quantity
(5) The limits of error set out in the table to this subsection apply in respect of
(a) meters used to dispense liquid propane to motor vehicles, when tested using the liquefied gases the meters are intended to measure and using a gravimetric prover standard; and
(b) meters of a size smaller than 25 mm or one inch, when tested using the liquefied gases the meters are intended to measure and using a gravimetric prover standard.
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error Litres Gallons 1 from 25 to 45 from 5 to 10 1.5% 2% 2 more than 45 more than 10 0.8% 1%
(6) The limits of error set out in the table to this subsection apply in respect of meters of a size 25 mm or larger or one inch or larger, except meters referred to in paragraph (5)(a), when tested using the liquefied gases the meters are intended to measure and using a gravimetric prover standard.
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error Litres Gallons 1 from 100 to 200 from 20 to 45 0.75% 1% 2 more than 200 more than 45 0.75% 0.75% (7) The acceptance limits of error set out in Column II of the tables to subsections (2) to (6) apply only in respect of tests made under controlled conditions, and for tests made under other conditions the limits of error set out in Column III of those tables apply.
- SOR/89-570, s. 6(F)
- SOR/90-118, s. 36
269 (1) Subject to subsection (4), the limits of error set out in the tables to subsections (2) and (3) apply to all slow-flow meters rated at 115 litres per hour or less or 25 gallons per hour or less for use in trade in pipeline systems to measure fuel oil, stove oil or kerosene supplied to furnaces and other heating equipment, or in any other system where the meter is permanently connected to a point of fuel consumption.
(2) The limits of error set out in the following table apply to meters designed to operate at 2.25 to 22.5 litres or more per hour or 0.5 to five gallons or more per hour:
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error 1 Any volume above the minimum volume set out in specifications established by the Minister for slow-flow meters or for a design of slow-flow meter ½% of known test quantity ¾% of known test quantity (3) The limits of error set out in the following table apply to meters designed to operate at 0.10 to 3.4 litres or more per hour or 0.02 to 0.75 gallons or more per hour:
TABLE
Column I Column II Column III Item Known Test Quantity Acceptance Limits of Error In-Service Limits of Error 1 Any volume above the minimum volume set out in specifications established by the Minister for slow-flow meters or for a design of slow-flow meter (during tests made at 0.5 litres per hour or 0.10 gallons per hour) 3% of known test quantity 4% of known test quantity 2 Any volume above the minimum volume set out in specifications established by the Minister for slow-flow meters or for a design of slow-flow meter (during tests made at 3.4 litres per hour or 0.75 gallons per hour) ½% of known test quantity ¾% of known test quantity (4) The limits of error set out in the tables to subsections (2) and (3) refer to the results of bench tests made in accordance with specifications established by the Minister for slow-flow meters.
- SOR/89-570, s. 6(F)