Government of Canada / Gouvernement du Canada
Symbol of the Government of Canada

Search

Weights and Measures Regulations (C.R.C., c. 1605)

Regulations are current to 2024-10-30 and last amended on 2024-09-20. Previous Versions

PART VSpecifications Relating to Devices (continued)

DIVISION XIVolumetric Liquid Meters (continued)

Design, Composition and Construction (continued)

 Where an automatic temperature-compensated or other type of compensated register is installed on a meter, an uncompensated register shall also be provided that indicates the uncompensated volume measured by the meter.

 The purchaser’s cumulative register on a key-operated self-serve meter shall be designed so that it is readily visible to the purchaser at the time of delivery and will permit comparison with the quantity indication of the delivery shown by the primary indicator.

 A meter shall have a sealable means of adjustment that will permit the meter to be adjusted to within 0.05 per cent of the quantity dispensed through the meter.

  • SOR/90-118, s. 34

Performance

[
  • SOR/2005-297, s. 42(F)
]

 When a meter is tested for acceptance limits of error at any flow rate between its rated maximum flow and its rated minimum flow using a known test quantity of liquid set out in Column I of an item in the applicable table to sections 265 to 270, the meter is within the acceptance limits of error in respect of that quantity if the quantity registered by the meter does not differ from the known test quantity by an amount in excess of the amount set out in Column II of that item.

  • SOR/89-570, s. 6(F)

 When a meter is tested for in-service limits of error at any flow rate between its rated maximum flow and its rated minimum flow using a known test quantity of liquid set out in Column I of an item in the applicable table to sections 265 to 270, the meter is within the in-service limits of error in respect of that quantity if the quantity registered by the meter does not differ from the known test quantity by an amount in excess of the amount set out in Column III of that item.

  • SOR/89-570, s. 6(F)

 Where a meter is tested for acceptance limits of error or in-service limits of error using three consecutive tests conducted at any single rate of flow or at any flow rate of the same pattern that falls between its rated maximum flow and its rated minimum flow, the spread of the test results shall not exceed two fifths of the applicable limit of error for the meter.

  • SOR/89-570, s. 6(F)
  • SOR/2005-297, s. 19

 [Revoked, SOR/90-118, s. 35]

  •  (1) Subject to subsection (4), the limits of error set out in the tables to subsections (2) and (3) apply to slow-flow dispensers incorporating either meters or self-measuring pumps, dispensers for motor fuel of the service station type, meters on refuellers for small aircraft and boats, meters on vehicle tanks and to all other meters for use in measuring automotive or petroleum liquids in retail trade and that are not otherwise mentioned in the tables in this Division.

  • (2) The limits of error set out in the following table apply to meters that register in metric units of volume:

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    LitresMillilitresMillilitres
    10.1024
    20.202.55
    30.50612
    411020
    521530
    653060
    7104080
    82050100
    950100200
    10100220440
    11250 or more3/16% of known test quantity3/8% of known test quantity
  • (3) The limits of error set out in the following table apply to meters that register in Canadian units of volume:

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    PintsFluid OuncesFluid Ounces
    1½1/8¼
    21¼½
    Quarts
    313/8¾
    425/8
    Gallons
    5112
    62
    7524
    8107
    92013
    1050 or more3/16% of known test quantity3/8% of known test quantity
  • (4) The acceptance limits of error set out in Column II of the tables to subsections (2) and (3) apply only to tests made under controlled conditions such as bench tests, and for tests for acceptance limits of error under other conditions, the limits of error set out in Column III of those tables shall apply.

  • SOR/89-570, s. 6(F)
  •  (1) Subject to subsections (4) and (5), the limits of error set out in the tables to subsections (2) and (3) apply to all positive displacement bulk volumetric liquid meters, including vehicle-mounted meters, that are designed to measure volumes of 180 litres or more, or 40 gallons or more, of refined petroleum products such as gasoline and fuel oils, foodstuffs such as milk, liquid fertilizers of the solution type such as aqua ammonia, or other liquids of low viscosity.

  • (2) The limits of error set out in the following table apply to meters, generally of a size 65 mm or 2½ inches or smaller that are designed to measure liquids of low viscosity:

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    1Any volume of 225 litres or more up to 900 litres or of 50 gallons or more up to 200 gallons3/16% of known test quantity¼% of known test quantity
  • (3) The limits of error set out in the following table apply to meters, generally of a size 75 mm or three inches or larger that are designed to measure liquids of low viscosity:

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    1Any volume of 1 350 litres or more or of 300 gallons or more1/8% of known test quantity¼% of known test quantity
  • (4) The acceptance limits of error and in-service limits of error set out in the tables to subsections (2) and (3) apply to known test quantities equal to or greater than that delivered by the meter in one minute at the maximum operating rate.

  • (5) The acceptance limits of error set out in Column II of the tables to subsections (2) and (3) apply only to tests made under controlled conditions such as bench tests and, for tests for acceptance limits of error under other conditions, the limits of error set out in Column III of those tables shall apply.

  • SOR/89-570, s. 6(F)
  • SOR/2017-17, s. 16
  • SOR/2018-252, s. 5
  •  (1) Subject to subsections (4) and (5), the limits of error set out in the tables to subsections (2) and (3) apply to all positive displacement bulk volumetric liquid meters, including vehicle-mounted meters, that are designed to measure volumes of 180 litres or more or 40 gallons or more of heavy petroleum products, such as bunker oils, foodstuffs such as syrups, liquid cattle feed, fertilizers of high specific gravities or other liquids of high viscosity.

  • (2) The limits of error set out in the following table apply to meters, generally of a size 65 mm or 2½ inches or smaller, that are used to measure liquids of high viscosity:

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    1Any volume of 225 litres or more up to 900 litres or of 50 gallons or more up to 200 gallons¼% of known test quantity½% of known test quantity
  • (3) The limits of error set out in the following table apply to meters, generally of a size 75 mm or three inches or larger, that are used to measure liquids of high viscosity:

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    1Any volume of 1 350 litres or more, or of 300 gallons or more3/16% of known test quantity3/8% of known test quantity
  • (4) The acceptance limits of error and in-service limits of error set out in the tables to subsections (2) and (3) apply to known test quantities equal to or greater than that delivered by the meter in one minute at maximum operating rate.

  • (5) The acceptance limits of error set out in Column II of the tables to subsections (2) and (3) apply only to tests under controlled conditions such as bench tests, and for tests for acceptance limits of error under other conditions, the limits of error set out in Column III of those tables shall apply.

  • SOR/89-570, s. 6(F)
  • SOR/2017-17, s. 17
  • SOR/2018-252, s. 6(F)
  •  (1) Subject to subsection (7), the limits of error set out in the tables to subsections (2) to (6) apply in respect of all meters that are designed to measure liquefied gases at ambient temperatures.

  • (2) The limits of error set out in the table to this subsection apply in respect of meters of a size from 25 mm to 50 mm or one inch to two inches, when tested using the liquefied gases the meters are intended to measure and using a narrow neck metal standard of the vapour displacement type.

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    LitresGallons
    1225 or more50 or more0.5% of known test quantity1% of known test quantity
  • (3) The limits of error set out in the table to this subsection apply in respect of meters of a size 25 mm or larger or one inch or larger, except meters referred to in paragraph (4)(a), when tested using the liquefied gases the meters are intended to measure and using a pipe prover standard or master meter standard.

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    LitresGallons
    1225 or more50 or more0.25% of known test quantity0.5% of known test quantity
  • (4) The limits of error set out in the table to this subsection apply in respect of

    • (a) meters used to dispense liquid propane to motor vehicles, when tested using the liquefied gases the meters are intended to measure and using a pipe prover standard or master meter standard; and

    • (b) meters of a size smaller than 25 mm or one inch, when tested using the liquefied gases the meters are intended to measure and using a pipe prover standard or master meter standard.

      TABLE

      Column IColumn IIColumn III
      ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
      LitresGallons
      1  5160 mL or 2 fluid ounces120 mL or 4 fluid ounces
      210285 mL or 3 fluid ounces170 mL or 6 fluid ounces
      320 or more4 or more0.5% of known test quantity1% of known test quantity
  • (5) The limits of error set out in the table to this subsection apply in respect of

    • (a) meters used to dispense liquid propane to motor vehicles, when tested using the liquefied gases the meters are intended to measure and using a gravimetric prover standard; and

    • (b) meters of a size smaller than 25 mm or one inch, when tested using the liquefied gases the meters are intended to measure and using a gravimetric prover standard.

      TABLE

      Column IColumn IIColumn III
      ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
      LitresGallons
      1from 25 to 45from 5 to 101.5%2%
      2more than 45more than 100.8%1%
  • (6) The limits of error set out in the table to this subsection apply in respect of meters of a size 25 mm or larger or one inch or larger, except meters referred to in paragraph (5)(a), when tested using the liquefied gases the meters are intended to measure and using a gravimetric prover standard.

    TABLE

    Column IColumn IIColumn III
    ItemKnown Test QuantityAcceptance Limits of ErrorIn-Service Limits of Error
    LitresGallons
    1from 100 to 200from 20 to 450.75%1%
    2more than 200more than 450.75%0.75%
  • (7) The acceptance limits of error set out in Column II of the tables to subsections (2) to (6) apply only in respect of tests made under controlled conditions, and for tests made under other conditions the limits of error set out in Column III of those tables apply.

  • SOR/89-570, s. 6(F)
  • SOR/90-118, s. 36
 

Date modified: